ASTM D6186-2008 752 Standard Test Method for Oxidation Induction Time of Lubricating Oils by Pressure Differential Scanning Calorimetry (PDSC)《利用压差扫描量热法(PDSC)测试润滑油的氧化感应时间的标准试验方法》.pdf
《ASTM D6186-2008 752 Standard Test Method for Oxidation Induction Time of Lubricating Oils by Pressure Differential Scanning Calorimetry (PDSC)《利用压差扫描量热法(PDSC)测试润滑油的氧化感应时间的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D6186-2008 752 Standard Test Method for Oxidation Induction Time of Lubricating Oils by Pressure Differential Scanning Calorimetry (PDSC)《利用压差扫描量热法(PDSC)测试润滑油的氧化感应时间的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 6186 08An American National StandardStandard Test Method forOxidation Induction Time of Lubricating Oils by PressureDifferential Scanning Calorimetry (PDSC)1This standard is issued under the fixed designation D 6186; the number immediately following the designation indicates the year
2、oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of oxidationin
3、duction time of lubricating oils subjected to oxygen at 3.5MPa (500 psig) and temperatures between 130 and 210C.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.2.1 ExceptionPressure measurement appears in MPawith psig prov
4、ided in parentheses for information only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory
5、 limitations prior to use.2. Terminology2.1 Definitions of Terms Specific to This Standard:2.1.1 extrapolated onset time, na time determined on athermal curve, as the intersection of the extrapolated baselineand a line tangent to the oxidation exotherm constructed at itsmaximum rate.2.1.2 oxidation
6、induction time, (OIT), na period of timeduring which the oxidation rate accelerates from zero to amaximum and which corresponds to the extrapolated onsettime.2.1.3 thermal curve, na graph of sample heat flow versustime.3. Summary of Test Method3.1 A small quantity of oil is weighed into a sample pan
7、 andplaced in a test cell. The cell is heated to a specifiedtemperature and then pressurized with oxygen. The cell is heldat a regulated temperature and pressure until an exothermicreaction occurs. The extrapolated onset time is measured andreported as the oxidation induction time for the lubricatin
8、g oilat the specified test temperature.4. Significance and Use4.1 Oxidation induction time, as determined under theconditions of this test method, may be used as an indication ofoxidation stability.2This test method is faster than other oiloxidation tests and requires a very small amount of sample.
9、Itmay be used for research and development, quality control, andspecification purposes. However, no correlation has beenestablished between the results of this test method and serviceperformance.5. Apparatus5.1 Pressure Differential Scanning Calorimeter (PDSC),equipped with the following items:5.1.1
10、 Sample Enclosure, with capability to 3.5 6 0.2 MPa(500 6 25 psig) at 210C and pressure gauge graduated atintervals of 200 KPa (28.6 psig) or less.5.1.2 Thermal Analyzer.5.1.3 Aluminum Solid Fat Index (SFI) Sample PanSeeNote 1.5.1.4 Oxidation Stability Software.5.1.5 Calibration Software.5.1.6 Calib
11、rated Flowmeter, with a capacity of at least 200mL/min and graduated in intervals of 5 mL or less.5.1.7 Sample Encapsulation Press.NOTE 1It has been found that when oil samples are prepared with SFIpans which have more consistent surface areas than standard flat bottompans, reproducibility is improv
12、ed.NOTE 2Stainless steel or copper tubing is compatible with oxygen.NOTE 3See Fig. 1 for a diagram of a typical test unit.6. Reagents and Materials6.1 Oxygen, a minimum purity of 99.5 % oxygen by vol-ume. (WarningOxidizer. Gas under pressure.)6.2 Indium, of not less than 99.9 % indium by mass.7. Cal
13、ibration7.1 Sample Temperature Calibration:1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.09.0D on Oxidation of Lubricants.Current edition approved Dec. 1, 2008. Published January 2009. Origina
14、llyapproved in 1997. Last previous edition approved in 2003 as D 618698(2003)1.2Rhee, In-Sik, “Development of New Oxidation Stability Test Method forLubricating Oils Using a Pressure Differential Scanning Calorimeter (PDSC),”NLGI Spokesman, Vol 65, No. 3, June 2001, pp. 1623.1*A Summary of Changes s
15、ection appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.7.1.1 Weigh approximately 10 mg of indium into an alumi-num sample pan, insert a lid and crimp the lid to the pan usingthe encapsulation press.
16、 Place the crimped pan onto the sampleplatform in the pressure cell. Seal an empty pan in the samemanner and place it on the reference platform. Set the cellcover in place and close the cell.7.1.2 Open the oxygen cylinder valve slightly and set apressure of 3.5 6 0.2 MPa (500 6 25 psig) on the cell
17、inlet linewith the pressure regulator. Partially open the inlet valve on thecell and allow the pressure to slowly build up in the cell. Thisrequires approximately 2 min. Using the outlet valve, adjustand maintain the oxygen purge rate through the flowmeter at100 6 10 mL/min.7.1.3 Set the thermal ana
18、lyzer to heat from ambient tem-perature (approximately 22C) to 180C at a programmed rateof 10C/min. After completion of the run measure the meltingtemperature of the indium. If the melting temperature differsfrom 157.4 6 0.2C (see Note 4) correct the difference byusing either the hardware or softwar
19、e calibration proceduredescribed in the manufacturers instruction manual. If thehardware calibration procedure is used, perform the tempera-ture correction under 3.5 MPa (500 psig) oxygen pressure witha 100 mL/min purge rate. A typical melting calibration curve isshown in Fig. 2.NOTE 4The melting te
20、mperature of indium is 156.6C at atmosphericpressure, but has been found to be elevated to 157.4C under theconditions of this test method, 3.5 MPa (500 psig) of oxygen.7.2 Temperature Controller Calibration:7.2.1 Remove both the sample pan and the reference panfrom the cell, then close the cell. Slo
21、wly pressurize the cellwith 3.5 6 0.2 MPa (500 6 25 psig) oxygen and adjust thepurge rate to 100 6 10 mL/min using the cell outlet valve.Select the desired test temperature (either 210, 180, 155, or130C).7.2.2 Program the cell to maintain the selected test tempera-ture. If, after 10 min, the display
22、ed cell temperature differs bymore than 6 0.2C from the selected temperature, slowlyadjust the temperature controller until they agree.After makingan adjustment, wait at least 5 min to make certain that thetemperature is stable before continuing. If the PDSC equip-ment does not have this function, t
23、he control calibration shallbe followed according to the equipment manufacturers rec-ommendations.7.3 Cell Base Pressure Gauge CalibrationConduct thecalibration using a calibrated pressure transducer or a previ-ously calibrated gauge according to the pressure cell manufac-turers instructions.8. Proc
24、edure8.1 Before starting a test, the control thermocouple calibra-tion shall be conducted at the test temperature (either 210, 180,155, or 130C) according to 7.2.1 and 7.2.2. When the testtemperature is not known, conduct the calibration at 210C.8.2 Weigh 3.0 6 0.2 mg of oil into a new sample pan.Sp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD61862008752STANDARDTESTMETHODFOROXIDATIONINDUCTIONTIMEOFLUBRICATINGOILSBYPRESSUREDIFFERENTIALSCANNINGCALORIMETRYPDSC

链接地址:http://www.mydoc123.com/p-521800.html