ASTM D4492-2007 Standard Test Method for Analysis of Benzene by Gas Chromatography《气相色谱法分析苯的标准试验方法》.pdf
《ASTM D4492-2007 Standard Test Method for Analysis of Benzene by Gas Chromatography《气相色谱法分析苯的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4492-2007 Standard Test Method for Analysis of Benzene by Gas Chromatography《气相色谱法分析苯的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 4492 07Standard Test Method forAnalysis of Benzene by Gas Chromatography1This standard is issued under the fixed designation D 4492; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number
2、 in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of normallyoccurring trace impurities in, and the purity of, finishedbenzene by gas chromatograph
3、y.1.2 This test method is applicable for benzene purities of99.80 weight % or higher.1.3 The lower limit of detection for total non-aromaticimpurities is 0.005 weight %. The lower limit of detection forindividual aromatic hydrocarbon impurities in 0.001 weight %.The lower limit of detection for 1,4-
4、dioxane is 0.0005 weight%.1.4 In determining the conformance of the test results usingthis method to applicable specifications, results shall berounded off in accordance with the rounding off method ofPractice E29.1.5 The values stated in SI units are to be regarded asstandard. No other units of mea
5、surement are included in thisstandard.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory li
6、mitations prior to use. For specific hazardstatements, see Section 8.2. Referenced Documents2.1 ASTM Standards:2D 852 Test Method for Solidification Point of BenzeneD 3437 Practice for Sampling and Handling Liquid CyclicProductsD 6809 Guide for Quality Control and Quality AssuranceProcedures for Aro
7、matic Hydrocarbons and Related Ma-terialsE29 Practice for Using Significant Digits in Test Data toDetermine Conformance with SpecificationsE 260 Practice for Packed Column Gas ChromatographyE 355 Practice for Gas Chromatography Terms and Rela-tionshipsE 691 Practice for Conducting an Interlaboratory
8、 Study toDetermine the Precision of a Test MethodE 1510 Practice for Installing Fused Silica Open TubularCapillary Columns in Gas Chromatographs2.2 Other Document:OSHA Regulations, 29 CFR paragraphs 1910.1000 and1910.120033. Summary of Test Method3.1 A known amount of an internal standard is added t
9、o thespecimen. A small volume of this mixture is injected into a gaschromatograph equipped with a flame ionization detector (FID)and a capillary column.3.2 The peak area of each impurity and the internal standardis measured by an electronic integrator. The concentration ofeach impurity is calculated
10、 from the ratio of the peak area ofthe internal standard versus the peak area of the impurity.Purity is calculated by subtracting the sum of the impuritiesfound from 100.00 weight %. Results are reported in weightpercent.4. Significance and Use4.1 This test method is suitable for determining the con
11、cen-trations of known impurities in finished benzene and for use asan integral quality control tool where benzene is either pro-duced or used in a manufacturing procedure. It is generallyapplied to impurities such as nonaromatics containing ninecarbons or less, toluene, C8 aromatics, and 1,4-dioxane
12、.4.2 Absolute purity cannot be determined if unknown im-purities are present. Test Method D 852 is generally used as acriteria for determining the absolute purity.5. Interferences5.1 Benzene is typically resolved from naturally occurringcomponents with boiling points 138C. Naturally occurringcompone
13、nts include nonaromatic hydrocarbons, toluene, C81This test method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and Related Chemicals and is the direct responsibility ofSubcommittee D16.01 on Benzene, Toluene, Xylenes, Cyclohexane and TheirDerivatives.Current edition appro
14、ved June 1, 2007. Published June 2007. Originallyapproved in 1985. Last previous edition approved in 2003 as D 4492 - 03.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refe
15、r to the standards Document Summary page onthe ASTM website.3Available from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:/www.access.gpo.gov.1*A Summary of Changes section appears at the end of this standard.Copyright
16、ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.aromatics, and 1,4-dioxane. An adequate separation of knownimpurities from benzene should be evaluated for the columnselected.5.2 The internal standard chosen must be sufficiently re-solved from a
17、ny impurity and the benzene peak.6. Apparatus6.1 Gas ChromatographAny chromatograph having aflame ionization detector that can be operated at the conditionsgiven in Table 1. The system should have sufficient sensitivityto obtain a minimum peak height response for a 0.0005 weight% impurity three time
18、s the height of the signal backgroundnoise.6.2 Electronic Integrator, computer-based capable of han-dling internal standard calculations and peak grouping isrecommended.6.3 Column, fused silica capillary column with cross-linkedpolyethylene glycol stationary phase is recommended. Alter-nate stationa
19、ry phases may be used if they produce at least thesame aromatic separation and elute C9nonaromatic impuritiesbefore benzene.6.4 Microsyringes, 10 and 100 l capacity.7. Reagents and Materials7.1 Carrier GasChromatographic grade helium is recom-mended.7.2 High Purity Benzene, 99.99 weight % minimum, p
20、re-pared by multiple step recrystallization of commercially avail-able 99 + weight % benzene.7.3 Internal Standard, n-Nonane (nC9) with a purity of 99weight % minimum is recommended. Other compounds maybe acceptable provided they can be obtained in high purity andmeet the requirements of 5.2.7.4 Pur
21、e compounds for calibration should include toluene,benzene, ethylbenzene, cyclohexane, and 1,4-dioxane of apurity not less than 99 %. If the purity of the calibrationcompounds is less than 99 %, the concentration and identifi-cation of impurities must be known so that the composition ofthe final wei
22、ghed blends can be adjusted for the presence of theimpurities.8. Hazards8.1 Consult current OSHA regulations and suppliers Mate-rial Safety Data Sheets and local regulations for all materialsused in this test method.8.2 Benzene is considered a hazardous material. The sam-pling and testing of benzene
23、 should follow safety rules in orderto adhere to all safety precautions as outlined in current OSHAregulations.9. Sampling9.1 Sample the material in accordance with PracticeD 3437.10. Preparation of Apparatus10.1 Follow manufacturers instructions for mounting thecolumn into the chromatograph and adj
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD44922007STANDARDTESTMETHODFORANALYSISOFBENZENEBYGASCHROMATOGRAPHY 色谱 分析 标准 试验 方法 PDF

链接地址:http://www.mydoc123.com/p-517212.html