ASTM E834-2009(2015) 8084 Standard Practice for Determining Vacuum Chamber Gaseous Environment Using a Cold Finger《使用冷头测定真空室气体环境的标准实施规程》.pdf
《ASTM E834-2009(2015) 8084 Standard Practice for Determining Vacuum Chamber Gaseous Environment Using a Cold Finger《使用冷头测定真空室气体环境的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E834-2009(2015) 8084 Standard Practice for Determining Vacuum Chamber Gaseous Environment Using a Cold Finger《使用冷头测定真空室气体环境的标准实施规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E834 09 (Reapproved 2015)Standard Practice forDetermining Vacuum Chamber Gaseous Environment Usinga Cold Finger1This standard is issued under the fixed designation E834; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision,
2、the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers a technique for collecting samplesof materials that are part of the residual gas envi
3、ronment of anevacuated vacuum chamber. The practice uses a device desig-nated as a “cold finger” that is placed within the environmentto be sampled and is cooled so that constituents of theenvironment are retained on the cold-finger surface.1.2 The practice covers a method for obtaining a samplefrom
4、 the cold finger and determining the weight of the materialremoved from the cold finger.1.3 The practice contains recommendations as to ways inwhich the sample may be analyzed to identify the constituentsthat comprise the sample.1.4 By determining the species that constitute the sample,the practice
5、may be used to assist in defining the source of theconstituents and whether the sample is generally representativeof samples similarly obtained from the vacuum chamber itself.1.5 This practice covers alternative approaches and usagesto which the practice can be put.1.6 The degree of molecular flux a
6、nisotropy significantlyaffects the assurance with which one can attribute characteris-tics determined by this procedure to the vacuum chamberenvironment in general.1.7 The temperature of the cold finger significantly affectsthe quantity and species of materials collected.1.8 The values stated in SI
7、units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.9 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health p
8、ractices and determine the applica-bility of regulatory limitations prior to use. For specificwarning statements, see Section 8.2. Referenced Documents2.1 ASTM Standards:2E177 Practice for Use of the Terms Precision and Bias inASTM Test Methods3. Terminology3.1 pretest cold finger sample residue mas
9、s, Mithe mass ofmaterial collected from the cold finger during the pretestoperation and as measured by the techniques specified inSection 9. The mass is based on a sample volume of 50 mL.3.2 posttest stock sample residue mass, Mfthe mass ofresidue in a sample collected from the cold finger during th
10、eposttest operation and as measured by the technique specifiedin Section 9. The mass is based on a sample volume of 50 mL.3.3 pretest stock sample residue mass, Sithe mass ofresidue in a sample of the solvent (used to obtain the pretestcold finger sample) as measured by the technique specified inSec
11、tion 9. The mass is based on a sample volume of 50 mL.3.4 posttest stock sample residue mass, Sf the mass ofresidue in a sample of the solvent (used to obtain the posttestcold finger sample) as measured by the technique specified inSection 9. The mass is based on a sample volume of 50 mL.3.5 cold fi
12、ngerthe device that is used in collecting thesample of the residual gases in an evacuated vacuum chamber(see Fig. 1).3.6 CFRthe residue collected by the cold finger during thevacuum exposure given in milligrams.4. Summary of Practice4.1 The cold-finger technique provides a method for char-acterizing
13、 the ambiance in a vacuum chamber when thechamber is being operated with or without a test item.1This practice is under the jurisdiction of ASTM Committee E21 on SpaceSimulation and Applications of Space Technology and is the direct responsibility ofSubcommittee E21.05 on Contamination.Current editi
14、on approved Oct. 1, 2015. Published November 2015. Originallyapproved in 1981. Last previous edition approved in 2009 as E834 09. DOI:10.1520/E0834-09R15.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMS
15、tandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.2 In use, the cold finger is installed in the vacuumchamber in such a location as to be exp
16、osed to fluxesrepresentative of those in the general ambiance. (Chamberconditions that will exist under vacuum conditions must beconsidered so as to assess the effects of molecular fluxanisotropy.)4.3 The cold finger is cleaned before the vacuum exposureand a sample of any residue on the surface is
17、obtained. Thepretest cleaning and sampling procedure consists of (a) heatingthe cold finger and scrubbing it with a solution of laboratorydetergent and water; (b) rinsing the cold finger with deminer-alized or distilled water; (c) rinsing the cold finger withisolpropanl as the solvent; (d) obtaining
18、 a sample of anyresidue contained in a second rinse with solvent; and (e)obtaining a sample of the solvent.4.4 The vacuum chamber is then sealed and evacuated; afterreaching a pressure of less than 1 mPa (8 106torr), a coolantis flowed through the cold finger so that materials in theambient environm
19、ent can adhere to the surface. Generally,liquid nitrogen is used as the coolant. Other coolants may beused provided that the coolant temperature is controlled andreported. This coolant flow is continued until the chamberpressure rises to greater than 80 kPa (600 torr) as the chamberis being returned
20、 to room ambient conditions using dry gaseousnitrogen. (WarningToo rapid a repressurization may dis-lodge some of the condensate.)4.5 As soon as possible after the chamber door is opened,the solvent is poured over the cold finger and a samplecontaining any residue from the cold finger is collected.
21、Asecond sample of the solvent is obtained if the solvent is takenfrom a container different than that used under 4.3.4.6 Both the pretest and posttest samples are placed inpreviously cleaned and weighed evaporating dishes. The dishescontaining the samples are placed on a steam bath and thesolvent is
22、 evaporated. The dishes containing the residue arethen weighed using an analytical balance. The samples of thesolvent are similarly handled and any residue weighed. Thedifferences of mass between the pretest residue and posttestresidue is then determined (corrected if necessary for anysignificant re
23、sidue found in the solvent); this difference in massis taken as the residue collected by the cold finger during itsexposure to the vacuum environment, CFR.4.7 Analytical procedures such as infrared spectroscopy orgas chromatography-mass spectrometry may be used to iden-tify those species that consti
24、tute the residue.5. Significance and Use5.1 When applied in the case in which there is no test itemin the vacuum chamber (such as during bake-out operations),this procedure may be used to evaluate the performance of thevacuum chamber in relation to other data from the same orother chambers given tha
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME834200920158084STANDARDPRACTICEFORDETERMININGVACUUMCHAMBERGASEOUSENVIRONMENTUSINGACOLDFINGER 使用

链接地址:http://www.mydoc123.com/p-533637.html