ASTM E2995-2014 Standard Specification for ASTM Thermohydrometers with Integral Low-Hazard Thermometers《带集成低危害温度计的ASTM温差比重计标准规格》.pdf
《ASTM E2995-2014 Standard Specification for ASTM Thermohydrometers with Integral Low-Hazard Thermometers《带集成低危害温度计的ASTM温差比重计标准规格》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2995-2014 Standard Specification for ASTM Thermohydrometers with Integral Low-Hazard Thermometers《带集成低危害温度计的ASTM温差比重计标准规格》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E2995 14Standard Specification forASTM Thermohydrometers with Integral Low-HazardThermometers1This standard is issued under the fixed designation E2995; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers glass thermohydrometers ofvarious scale graduation systems, intended for use in a varietyof ASTM
3、Test Methods.1.2 The thermohydrometers presented in this standard arenot exact replacements for the thermohydrometers in Specifi-cation E100; the thermometers have larger maximum permis-sible scale errors than do their Hg-filled counterparts. The usershould decide if these thermohydrometers are appr
4、opriate foruse in his or her application.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
5、 limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D287 Test Method for API Gravity of Crude Petroleum andPetroleum Products (Hydrometer Method)D1250 Guide for Use of the Petroleum Measurement TablesD3290 Specification for Bond and Ledger Papers for Perma-nent Records (Withdrawn 20
6、10)3E77 Test Method for Inspection and Verification of Ther-mometersE100 Specification for ASTM HydrometersE126 Test Method for Inspection, Calibration, and Verifica-tion of ASTM HydrometersE344 Terminology Relating to Thermometry and Hydrom-etry3. Terminology3.1 DefinitionsThe definitions given in
7、Terminology E344apply.3.2 Definitions of Terms Specific to This Standard:3.2.1 ledger paper, na paper characterized by strength,high tearing resistance, eraseability, water resistance, inkreceptivity, uniformity of surface, and smoothness.3.2.1.1 DiscussionOriginally, ledger paper was used espe-cial
8、ly for pen and ink records. Most ledger papers are surfacesized, frequently subjected to appreciable wear, and must havea high degree of permanence and durability.3.2.2 length of the scale, nlength of the nominal range inthe stem, not including graduations extending above and belowthe nominal limits
9、.3.2.3 thermohydrometer, nglass hydrometer having anintegral thermometer.3.2.4 top of the thermohydrometer, ntop of the finishedinstrument.3.2.5 total length, noverall length of the finished thermo-hydrometer.4. Specifications4.1 Thermohydrometers shall conform to the detailed speci-fications in Tab
10、le 1 and to the general requirements specified inSections 515.4.2 Thermohydrometers shall conform to the inspectioncriteria found in Section 16 and the calibration and verificationcriteria found in Section 14.4.3 Thermohydrometers manufactured to previous revisionsof this specification shall retain
11、the same ASTM status as thosemeeting current specifications.4.4 At time of purchase, scale errors shall not exceed themaximum permissible scale error found in Table 1.NOTE 1CautionUsers should be aware that both temperature anddensity indications of thermohydrometers may change with roughhandling, s
12、hock, exposure to aggressive liquids, and thermal cycling,among other factors. Consequently, test results and performance obtainedat the time of manufacture may not necessarily apply throughout anextended period of use. Periodic calibration or verification of theseinstruments, in accordance with pro
13、cedures set forth in Standard TestMethod E126, is recommended1This specification is under the jurisdiction of ASTM Committee E20 onTemperature Measurement and is the direct responsibility of Subcommittee E20.05on Liquid-in-Glass Thermometers and Hydrometers.Current edition approved Dec. 1, 2014. Pub
14、lished December 2014. DOI:10.1520/E2995-142For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved
15、version of this historical standard is referenced onwww.astm.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1TABLE 1 Specifications for ASTM ThermohydrometersAPI Gravity Thermohydrometersfor Petroleum Products and Other Liquids of
16、 Similar Surface Tensions (33 dynes/cm or less)Thermometer Scale in BodyASTM Thermohydrometer No. Nominal API Gravity Range, degS550HL-14 -6 to +6S551HL-14 -1 to 11S552HL-14 9to21S553HL-14 19 to 31S554HL-14 29 to 41S555HL-14 39 to 51S556HL-14 49 to 61S557HL-14 59 to 71S558HL-14 69 to 81S559HL-14 79
17、to 91S560HL-14 89 to 101S561HL-14 37 to 49S562HL-14 64 to 76Thermohydrometer DimensionsTotal length, mm 374 to 387Body diameter, mm 18 to 25Stem diameter, min, mm 4.0Hydrometer ScaleStandard temperature, F 60Subdivisions, API 0.1Intermediate lines at, API 0.5Main (numbered) lines at, API 1.0Scale er
18、ror at any point not to exceed, API 0.1Length of nominal scale, mm 125 to 145Thermometer ScaleRange, F 0to150Immersion totalSubdivisions, F 2Intermediate lines at, F 10Main (numbered) lines at, F 20Scale error at any point not to exceed, F 2Scale length, mm 80 to 110Density Thermohydrometersfor Petr
19、oleum Products and Other Liquids of Similar Surface Tensions (33 dynes/cm or less)Thermometer Scale in BodyASTM Thermohydrometer No. Nominal Range, kg/m3at 15CS500HL-14 600 to 650S501HL-14 650 to 700S502HL-14 700 to 750S503HL-14 750 to 800S504HL-14 800 to 850S505HL-14 850 to 900S506HL-14 900 to 950S
20、507HL-14 950 to 1000S508HL-14 1000 to 1050S509HL-14 1050 to 1100S510HL-14 775 to 825Thermohydrometer DimensionsTotal length, mm 374 to 387Body diameter, mm 18 to 25Stem diameter, min, mm 4.0Hydrometer ScaleStandardization temperature, C 15Subdivisions, kg/m3at 15C 0.5Intermediate lines, kg/m3at 15C
21、1.0Number lines, kg/m3at 15C 5Main (numbered) lines, kg/m3at 15C 10Scale error at any point not to exceed, kg/m3at 15C 0.5Length of nominal scale, mm 125 to 145Thermometer ScaleRange, C -20to65Immersion totalSubdivisions, C 1Intermediate lines at, C 5Main (numbered) lines at, C 10Scale error at any
22、point not to exceed, C 1Scale length, mm 80 to 110E2995 1425. Type5.1 Thermohydrometers shall be of the constant mass,variable displacement type. Thermohydrometers shall be madeof glass, except for the scale, ballast, cement, and the thermo-metric liquid of the thermometer.5.2 The outer surface of t
23、he stem and body shall besymmetrical about the vertical axis. There shall be no unevenor unnecessary thickening of the walls, and no abrupt changesor constrictions that would hinder thorough cleaning or tend totrap air bubbles when the instrument is immersed.5.3 The thermohydrometer shall float with
24、 its axis verticalin the liquids for which it is intended.5.4 The thermohydrometer shall be thoroughly dry on theinside when sealed. The top of the stem shall be neatly roundedwithout unnecessary thickening.5.5 The glass shall be smooth, transparent, and free ofbubbles, strain patterns, cracks, or o
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME29952014STANDARDSPECIFICATIONFORASTMTHERMOHYDROMETERSWITHINTEGRALLOWHAZARDTHERMOMETERS 集成 危害 温度计

链接地址:http://www.mydoc123.com/p-532216.html