ASTM F2079-2002 Standard Test Method for Measuring Intrinsic Elastic Recoil of Balloon-Expandable Stents《测量气球反冲力的标准试验方法 可膨胀幅度》.pdf
《ASTM F2079-2002 Standard Test Method for Measuring Intrinsic Elastic Recoil of Balloon-Expandable Stents《测量气球反冲力的标准试验方法 可膨胀幅度》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2079-2002 Standard Test Method for Measuring Intrinsic Elastic Recoil of Balloon-Expandable Stents《测量气球反冲力的标准试验方法 可膨胀幅度》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2079 02Standard Test Method forMeasuring Intrinsic Elastic Recoil of Balloon-ExpandableStents1This standard is issued under the fixed designation F 2079; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 The purpose of this test method is to quantify thepercentage by which the diameter of a stent decreases from itsexpande
3、d diameter while still on the delivery balloon to itsrelaxed diameter after deflating the balloon. This test method isappropriate for stents manufactured from a material that isplastically deformed when the stents diameter is increasedfrom its predeployed size to its postdeployed size by mechani-cal
4、 means. This test method may be performed in air at roomtemperature unless there is a known temperature dependence ofthe material, in which case, the temperature at which the test isconducted shall be stated in the report.1.2 This standard does not purport to address all of thesafety concerns, if an
5、y, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Terminology2.1 Definitions:2.1.1 labeled diameter, nthe nominal deployed size of astent as
6、 indicated on its manufacturers label.2.1.2 stent recoil, nthe amount, expressed as a percentage,by which the diameter of a stent changes from the expandeddiameter measured with the stent on the inflated deliveryballoon to the final value measured after deflating the balloon.3. Summary of Test Metho
7、d3.1 A sample device representative of product that will bemarketed is either premounted or mounted on the deliveryballoon at the time of use. The delivery balloon is inflated tothe nominal expansion pressure indicated for the labeled stent.The outer diameter of the stent is measured in two approxi-
8、mately orthogonal rotational positions while the stent is still onthe inflated delivery balloon. The balloon is deflated and theouter diameter of the stent is remeasured in the same positionsat approximately the same location.4. Significance and Use4.1 Minimal stent recoil is a desirable feature of
9、a stentbecause it minimizes the maximum diameter to which a stentmust be expanded to achieve its final relaxed diameter. A stenthaving a high recoil must be expanded to a greater diameter toachieve its final relaxed diameter than a stent having lowrecoil. Practically, excessive expansion of the vess
10、el into whichthe stent is to be implanted may cause tissue damage resultingin a poor immediate result or poor long-term outcome. Stentrecoil is affected by intrinsic properties of the material used toconstruct the stent and the specific geometric design of thestent; therefore, measuring stent recoil
11、 is an essential part ofevaluating the design.5. Apparatus5.1 A means to inflate with noncompressible fluid, typicallywater, the delivery balloon on which the stent is mounted. Themeans used must be capable of achieving the pressure requiredto maintain the expanded diameter of the stent until it can
12、 bemeasured and may include a device to monitor pressure.5.2 A means to measure the outer diameter of the stentwithout deforming the stent. Typically, a calibrated opticalsystem, which does not require contact with the stent, is used.The resolution of the measurement system shall be 0.01 mm orbetter
13、. The accuracy of the system shall be 2 % of reading orbetter.6. Sampling, Test Specimens, and Test Units6.1 Unless otherwise justified, all samples selected fortesting should be taken from fully processed, clinical qualityproduct. It is not required that these devices undergo terminalsterilization.
14、 Cosmetic rejects or other nonclinical samples maybe used if the cause for rejection has been shown not to affectstent recoil.6.2 The number of specimens tested for each unique stentgeometry should be sufficient to meet sampling requirementsfor desired specification limits. In general, a minimum of
15、tenspecimens is recommended. If a single stent geometry isintended to be used for more than one labeled diameter, recoilshall be evaluated for test specimens expanded to the smallestand largest diameters intended for that geometry.6.3 Most stents are comprised of a repetitious continuous1This test m
16、ethod is under the jurisdiction of ASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.30 on Cardiovascular Standards.Current edition approved May 10, 2002. Published July 2002. Originallypublished as F 2079 01. Last previous edition F
17、2079 01.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.pattern or of repeating subunits. If stents are made longer byadditional repetition of the basic geometric pattern or byadding identical subunits, then recoil need be measured o
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF20792002STANDARDTESTMETHODFORMEASURINGINTRINSICELASTICRECOILOFBALLOONEXPANDABLESTENTS 测量 气球 反冲力

链接地址:http://www.mydoc123.com/p-536491.html