ASTM F983-1986(2005)e1 Standard Practice for Permanent Marking of Orthopaedic Implant Components.pdf
《ASTM F983-1986(2005)e1 Standard Practice for Permanent Marking of Orthopaedic Implant Components.pdf》由会员分享,可在线阅读,更多相关《ASTM F983-1986(2005)e1 Standard Practice for Permanent Marking of Orthopaedic Implant Components.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 983 86 (Reapproved 2005)1Standard Practice forPermanent Marking of Orthopaedic Implant Components1This standard is issued under the fixed designation F 983; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、 last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1NOTEUnits information was editorially corrected in August 2009.1. Scope1.1 It is common practice for orthopaedic implant manufac
3、-turers to apply permanent identification to implant compo-nents. In this regard, Practice F86 describes recommendedlocations and methods of marking for metallic implants.1.2 The purpose of this practice is to (1) recommend thatorthopaedic implants be permanently marked, and (2) recom-mend practical
4、 amounts of information that should be includedin the marking. It is recognized, however, that marking is notpractical in some cases (see 4.1).1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purpor
5、t to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2F86 P
6、ractice for Surface Preparation and Marking of Me-tallic Surgical Implants3. Methods of Marking3.1 For metallic implants, the procedures described inPractice F86should be followed.3.2 For nonmetallic implants, other methods should bedevised and utilized.3.3 In any case, however, the marking method s
7、hould (a)not compromise implant performance significantly, and (b)provide legibility over the anticipated service life of theimplant.4. Information Included in Permanent Marking4.1 Orthopaedic implants vary widely in size (for example,from wire to total joint prostheses), and the amount ofinformatio
8、n that practically can be included in marking variesaccordingly. Some implants, such as threaded pins and cerclagewire and very small bone screws, do not provide any surfaceswhich can be marked practically.4.2 Standard InformationWhere implant size and shapeallow, it is recommended that the followin
9、g information beincluded in permanent marking:4.2.1 Manufacturer:4.2.2 MaterialThe use of generic names or ASTM stan-dards, or both, in addition to or in place of trade names isrecommended, where applicable.4.2.3 Implant component catalog number or model number.4.2.4 Implant component serial number
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF98319862005E1STANDARDPRACTICEFORPERMANENTMARKINGOFORTHOPAEDICIMPLANTCOMPONENTSPDF

链接地址:http://www.mydoc123.com/p-537664.html