ASTM E459-2005(2016) 0867 Standard Test Method for Measuring Heat Transfer Rate Using a Thin-Skin Calorimeter《采用薄皮热量计测量传热率的标准试验方法》.pdf
《ASTM E459-2005(2016) 0867 Standard Test Method for Measuring Heat Transfer Rate Using a Thin-Skin Calorimeter《采用薄皮热量计测量传热率的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E459-2005(2016) 0867 Standard Test Method for Measuring Heat Transfer Rate Using a Thin-Skin Calorimeter《采用薄皮热量计测量传热率的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E459 05 (Reapproved 2016)Standard Test Method forMeasuring Heat Transfer Rate Using a Thin-SkinCalorimeter1This standard is issued under the fixed designation E459; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the y
2、ear of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the design and use of a thinmetallic calorimeter for measuring heat transfer rate (also
3、called heat flux). Thermocouples are attached to the unexposedsurface of the calorimeter. A one-dimensional heat flow analy-sis is used for calculating the heat transfer rate from thetemperature measurements. Applications include aerodynamicheating, laser and radiation power measurements, and firesa
4、fety testing.1.2 Advantages:1.2.1 Simplicity of ConstructionThe calorimeter may beconstructed from a number of materials.The size and shape canoften be made to match the actual application. Thermocouplesmay be attached to the metal by spot, electron beam, or laserwelding.1.2.2 Heat transfer rate dis
5、tributions may be obtained ifmetals with low thermal conductivity, such as some stainlesssteels, are used.1.2.3 The calorimeters can be fabricated with smoothsurfaces, without insulators or plugs and the attendant tempera-ture discontinuities, to provide more realistic flow conditionsfor aerodynamic
6、 heating measurements.1.2.4 The calorimeters described in this test method arerelatively inexpensive. If necessary, they may be operated toburn-out to obtain heat transfer information.1.3 Limitations:1.3.1 At higher heat flux levels, short test times are neces-sary to ensure calorimeter survival.1.3
7、.2 For applications in wind tunnels or arc-jet facilities,the calorimeter must be operated at pressures and temperaturessuch that the thin-skin does not distort under pressure loads.Distortion of the surface will introduce measurement errors.1.4 The values stated in SI units are to be regarded assta
8、ndard. No other units of measurement are included in thisstandard.1.4.1 ExceptionThe values given in parentheses are forinformation only.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to e
9、stablish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Summary of Test Method2.1 This test method for measuring the heat transfer rate toa metal calorimeter of finite thickness is based on the assump-tion of one-dimensional heat f
10、low, known metal properties(density and specific heat), known metal thickness, and mea-surement of the rate of temperature rise of the back (orunexposed) surface of the calorimeter.2.2 After an initial transient, the response of the calorimeteris approximated by a lumped parameter analysis:q 5 CpdTd
11、(1)where:q = heat transfer rate, W/m2, = metal density, kg/m3, = metal thickness, m,Cp= metal specific heat, J/kgK, anddT/d = back surface temperature rise rate, K/s.3. Significance and Use3.1 This test method may be used to measure the heattransfer rate to a metallic or coated metallic surface for
12、avariety of applications, including:3.1.1 Measurements of aerodynamic heating when the calo-rimeter is placed into a flow environment, such as a windtunnel or an arc jet; the calorimeters can be designed to havethe same size and shape as the actual test specimens tominimize heat transfer corrections
13、;3.1.2 Heat transfer measurements in fires and fire safetytesting;3.1.3 Laser power and laser absorption measurements; aswell as,3.1.4 X-ray and particle beam (electrons or ions) dosimetrymeasurements.1This test method is under the jurisdiction of ASTM Committee E21 on SpaceSimulation and Applicatio
14、ns of Space Technology and is the direct responsibility ofSubcommittee E21.08 on Thermal Protection.Current edition approved April 1, 2016. Published April 2016. Originallyapproved in 1972. Last previous edition approved in 2011 as E459 05 (2011).DOI: 10.1520/E0459-05R16.Copyright ASTM International
15、, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2 The thin-skin calorimeter is one of many concepts usedto measure heat transfer rates. It may be used to measureconvective, radiative, or combinations of convective and ra-diative (usually called mixed or total)
16、 heat transfer rates.However, when the calorimeter is used to measure radiative ormixed heat transfer rates, the absorptivity and reflectivity of thesurface should be measured over the expected radiation wave-length region of the source.3.3 In 4.6 and 4.7, it is demonstrated that lateral heatconduct
17、ion effects on a local measurement can be minimizedby using a calorimeter material with a low thermal conductiv-ity.Alternatively, a distribution of the heat transfer rate may beobtained by placing a number of thermocouples along the backsurface of the calorimeter.3.4 In high temperature or high hea
18、t transfer rateapplications, the principal drawback to the use of thin-skincalorimeters is the short exposure time necessary to ensuresurvival of the calorimeter such that repeat measurements canbe made with the same sensor. When operation to burnout isnecessary to obtain the desired heat flux measu
19、rements, thin-skin calorimeters are often a good choice because they arerelatively inexpensive to fabricate.4. Apparatus4.1 Calorimeter DesignTypical details of a thin-skin calo-rimeter used for measuring aerodynamic heat transfer rates areshown in Fig. 1. The thermocouple wires (0.127 mm OD,0.005 i
20、n., 36 gage) are individually welded to the back surfaceof the calorimeter using spot, electron beam, or laser tech-niques. This type of thermocouple joint (called an intrinsicthermocouple) has been found to provide superior transientresponse as compared to a peened joint or a beaded thermo-couple t
21、hat is soldered to the surface (1, 2).2The wires shouldbe positioned approximately 1.6 mm apart along an expectedisotherm. The use of a small thermocouple wire minimizes heatconduction into the wire but the calorimeter should still berugged enough for repeated measurements. However, when thethicknes
22、s of the calorimeter is on the order of the wire diameterto obtain the necessary response characteristics, the recommen-dations of Sobolik, et al. 1989, Burnett 1961, and Kidd1985 (2-4) should be followed.4.2 When heating starts, the response of the back (unheated)surface of the calorimeter lags beh
23、ind that of the front (heated)surface. For a step change in the heat transfer rate, the initialresponse time of the calorimeter is the time required for thetemperature rise rate of the unheated surface to approach thetemperature rise rate of the front surface within 1 %. Ifconduction heat transfer i
24、nto the thermocouple wire is ignored,the initial response time is generally defined as:r5 0.5Cp2k(2)where:r= initial response time, s, and2The boldface numbers in parentheses refer to the list of references at the end ofthis standard.FIG. 1 Typical Thin-Skin Calorimeter for Heat Transfer Measurement
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME459200520160867STANDARDTESTMETHODFORMEASURINGHEATTRANSFERRATEUSINGATHINSKINCALORIMETER 采用 热量计 测量

链接地址:http://www.mydoc123.com/p-532816.html