1、Designation: E 319 85 (Reapproved 2008)Standard Practice for theEvaluation of Single-Pan Mechanical Balances1This standard is issued under the fixed designation E 319; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of las
2、t revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.INTRODUCTIONThe balance performs two basic functions: (1) it compares an unknown load with one or moreweights, and (2) it indicates t
3、he difference between the two loads for differences smaller than thesmallest weights normally used on the balance. The test procedure given herein measures the precisionwith which the balance can compare the two loads, and the rates at which systematic errors may affectthe observed difference.1. Sco
4、pe1.1 This practice covers testing procedures for evaluatingthe performance of single-arm balances required by ASTMstandards.1.2 This practice is intended for but not limited to sensitivityratios of 106or better and on-scale ranges of 1000xd or morewhere d = reability either directly or by estimatio
5、n.1.3 This practice can also be applied to other single-panbalances with mechanical weight changing of different capaci-ties or sensitivities with appropriate test loads and calibrationweights.NOTE 1Mechanical balances of this type have largely been replacedby automatic electronic balances incorpora
6、ting a variety of operationalprinciples. Nevertheless, some single-pan mechanical balances are stillmanufactured and many older balances will remain in service for years tocome. One type of automatic electronic balance, the so-called “hybrid,”bears considerable similarity to single-pan mechanical ba
7、lances of the nulltype. (1)21.4 This standard does not purport to address all of thesafety problems, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations
8、prior to use.2. Terminology2.1 Definitions (1):2.1.1 accuracythe degree of agreement of the measure-ments with the true value of the magnitude of the quantitymeasured (2).2.1.2 correction for a weightthe correction for the error inadjustment is:Cr.W 5 A 2 N (1)where:Cr.W = correction for the error i
9、n adjustment to nominalvalue,A = actual value of the weight, andN = nominal value.NOTE 2In practice it is not possible to adjust weights exactly to theirnominal values.2.1.3 correction for error in scale indicationthe correc-tion for the scale indication, I, is:Cr.I 5 A 2 I (2)NOTE 3The correction f
10、or the scale is taken with reference to themeasured value of a weight used as a test load during calibration of theon-scale range.2.1.4 index of precisionthe standard deviation, computedin any acceptable manner, for a collection of measurementsinvolving a given pair of mass standards (3).NOTE 4The s
11、tandard deviation is computed from the data provided bythe instrument precision test (see Section 7) index of precision.2.1.5 null-type balancea balance which requires, as thefinal step in its operation, that the observer restore the angle ofthe balance beam to its original (or null) position. The l
12、eastsignificant figures of the balance indication are obtained fromthis operation.1This practice is under the jurisdiction of ASTM Committee E41 on LaboratoryApparatus and is the direct responsibility of Subcommittee E41.06 on WeighingDevices.Current edition approved Nov. 1, 2008. Published January
13、2009. Originallyapproved in 1968. Last previous edition approved in 2003 as E 319 85 (2003).2The boldface numbers in parentheses refer to the list of references at the end ofthis practice.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United Stat
14、es.2.1.6 optical-type balancein this type the least significantfigures of the balance indication are related to the deflectionangle of the beam from its original (or null) position. A scaleplaced on the moving beam is optically projected onto the(stationary) balance case to provide this indication.2
15、.1.7 precisionthe repeatability of the balance indicationwith the same load under essentially the same conditions.NOTE 5The more closely the measurements are grouped, the smallerthe index of precision will be. The precision must be measured underenvironmental conditions that represent the conditions
16、 under which thebalance is normally used.2.1.8 readabilitythe value of the smallest decimal subdi-vision of a scale division in terms of mass units, that can beread, when the balance is read in the intended manner.NOTE 6The readability of a particular instrument is not a measure ofits performance as
17、 a weighing device. The relationship between thenumerical value obtained by reading devices and the ability of the operatorto estimate the location of the reference device or index is important. It ispossible to introduce a large number of readable subdivisions of the mainscale divisions that would
18、increase the “readability” as defined but if thereading device cannot be reset to the same numerical value when the beamis in an immovable condition, or when the load on the beam is a constantvalue, the readability becomes meaningless. Readability substantially lessthan 1 standard deviation as deter
19、mined by repeated measurement with agiven test weight is usually superfluous.2.1.9 scale divisionthe smallest graduated interval subdi-vided either by estimation or with the aid of a vernier.Subdivisions which appear as divisions on the vernier are notconsidered to be scale divisions, but rather par
20、ts of scaledivisions.2.1.10 sensitivity weighta small weight used to measurethe “on-scale” deflection of the balance indicator.NOTE 7With single-pan balances the sensitivity weight should beequal to the value of the smallest built-in weight represented by the firststep on the dial for the smallest w
21、eights.2.1.11 test loada load chosen to represent the sample loadin the test procedure.2.1.12 value of the divisionthe change in load required tochange the balance indication by one scale division. Thereciprocal of the sensitivity is its most useful function.3. Summary of Practice3.1 The accuracy of
22、 the direct-reading scale, the smallestweight of the set of built-in weights, and uniformity ofsensitivity between the upper and lower halves of full-scaledeflections are verified by preliminary tests.3.2 Estimates of rate of change of the zero with time, rate ofchange in the value of the scale divi
23、sion with time, and aquantitative measure of the variability or random error areprovided by short tests for precision and bias.3.3 An overall test of the direct-reading capability is pro-vided by tests of the built-in weights.4. Significance and Use4.1 Monitoring Weighing PerformanceThis practice pr
24、o-vides results in the form of control charts which measure theweighing capability at the time of the test. A series of tests atappropriate intervals will monitor balance performance over aperiod of time. A marked change from expected performancemay result from a variety of causes including: maladju
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME31919852008952STANDARDPRACTICEFORTHEEVALUATIONOFSINGLEPANMECHANICALBALANCES 机械 天平 评定 标准 实施 规程 PDF

链接地址:http://www.mydoc123.com/p-527392.html