ASTM E1965-1998(2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature《间歇测定病人体温用的红外温度计的标准规范》.pdf
《ASTM E1965-1998(2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature《间歇测定病人体温用的红外温度计的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1965-1998(2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature《间歇测定病人体温用的红外温度计的标准规范》.pdf(17页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1965 98 (Reapproved 2009)Standard Specification forInfrared Thermometers for Intermittent Determination ofPatient Temperature1This standard is issued under the fixed designation E 1965; the number immediately following the designation indicates the year oforiginal adoption or, in the
2、case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers electronic instruments in-tended for intermittent measuring an
3、d monitoring of patienttemperatures by means of detecting the intensity of thermalradiation between the subject of measurement and the sensor.1.2 The specification addresses assessing subjects bodyinternal temperature through measurement of thermal emissionfrom the ear canal. Performance requirement
4、s for noncontacttemperature measurement of skin are also provided.1.3 The specification sets limits for laboratory accuracy andrequires determination and disclosure of clinical accuracy ofthe covered instruments.1.4 Performance and storage limits under various environ-mental conditions, requirements
5、 for labeling and test proce-dures are established.NOTE 1For electrical safety consult Underwriters Laboratory Stan-dards.2NOTE 2For electromagnetic emission requirements and tests refer toCISPR 11: 1990 Lists of Methods of Measurement of ElectromagneticDisturbance Characteristics of Industrial, Sci
6、entific, and Medical (ISM)Radiofrequency Equipment.31.5 The values of quantities stated in SI units are to beregarded as the standard. The values of quantities in parenthe-ses are not in SI and are optional.1.6 The following precautionary caveat pertains only to thetest method portion, Section 6, of
7、 this specification: Thisstandard does not purport to address all of the safety concerns,if any, associated with its use. It is the responsibility of the userof this standard to establish appropriate safety and healthpractices and determine the applicability of regulatory limita-tions prior to use.2
8、. Referenced Documents2.1 ASTM Standards:4E 177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE 344 Terminology Relating to Thermometry and Hydrom-etryE 667 Specification for Mercury-in-Glass, Maximum Self-Registering Clinical ThermometersE 1112 Specification for Electronic The
9、rmometer for Inter-mittent Determination of Patient Temperature2.2 International Electrotechnical Commission Standards:IEC 601-1-2: 1993 Medical Electrical Equipment, Part 1;General Requirements for Safety. Collateral Standard:Electromagnetic CompatibilityRequirements and Tests3IEC 1000-4-2: 1995 El
10、ectromagnetic Compatibility(EMC)Part 4: Testing and Measurement Techniques;Section 2: Electrostatic Discharge Immunity Test: BasicEMC Publication (Rev. of IEC 801-2)3IEC 1000-4-3: 1995 Electromagnetic Compatibility32.3 Other Standards:International Vocabulary of Basic and General Terms inMetrology (
11、VIM)33. Terminology3.1 DefinitionsThe definitions given in TerminologyE 344 apply.3.2 Definitions of Terms Specific to This StandardTheterms defined below are for the purposes of this specificationonly. Manufacturers should use this terminology in labelinginstruments and in technical and sales liter
12、ature.3.2.1 accuracy, nability of an infrared thermometer togive a reading close to the true temperature.3.2.2 adjusted mode, noutput of an IR thermometer thatgives temperature measured and calculated from a subject or1This specification is under the jurisdiction of ASTM Committee E20 onTemperature
13、Measurement and is the direct responsibility of Subcommittee E20.08on Medical Thermometry.Current edition approved May 1, 2009. Published July 2009. Originally approvedin 1998. Last previous edition approved in 2003 as E 1965 98 (2003).2Available from Underwriters Laboratories Inc., 1655 Scott Blvd.
14、, Santa Clara,CA 95050.3Available from Global Engineering Documents, 15 Inverness Way East,Englewood, CO 80112.4For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the s
15、tandards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.object, by correcting such temperature for variations in ambi-ent temperature, subjects temperature, emissivity, body site(that is, ora
16、l,orrectal), etc.3.2.3 axillary temperature tba, ntemperature at the apexof either axilla (armpit) as measured by a contact thermometer.3.2.4 blackbody, na reference source of infrared radiationmade in the shape of a cavity and characterized by preciselyknown temperature of the cavity walls and havi
17、ng effectiveemissivity at the cavity opening arbitrarily considered equal tounity.3.2.5 blackbody temperature tBB, ntemperature ofblackbody cavity walls as measured by an imbedded orimmersed contact thermometer.3.2.6 bladder temperature, ntemperature of the interior ofurinary bladder as measured by
18、a contact thermometer.3.2.7 body temperature, ntemperature measured from theinterior of a human body cavity, such as pulmonary artery,distal esophagus, urinary bladder, ear canal, oral, or rectal.3.2.8 clinical accuracy, nability of an infrared ear canalthermometer to give a reading close to true te
19、mperature of thesite that it purports to represent.3.2.9 clinical bias xd, nmean difference between IRthermometer output and an internal body site temperature fromsubjects at specified conditions of ambient temperature andhumidity and averaged over a selected group of subjects.3.2.10 clinical repeat
20、ability sr, npooled standard devia-tion of changes in multiple ear canal temperature readings astaken from the same subject from the same ear with the sameinfrared thermometer by the same operator within a relativelyshort time.3.2.11 combined site offset s, ncalculated difference indegrees of measur
21、ed temperature between a selected referencebody site and ear canal temperature and averaged over thepopulation of representative study samples.3.2.12 contact thermometer, nan instrument that isadapted for measuring temperature by means of thermalconductivity by determining temperature at the moment
22、whennegligible thermal energy flows between the thermometer andthe object of measurement.3.2.13 core temperature tc, ntemperature at a subjectsbody site, such as pulmonary artery, distal esophagus, urinarybladder, or tympanic membrane, recognized as indicative ofinternal body temperature and obtaine
23、d with a contact ther-mometer.3.2.14 mode, nan output of an IR thermometer that givesa representation of a temperature using a disclosed calculationtechnique with respect to selected reference (for example,blackbody, oral, rectal, etc.).3.2.15 displayed temperature range, ntemperature rangein degree
24、s Celsius or Fahrenheit that can be shown by an IRthermometer.3.2.16 IR thermometer type, nan optoelectronic instru-ment that is capable of noncontact infrared temperature mea-surement when placed into the auditory canal of a subject (earcanal type) or from the subjects body surface (skin type).3.2.
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME196519982009STANDARDSPECIFICATIONFORINFRAREDTHERMOMETERSFORINTERMITTENTDETERMINATIONOFPATIENTTEMPERATURE

链接地址:http://www.mydoc123.com/p-529938.html