ASTM E3061-2017 Standard Test Method for Analysis of Aluminum and Aluminum Alloys by Inductively Coupled Plasma Atomic Emission Spectrometry (Performance Based Method)《用电感耦合等离子体原子发.pdf
《ASTM E3061-2017 Standard Test Method for Analysis of Aluminum and Aluminum Alloys by Inductively Coupled Plasma Atomic Emission Spectrometry (Performance Based Method)《用电感耦合等离子体原子发.pdf》由会员分享,可在线阅读,更多相关《ASTM E3061-2017 Standard Test Method for Analysis of Aluminum and Aluminum Alloys by Inductively Coupled Plasma Atomic Emission Spectrometry (Performance Based Method)《用电感耦合等离子体原子发.pdf(11页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E3061 17Standard Test Method forAnalysis of Aluminum and Aluminum Alloys by InductivelyCoupled Plasma Atomic Emission Spectrometry(Performance Based Method)1This standard is issued under the fixed designation E3061; the number immediately following the designation indicates the year ofo
2、riginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method describes the inductively coupledplasma ato
3、mic emission spectrometric analysis of aluminumand aluminum alloys for the following elements:ElementsApplication Range, %Minimum MaximumSi 0.02 16.8Fe 0.02 3.06Cu 0.005 7.0Mn 0.003 1.41Mg 0.006 8.2Cr 0.004 0.52Ni 0.004 2.71Zn 0.02 9.65Ti 0.009 0.20Ag 0.003 0.4As 0.005 0.012B 0.009 0.027Ba 0.002 0.0
4、3Be 0.002 0.11Bi 0.01 0.59Ca 0.003 0.048Cd 0.002 0.055Co 0.002 0.034Ga 0.01 0.019Li 0.001 2.48Mo 0.02 0.15Na 0.008 0.026P 0.01 0.025Pb 0.009 0.51Sb 0.01 0.28Sc 0.01 0.065Sn 0.008 6.28Sr 0.0008 0.028Ti 0.005 0.20Tl 0.009 0.13V 0.01 0.12Zr 0.004 0.251.2 This test method has only been interlaboratory t
5、ested forthe elements and ranges specified. It may be possible to extendthis test method to other elements or different compositionranges if method validation, which includes evaluation ofmethod sensitivity and precision and bias (as described inSection 14), is performed. Additionally, the validatio
6、n studymust evaluate the acceptability of sample preparation method-ology using reference materials and/or spike recoveries. Theuser should carefully evaluate the validation data against thelaboratorys data quality objectives. Method validation ofscope extensions is also a requirement of ISO/IEC 170
7、25.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish a
8、ppro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Safety hazardstatements are given in Section 10 and specific warningstatements are given in Sections 15, 17, 18, 19, 20 and 21.2. Referenced Documents2.1 ASTM Standards:2B985 Practice for
9、 Sampling Aluminum Ingots, Billets, Cast-ings and Finished or Semi-Finished Wrought AluminumProducts for Compositional AnalysisD1193 Specification for Reagent WaterE34 Test Methods for Chemical Analysis of Aluminum andAluminum-Base AlloysE50 Practices for Apparatus, Reagents, and Safety Consid-erati
10、ons for Chemical Analysis of Metals, Ores, andRelated MaterialsE135 Terminology Relating to Analytical Chemistry forMetals, Ores, and Related MaterialsE177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE406 Practice for Using Controlled Atmospheres in Spec-trochemical AnalysisE
11、691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method1This test method is under the jurisdiction of ASTM Committee E01 onAnalytical Chemistry for Metals, Ores, and Related Materials and is the directresponsibility of Subcommittee E01.04 on Aluminum and Magne
12、sium.Current edition approved Jan. 15, 2017. Published March 2017. DOI: 10.1520/E3061172For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summar
13、y page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles f
14、or theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.1E716 Practices for Sampling and Sample Preparation ofAluminum and Aluminum Alloys for Determination ofChemical Composition by Spark Atomic Emis
15、sion Spec-trometryE1329 Practice for Verification and Use of Control Charts inSpectrochemical AnalysisE1452 Practice for Preparation of Calibration Solutions forSpectrophotometric and for Spectroscopic Atomic Analy-sis (Withdrawn 2005)3E1479 Practice for Describing and Specifying InductivelyCoupled
16、Plasma Atomic Emission SpectrometersE2857 Guide for Validating Analytical Methods2.2 ISO Standards4ISO/IEC 17025 General Requirements for the Competenceof Calibration and Testing LaboratoriesISO Guide 98-3 Uncertainty of Measurement Part 3:Guide to the Expression of Uncertainty in Measurement(GUM:19
17、95) - First Edition3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology E135.4. Summary of Test Method4.1 The test specimen, in the form of drillings, chips,millings, turnings, small pieces or powder, is dissolved in acaustic solution or a mixture of di
18、lute mineral acids andhydrogen peroxide or sodium nitrite and the resulting solutionsare measured using inductively coupled plasma atomic emis-sion spectrometry. The spectrometer is calibrated using cali-bration solutions prepared to match the sample matrix, using apure aluminum stock solution prepa
19、red in 15.2 and stocksolutions traceable to an SI unit through a national metrologylaboratory or stock solutions prepared as directed in PracticeE1452.5. Significance and Use5.1 This test method for the analysis of aluminum andaluminum alloys is primarily intended to test material forcompliance with
20、 The Aluminum Association Inc.5registeredcomposition limits or other specified composition limits foraluminum and aluminum alloys.5.2 It is assumed that all who use this test method will betrained analysts capable of performing common laboratoryprocedures skillfully and safely, and that the work wil
21、l beperformed in a properly equipped laboratory.5.3 This is a performance-based test method that relies moreon the demonstrated quality of the test result than on strictadherence to specific procedural steps. It is expected thatlaboratories using this test method will prepare their own workinstructi
22、ons. These work instructions should include detailedoperating instructions for the specific laboratory, the specificreference materials employed, and performance acceptancecriteria.6. Interferences6.1 The effect of potential spectral overlap interferences andbackground will vary based on the wavelen
23、gths selected,instrument design, and may vary from instrument to instrumentof the same design. Variation of excitation conditions oroperating parameters may enhance or minimize these interfer-ences. For these reasons, the effect of the potential interfer-ences must be thoroughly investigated for eac
24、h element andmatrix on the instrument chosen for analysis. Practice E1479describes the typical physical and spectral interferences en-countered during the inductively coupled plasma spectrometricanalysis of metal alloys. Potential spectral interferences forrecommended wavelengths are given in Table
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME30612017STANDARDTESTMETHODFORANALYSISOFALUMINUMANDALUMINUMALLOYSBYINDUCTIVELYCOUPLEDPLASMAATOMICEMISSIONSPECTROMETRYPERFORMANCEBASEDMETHOD

链接地址:http://www.mydoc123.com/p-532299.html