ASTM E77-2014 red 3405 Standard Test Method for Inspection and Verification of Thermometers《温度计检验和验证的标准试验方法》.pdf
《ASTM E77-2014 red 3405 Standard Test Method for Inspection and Verification of Thermometers《温度计检验和验证的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E77-2014 red 3405 Standard Test Method for Inspection and Verification of Thermometers《温度计检验和验证的标准试验方法》.pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E77 07E77 14Standard Test Method forInspection and Verification of Thermometers1This standard is issued under the fixed designation E77; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last revision.Anumber
2、 in parentheses indicates the year of last reapproval.Asuperscriptepsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope1.1 This test method covers visual and dimensional inspection
3、, test for permanency of pigment, test for bulb stability, inspectionand test for scale accuracy to be used in the verification of liquid-in-glass thermometers as specified in SpecificationSpecificationsE1 and E2251. However, these procedures may be applied to other liquid-in-glass thermometers.2NOT
4、E 1The use of NIST SP250-232 is recommended.1.2 WarningMercury has been designated by EPA and many state agencies as a hazardous material that can cause centralnervous system, kidney and liver damage. Mercury, or its vapor, may be hazardous to health and corrosive to materials. Cautionshould be take
5、n when handling mercury and mercury containing products. See the applicable product Material Safety Data Sheet(MSDS) for details and EPAs website- http:/www.epa.gov/mercury/faq.htm - for additional information. Users should be awarethat selling mercury and/or mercury containing products into your st
6、ate may be prohibited by state law.-1.3 This standard does not purport to address all of the safety problems, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatorylimita
7、tions prior to use.2. Referenced Documents2.1 ASTM Standards:3E1 Specification for ASTM Liquid-in-Glass ThermometersE344 Terminology Relating to Thermometry and HydrometryE2251 Specification for Liquid-in-Glass ASTM Thermometers with Low-Hazard Precision Liquids3. Terminology3.1 Definitions:3.1.1 Th
8、e definitions given in Terminology E344 apply. Some that are considered essential to this standard are given below.3.1.2 calibration, nof a thermometer or thermometric system, the set of operations that establish, under specified conditions,the relationship between the values of a thermometric quant
9、ity indicated by a thermometer or thermometric system and thecorresponding values of temperature realized by standards.3.1.2.1 Discussion(1) The result of a calibration permits either the assignment of values of temperature to indicated values of thermometric quantityor determination of corrections
10、with respect to indications. (2)Acalibration may also determine other metrological properties suchas the effect of influence quantities. (3) The result of a calibration may be communicated in a document such as a calibrationcertificate or a calibration report. (4) The term calibration has also been
11、used to refer to the result of the operations, torepresentations of the result, and to the actual relationship between values of the thermometric quantity and temperature.1 This test method is under the jurisdiction of ASTM Committee E20 on Temperature Measurement and is the direct responsibility of
12、 Subcommittee E20.05 onLiquid-in-Glass Thermometers and Hydrometers.Current edition approved Dec. 1, 2007May 1, 2014. Published January 2008September 2014. Originally approved in 1949. Last previous edition approved in 20032007as E77 98E77 07.(2003). DOI: 10.1520/E0077-07.10.1520/E0077-14.2 “Liquid-
13、in-Glass Thermometer Calibration Service,” NIST Special Publication 250-23, 1988, Superintendent of Documents, U.S. Government Printing Office,Washington, DC 20402-9325.3 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual
14、 Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be te
15、chnically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, P
16、O Box C700, West Conshohocken, PA 19428-2959. United States13.1.3 complete-immersion thermometer, n a liquid-in-glass thermometer, not specified in ASTM documents, designed toindicate temperature correctly when the entire thermometer is exposed to the temperature being measured.3.1.4 partial-immersi
17、on thermometer, n a liquid-in-glass thermometer designed to indicate temperature correctly when thebulb and a specified part of the stem are exposed to the temperature being measured.3.1.5 total-immersion thermometer, na liquid-in-glass thermometer designed to indicate temperature correctly when jus
18、t thatportion of the thermometer containing the liquid is exposed to the temperature being measured.3.2 Definitions of Terms Specific to This Standard:3.2.1 calibration, nthe determination of the indications of a thermometer with respect to temperatures established by astandard resulting in scale co
19、rrections to be applied when maximum accuracy is required.3.2.1 reference point, na temperature at which a thermometer is checked for changes in the bulb volume.3.2.2 verification, nthe process of testing a thermometer for compliance with specifications.3.2.3 verification temperatures, nthe specifie
20、d temperatures at which thermometers are tested for compliance with scale errorlimits.3.2.4 Other descriptions of terms relating to thermometers are included in Sections 3 and 17 of Specification E1.4. Significance and Use4.1 The test method described in this standard will ensure that the thermomete
21、rs listed in SpecificationSpecifications E1 andE2251 will indicate temperatures within the maximum scale errors listed, be compatible with the apparatus, and serve the purposefor which they were designed.Fig. 14.2 Thermometers that do not pass the visual and dimensional inspection tests may give err
22、oneously high or low temperaturereadings, or may not fit into existing equipment used in ASTM methods. If the pigment in the scale etchings washes out or fades,the thermometer will be difficult to read. Improper annealing of the bulb, as determined by the bulb stability test, will result inthermomet
23、er readings rapidly changing with time and use. For accurate temperature measurements the scale readings of thethermometer should be verified as described in this test method.5. Apparatus5.1 Graduated Metal Scales or TemplatesMaximum and minimum specified linear dimensions are measured with graduate
24、dmetal sales and templates on which lines are ruled at suitable distances from reference points corresponding to the maximum andminimum values of the several specified dimensions.5.2 Micrometers and Ring GagesSpecified diameters of ASTM thermometers are checked using micrometers, or moreconveniently
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 1人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME772014RED3405STANDARDTESTMETHODFORINSPECTIONANDVERIFICATIONOFTHERMOMETERS 温度计 检验 验证 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-527626.html