ASTM F377-2003 Standard Practice for Calibration of Braking Tractive Measuring Devices for Testing Tires《试验轮胎用刹车 牵引力测量设备校准的标准规程》.pdf
《ASTM F377-2003 Standard Practice for Calibration of Braking Tractive Measuring Devices for Testing Tires《试验轮胎用刹车 牵引力测量设备校准的标准规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F377-2003 Standard Practice for Calibration of Braking Tractive Measuring Devices for Testing Tires《试验轮胎用刹车 牵引力测量设备校准的标准规程》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 377 03Standard Practice forCalibration of Braking/Tractive Measuring Devices forTesting Tires1This standard is issued under the fixed designation F 377; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice gives procedures for the calibration of:reference load cellscalibration platform systems by using a refere
3、nce load cellstatic calibration of braking/tractive force on locked wheelsof tire test trailers, instrumented vehicles, and laboratory tiretesting machines by using the calibration platform system as acalibration fixture.1.2 The values stated in SI units are to be regarded asstandard. The values in
4、parentheses are for information only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory lim
5、itations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 74 Practice for Calibration of Force-Measuring Instru-ments for Verifying the Force Indication of Testing Ma-chinesF 538 Terminology Relating to Characteristics and Perfor-mance of Tires3. Terminology3.1 Definitions:3.2 bias, nthe di
6、fference between the average measuredtest result and the accepted reference value; it measures in aninverse manner the accuracy of a test.3.3 longitudinal force, F, of a tire, nthe component ofthe tire force vector in the X direction.3.4 vertical load, nthe normal reaction of the tire on theroad whi
7、ch is equal to the negative of normal force.4. Summary of Practice4.1 Reference-load cells shall be calibrated using proce-dures and equipment traceable to the National Institute ofStandards and Technology (NIST), or appropriate nationalstandards organization.NOTE 1Practice E 74 may be used as an al
8、ternative method forload-cell calibration.4.2 The calibrated reference-load cell is used to calibrate thecalibration-platform systems or used as the longitudinal forcesensor. The tire test trailer, instrumented vehicle, or laboratorytire testing machine is calibrated with the platform.5. Significanc
9、e and Use5.1 Calibration is essential in the use of various test plat-forms and devices to insure that the test results generated bythese test devices are accurate, repeatable and meaningful. Thisstandard gives the necessary instructions for the calibration ofall of the test devices cited in the Sco
10、pe.6. Apparatus6.1 The calibration system consists of the following basiccomponents:6.1.1 PlatformThe platform on which the test wheel isplaced shall have a flat high friction top surface. The platformshall be of sufficient dimensions to support the entire tirecontact patch throughout the calibratio
11、n process. A low frictionbearing (an air bearing is recommended) permitting freehorizontal motion and capable of sustaining a vertical loadequal to the largest anticipated wheel load shall support theplatform. The longitudinal movement shall be sufficient toobtain the necessary required force levels
12、. The platform mayalso be instrumented with integrated transducers (Force Sen-sors) to measure vertical and longitudinal forces.6.1.2 Force Generation ApparatusA system, capable ofdeveloping a longitudinal force sufficient for the calibration ofthe operating range of the device, shall be used. The f
13、orce shallbe applied along the longitudinal centerline of the platform.6.1.3 Force SensorsSensors to measure applied forces inthe vertical and longitudinal directions that shall be used tocalibrate braking/tractive measuring devices. These sensorscan be integrated force transducers located in the pl
14、atform,1This practice is under the jurisdiction of ASTM Committee F09 on Tires and isthe direct responsibility of Subcommittee F09.10 on Structural Integrity.Current edition approved Dec. 1, 2003. Published January 2004. Originallypublished as F 377 74. Last previous edition F 377 94.2For referenced
15、 ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Cons
16、hohocken, PA 19428-2959, United States.scales or tension load cells. Sensors shall have sufficientmeasurement range to cover the maximum anticipated testwheel forces. Sensors shall be mounted in a manner such thatcross-axis forces will not adversely affect the accuracy. Theforce sensor instrumentati
17、on shall have the capability to allowfor the recording of data, such as a chart recorder, data logger,or visual output that allows for manual logging of data.6.1.4 Reference Load CellThe force standard that pro-vides direct traceability to NIST or appropriate national stan-dards organization that sh
18、all be used to certify other forcesensors or weights used within this standard. If integratedplatform transducers are not used, the reference load cell maybe used as the longitudinal force sensor during calibration ofthe braking/tractive measuring device. The reference load cellshall have sufficient
19、 measurement range to cover the maximumanticipated forces.7. Calibration of the Reference Load Cell and ForceSensors7.1 Three procedures are covered: (1) calibration of thereference-load cell, (2) calibration of the longitudinal forcesensor, and (3) calibration of the vertical load sensor. Repeatthe
20、 appropriate calibrations when any relevant component ischanged or altered (see ASTM Manual 7). Repeat the calibra-tions annually unless the changes between the most recentcalibration equation values and those from the previous cali-bration do not exceed 0.1%. Calibration intervals may belengthened
21、to a maximum of 2 years provided that thesechanges do not exceed 0.1% of the value. Record the ambienttemperature during load cell calibrations. Force sensor gainvalues (or their equivalent) shall be consistent throughoutsections 7.3 and 7.4.7.2 Reference Load Calibration:7.2.1 Set up the reference
22、load cell, indicator and requiredequipment to perform the calibration. List all equipment used,including signal conditioning and output devices. Recordequipment identification and calibration dates.7.2.2 Set the bridge-excitation voltage and gain, or itsequivalent, in accordance with the instruction
23、s on the unitbeing used.7.2.3 Adjust the 9bridge zero,9 or its equivalent, to 9zero.9Be certain that the reference load cell is not being stressed.7.2.4 Using force measuring instruments or weights knownto 0.01% or better of applied force, perform a calibration on thereference load cell. Class F wei
24、ghts of 8.9N (2 lb) or greaterhave a tolerance of 0.01%. Apply the calibration forces in atleast 8 appropriately spaced increments to a maximum of notless than the anticipated maximum force seen in use. Allow allforce indicators to come to rest at each increment and recordthe force readings.7.2.5 De
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF3772003STANDARDPRACTICEFORCALIBRATIONOFBRAKINGTRACTIVEMEASURINGDEVICESFORTESTINGTIRES 试验 轮胎 刹车 牵引力

链接地址:http://www.mydoc123.com/p-532624.html