ASTM D5128-2014 Standard Test Method for On-Line pH Measurement of Water of Low Conductivity《在线测量低电导率水pH值的标准试验方法》.pdf
《ASTM D5128-2014 Standard Test Method for On-Line pH Measurement of Water of Low Conductivity《在线测量低电导率水pH值的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5128-2014 Standard Test Method for On-Line pH Measurement of Water of Low Conductivity《在线测量低电导率水pH值的标准试验方法》.pdf(12页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D5128 14Standard Test Method forOn-Line pH Measurement of Water of Low Conductivity1This standard is issued under the fixed designation D5128; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision.
2、A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the precise on-line determina-tion of pH in water samples of conductivity lower than 100S/cm (see Table 1 and
3、 Table 2) over the pH range of 3 to 11(see Fig. 1), under field operating conditions. pH measure-ments of water of low conductivity are problematic forconventional pH electrodes, methods, and related measurementapparatus.1.2 This test method includes the procedures and equipmentrequired for the cont
4、inuous pH measurement of low conduc-tivity water sample streams including the requirements for thecontrol of sample stream pressure, flow rate, and temperature.For off-line pH measurements in low conductivity samples,refer to Test Method D5464.1.3 The values stated in SI units are to be regarded ass
5、tandard. No other units of measurement are included in thisstandard.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the a
6、pplica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD1193 Specification for Reagent WaterD1293 Test Methods for pH of WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19
7、on WaterD3864 Guide for On-Line Monitoring Systems for WaterAnalysisD4453 Practice for Handling of High Purity Water SamplesD5464 Test Method for pH Measurement of Water of LowConductivity3. Terminology3.1 DefinitionsFor definitions of other terms used in thistest method, refer to Terminology D1129
8、and Practice D3864.3.2 Definitions of Terms Specific to This Standard:3.2.1 liquid junction potential, na dc potential that ap-pears at the point of contact between the reference electrodessalt bridge (sometimes called diaphragm) and the samplesolution.3.2.1.1 DiscussionIdeally this potential is nea
9、r zero, andis stable. However, in low conductivity water it becomes largerby an unknown amount, and is a zero offset. As long as itremains stable its effect can be minimized by “grab sample”calibration (3).33.2.2 streaming potential, nthe static electrical charge thatis induced by the movement of a
10、low ionic strength solutionhaving a high electrical resistivity or low electrical conductiv-ity (such as pure water), across relatively non-conductivesurfaces such as the pH electrode or other non-conductivewetted materials found in flowing sample streams.4. Summary of Test Method4.1 pH is measured
11、by electrodes contained in an allstainless steel flow cell. The pH measurement half-cell isconstructed of a glass membrane suitable for continuousservice in low conductivity water. The reference half-cell isconstructed in such a manner that the salt bridge uses either aflowing liquid electrolyte, or
12、 a pressurized gel electrolyte thatresists significant dilution for periods up to several months ofcontinuous operation.4.2 This test method describes the apparatus and proceduresto be used for the continuous on-line pH measurement of lowconductivity water sample streams. The requirements forconditi
13、oning sample pressure and flow rate are defined, andarrangements for this associated equipment are illustrated.1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.03 on Sampling Water andWater-Formed Deposits, Analysis of Wat
14、er for Power Generation and Process Use,On-Line Water Analysis, and Surveillance of Water.Current edition approved Jan. 1, 2014. Published January 2014. Originallyapproved in 1990. Last previous edition approved in 2009 as D5128 09. DOI:10.1520/D5128-14.2For referenced ASTM standards, visit the ASTM
15、 website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The boldface numbers given in parentheses refer to a list of references at theend of this standard.Copyrig
16、ht ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States14.3 The apparatus and procedures described in this testmethod are intended to be used with process-grade, pHanalyzer/transmitter instruments.5. Significance and Use5.1 Continuous pH measurement
17、s in low conductivitysamples are sometimes required in pure water treatment usingmultiple pass reverse osmosis processes. Membrane rejectionefficiency for several contaminants depends on pH measure-ment and control between passes where the conductivity is low.5.2 Continuous pH measurement is used to
18、 monitor powerplant cycle chemistry where small amounts of ammonia oramines or both are added to minimize corrosion by hightemperature pure water and steam.5.3 Conventional pH measurements are made in solutionsthat contain relatively large amounts of acid, base, or dissolvedsalts. Under these condit
19、ions, pH determinations may be madequickly and precisely. Continuous on-line pH measurements inwater samples of low conductivity are more difficult (4, 5).These low ionic strength solutions are susceptible to contami-nation from the atmosphere, sample stream hardware, and thepH electrodes. Variation
20、s in the constituent concentration oflow conductivity waters cause liquid junction potential shifts(see 3.2.1) resulting in pH measurement errors. Special pre-cautions are required.6. Interferences6.1 Sample systems for high purity, low conductivity watersare especially sensitive to contamination fr
21、om atmosphericgases (especially carbon dioxide, see Appendix X1 and Table3), and to accumulation of power plant corrosion particles insample lines that can absorb or desorb contaminants. ExcessiveTABLE 1 Calculated Conductivity and pH Values at 25C of LowConcentrations of NaOH in Pure WaterANOTE 1Th
22、is table tabulates the theoretical conductivity and pH valuesof low levels of NaOH in pure water as calculated from availablethermodynamic data.NOTE 2To illustrate the high sensitivity of the sample pH at these lowconcentrations to contaminants, the last column lists errors that wouldresult if the s
23、ample were contaminated with an additional 1 mg/L throughsample or equipment handling errors.SampleConcentration,mg/LSampleConductivity,S/cmSamplepH pH Error from Addi-tional 1 mg/L NaOHContaminate0.001 0.055 7.05 2.350.010 0.082 7.45 1.950.100 0.625 8.40 1.031.0 6.229 9.40 0.308.0 49.830 10.30 0.05
24、AData courtesy of Ref (1).BThis data developed from algorithms originallypublished in Ref (2).BThe boldface numbers in parentheses refer to a list of references at the end ofthis standard.TABLE 2 Calculated Conductivity and pH Values at 25C of LowConcentrations of HCl in Pure WaterANOTE 1This table
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD51282014STANDARDTESTMETHODFORONLINEPHMEASUREMENTOFWATEROFLOWCONDUCTIVITY 在线 测量 电导率 PH 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-518930.html