ASTM E3143-2018a Standard Practice for Performing Cryo-Transmission Electron Microscopy of Liposomes.pdf
《ASTM E3143-2018a Standard Practice for Performing Cryo-Transmission Electron Microscopy of Liposomes.pdf》由会员分享,可在线阅读,更多相关《ASTM E3143-2018a Standard Practice for Performing Cryo-Transmission Electron Microscopy of Liposomes.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E3143 18E3143 18aStandard Practice forPerforming Cryo-Transmission Electron Microscopy ofLiposomes1This standard is issued under the fixed designation E3143; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice covers procedures for vitrifying and recording images of a suspension of liposomes with a cryo-transmiss
3、ionelectron microscope (cryo-TEM) for the purpose of evaluating their shape, size distribution and lamellarity for quality assessment.The sample is vitrified in liquid ethane onto specially prepared holey, ultra-thin, or continuous carbon TEM grids, and imaged ina cryo-holder placed in a cryo-TEM.1.
4、2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish app
5、ropriate safety, health, and environmental practices and determine the applicability ofregulatory limitations prior to use.1.4 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Develo
6、pment of International Standards, Guides and Recommendations issuedby the World Trade Organization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E1617 Practice for Reporting Particle Size Characterization Data2.2 ISO Standards:313322-1 Particle Size Analysis
7、Image Analysis Methods Static Image Analysis Methods3. Terminology3.1 Definitions:3.1.1 anti-contaminator, na specially designed device built into the column of a cryo-transmission electron microscopedesigned to pull contamination produced by outgassing within the column away from the frozen specime
8、n during imaging. Thedevice is cooled with liquid nitrogen to a temperature below that of the specimen creating a colder surface for contamination tobuild on.3.1.2 carbon evaporator, na device used to evaporate carbon in a high vacuum chamber generally by applying current throughtwo carbon rods pres
9、sed against one another, with one rod being sharpened in order to provide resistance to the current; this causesthe rods to heat up and evaporate carbon. The same device can also be used as a glow discharge device (3.1.15) in order to glowdischarge surfaces.3.1.3 continuous carbon grid, na copper el
10、ectron microscope grid coated with a self-supporting layer of continuous carbonover the square mesh of the grid.3.1.4 copper electron microscope grid, ncommonly referred to as an “EM grid,” a thin 3-mm diameter copper foil disk,usually manufactured with a pattern of square holes called a meshthrough
11、 which imaging is conducted in an electron microscope.The number, pattern, and shape of the holes can vary depending on imaging conditions and sample requirements.1 This practice is under the jurisdiction of ASTM Committee E56 on Nanotechnology and is the direct responsibility of Subcommittee E56.02
12、 on Physical and ChemicalCharacterization.Current edition approved Feb. 1, 2018June 1, 2018. Published February 2018July 2018. Originally approved in 2018. Last previous edition approved in 2018 as E3143 18. DOI: 10.1520/E1343-18.10.1520/E1343-18A.2 For referencedASTM standards, visit theASTM websit
13、e, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 Available from International Organization for Standardization (ISO), ISO Central Secretariat, BIBC II, Chemin d
14、e Blandonnet 8, CP 401, 1214 Vernier, Geneva,Switzerland, http:/www.iso.org.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately
15、 depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, P
16、A 19428-2959. United States13.1.5 cryo-grid storage device, nany device used to store cryo-EM grids indefinitely in a liquid nitrogen dewar.3.1.6 cryo-TEM, nalso referred to simply as cryo-EM, or cryo electron microscopy, the process of imaging frozen vitrifiedsamples using a cryo-transmission elect
17、ron microscope.3.1.7 cryo-TEM holder, na liquid nitrogen refrigerated device specifically designed to hold and maintain a prepared gridcontaining a frozen, hydrated, vitrified specimen in a cryo-TEM while imaging is conducted.3.1.8 cryo-TEM plunger, na device used to vitrify a sample onto a holey, u
18、ltra-thin, or continuous carbon grid for cryo-TEMby plunging the grid containing a thin layer of aqueous sample into an ethane slush or other cryogen(s). There are many types ofplunger designs both homemade and commercial. The simplest is the homemade guillotine type (see Fig. 3(B), 8.3).3.1.9 cryo-
19、transmission electron microscope, na specially designed transmission electron microscope (see 3.1.25) capable ofimaging frozen vitrified specimens under low electron dose conditions in order to prevent beam-induced specimen damage.3.1.10 electron micrograph, nthe individual image recorded by the ele
20、ctron microscope as data are collected.3.1.11 electron tomography, nthe process of using images derived from an electron microscope in which the images wererecorded of a sample through an incremental series of tilt angles that are then used with image analysis software to reconstruct thestructure of
21、 the specimen in 3D.3.1.12 ethane, nchemical formula C2H6. A colorless, odorless flammable gas.3.1.13 forceps, na very fine pointed pair of tweezers or pincer used to pick up electron microscope grids.3.1.14 glow discharge, na plasma generated by passing an electric current through a low-pressure ga
22、s environment within achamber, usually a bell jar and with respect to this standard, is used to clean and statically charge the carbon surface coating ofcopper electron microscope grids in order to make them hydrophilic so that they will wet during sample application.3.1.15 glow discharge device, na
23、 device, such as a carbon evaporator (see 3.1.2), designed to generate a glow dischargeplasma.3.1.16 hang time, vthe amount of time after blotting of the sample from the prepared grid, prior to plunging the grid into theliquid ethane for vitrification.3.1.17 holey carbon grid, na copper electron mic
24、roscope grid specially prepared for cryo-TEM consisting of a thin electronsemi-transparent carbon film containing small holes that is suspended over the larger mesh of square holes of the copper electronmicroscope grid (see Fig. 3(A), 8.2.1).3.1.18 image analysis, nthe process of analyzing digital i
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME31432018ASTANDARDPRACTICEFORPERFORMINGCRYOTRANSMISSIONELECTRONMICROSCOPYOFLIPOSOMESPDF

链接地址:http://www.mydoc123.com/p-532375.html