《ASTM D6186-1998(2003)e1 Standard Test Method for Oxidation Induction Time of Lubricating Oils by Pressure Differential Scanning Calorimetry (PDSC)《利用压差扫描量热法(PDSC)测试润滑油的氧化感应时间的标准试验方.pdf》由会员分享,可在线阅读,更多相关《ASTM D6186-1998(2003)e1 Standard Test Method for Oxidation Induction Time of Lubricating Oils by Pressure Differential Scanning Calorimetry (PDSC)《利用压差扫描量热法(PDSC)测试润滑油的氧化感应时间的标准试验方.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 6186 98 (Reapproved 2003)e1An American National StandardStandard Test Method forOxidation Induction Time of Lubricating Oils by PressureDifferential Scanning Calorimetry (PDSC)1This standard is issued under the fixed designation D 6186; the number immediately following the designation
2、 indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEWarning notes were editorially moved into
3、the standard text in August 2003.1. Scope1.1 This test method covers the determination of oxidationinduction time of lubricating oils subjected to oxygen at 3.5MPa (500 psig) and temperatures between 130 and 210C.1.2 The values stated in SI units are to be regarded as thestandard.1.3 This standard d
4、oes not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Terminology2.1 Definitions of T
5、erms Specific to This Standard:2.1.1 extrapolated onset time, na time determined on athermal curve, as the intersection of the extrapolated baselineand a line tangent to the oxidation exotherm constructed at itsmaximum rate.2.1.2 oxidation induction time, (OIT), na period of timeduring which the oxi
6、dation rate accelerates from zero to amaximum and which corresponds to the extrapolated onsettime.2.1.3 thermal curve, na graph of sample heat flow versustime.3. Summary of Test Method3.1 A small quantity of oil is weighed into a sample pan andplaced in a test cell. The cell is heated to a specified
7、temperature and then pressurized with oxygen. The cell is heldat a regulated temperature and pressure until an exothermicreaction occurs. The extrapolated onset time is measured andreported as the oxidation induction time for the lubricating oilat the specified test temperature.4. Significance and U
8、se4.1 Oxidation induction time, as determined under theconditions of this test method, may be used as an indication ofoxidation stability.2This test method is faster than other oiloxidation tests and requires a very small amount of sample. Itmay be used for research and development, quality control,
9、 andspecification purposes. However, no correlation has beenestablished between the results of this test method and serviceperformance.5. Apparatus5.1 Pressure Differential Scanning Calorimeter (PDSC),equipped with the following items:5.1.1 Sample Enclosure, with capability to 3.5 6 0.2 MPa(500 6 25
10、 psig) at 210C and pressure gage graduated atintervals of 200 KPa (28.6 psig) or less.5.1.2 Thermal Analyzer.5.1.3 Aluminum Solid Fat Index (SFI) Sample PanSeeNote 1.5.1.4 Oxidation Stability Software.5.1.5 Calibration Software.5.1.6 Calibrated Flowmeter, with a capacity of at least 200mL/min and gr
11、aduated in intervals of 5 mL or less.5.1.7 Sample Encapsulation Press.NOTE 1It has been found that when oil samples are prepared with SFIpans which have more consistent surface areas than standard flat bottompans, reproducibility is improved.NOTE 2Stainless steel or copper tubing is compatible with
12、oxygen.NOTE 3See Fig. 1 for a diagram of a typical test unit.6. Reagents and Materials6.1 Oxygen, a minimum purity of 99.5 % oxygen by vol-ume. (WarningOxidizer. Gas under pressure.)6.2 Indium, of not less than 99.9 % indium by mass.1This test method is under the jurisdiction of ASTM Committee D02 o
13、nPetroleum Products and Lubricants and is the direct responsibility of SubcommitteeD02.09 on Oxidation.Current edition approved May 10, 2003. Published August 2003. Originallyapproved in 1997. Last previous edition approved in 1998 as D 618698.2Rhee, In-Sik, “Development of New Oxidation Stability T
14、est Method forLubricating Oils Using a Pressure Differential Scanning Calorimeter (PDSC),”NLGI Spokesman, Vol 65, No. 3, June 2001, pp. 1623.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.7. Calibration7.1 Sample Temperature Calibra
15、tion:7.1.1 Weigh approximately 10 mg of indium into an alumi-num sample pan, insert a lid and crimp the lid to the pan usingthe encapsulation press. Place the crimped pan onto the sampleplatform in the pressure cell. Seal an empty pan in the samemanner and place it on the reference platform. Set the
16、 cellcover in place and close the cell.7.1.2 Open the oxygen cylinder valve slightly and set apressure of 3.5 6 0.2 MPa (500 6 25 psig) on the cell inlet linewith the pressure regulator. Partially open the inlet valve on thecell and allow the pressure to slowly build up in the cell. Thisrequires app
17、roximately 2 min. Using the outlet valve, adjustand maintain the oxygen purge rate through the flowmeter at100 6 10 mL/min.7.1.3 Set the thermal analyzer to heat from ambient tem-perature (approximately 22C) to 180C at a programmed rateof 10C/min. After completion of the run measure the meltingtempe
18、rature of the indium. If the melting temperature differsfrom 157.4 6 0.2C (see Note 4) correct the difference byusing either the hardware or software calibration proceduredescribed in the manufacturers instruction manual. If thehardware calibration procedure is used, perform the tempera-ture correct
19、ion under 3.5 MPa (500 psig) oxygen pressure witha 100 mL/min purge rate. A typical melting calibration curve isshown in Fig. 2.NOTE 4The melting temperature of indium is 156.6C at atmosphericpressure, but has been found to be elevated to 157.4C under theconditions of this test method, 3.5 MPa (500
20、psig) of oxygen.37.2 Temperature Controller Calibration:7.2.1 Remove both the sample pan and the reference panfrom the cell, then close the cell. Slowly pressurize the cellwith 3.5 6 0.2 MPa (500 6 25 psig) oxygen and adjust thepurge rate to 100 6 10 mL/min using the cell outlet valve.Select the des
21、ired test temperature (either 210, 180, 155, or130C).7.2.2 Program the cell to maintain the selected test tempera-ture. If, after 10 min, the displayed cell temperature differs bymore than 6 0.2C from the selected temperature, slowlyadjust the temperature controller until they agree. After makingan
22、adjustment, wait at least 5 min to make certain that thetemperature is stable before continuing. If the PDSC equip-ment does not have this function, the control calibration shallbe followed according to the equipment manufacturers rec-ommendations.7.3 Cell Base Pressure Gage CalibrationConduct theca
23、libration using a calibrated pressure transducer or a previ-ously calibrated gage according to the pressure cell manufac-turers instructions.8. Procedure8.1 Before starting a test, the control thermocouple calibra-tion shall be conducted at the test temperature (either 210, 180,155, or 130C) accordi
24、ng to 7.2.1 and 7.2.2. When the testtemperature is not known, conduct the calibration at 210C.8.2 Weigh 3.0 6 0.2 mg of oil into a new sample pan.Spread the sample evenly upon the flat portion. Do not spillany of the sample into the trough portion of the pan. A flatbottom pan can be used if the samp
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD618619982003E1STANDARDTESTMETHODFOROXIDATIONINDUCTIONTIMEOFLUBRICATINGOILSBYPRESSUREDIFFERENTIALSCANNINGCALORIMETRYPDSC

链接地址:http://www.mydoc123.com/p-521799.html