1、Designation: D3861 91 (Reapproved 2003)Standard Test Method forQuantity of Water-Extractable Matter in Membrane Filters1This standard is issued under the fixed designation D3861; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the
2、year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the gravimetric determinationof the water-extractable material present in membrane fil
3、tersand is applicable over the complete concentration range ofextractables.1.2 The analyst should be aware that adequate collaborativedata for precision and bias statements as required by PracticeD2777 is not provided. See Section 11 for details.1.3 This standard does not purport to address all of t
4、hesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating
5、 to WaterD1193 Specification for Reagent WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on Water3. Terminology3.1 DefinitionsFor definitions of terms used in this testmethod, refer to Terminology D1129.4. Summary of Test Method4.1 The quantity
6、of water-soluble extractables present inmembrane filters is determined by immersing the preweighedmembrane in boiling reagent grade water for an extended time,then drying and reweighing the membrane. A control mem-brane is employed to eliminate weighing errors caused bybalance changes or changing mo
7、isture content of the mem-brane in the weighing procedures. Weight changes of thecontrol membrane are applied as a correction factor to theweight change of the test membrane filters.5. Significance and Use5.1 The presence of water-soluble extractables in membranefilters can create errors in test pro
8、cedures employing membranefilters. However, these errors can be eliminated or significantlyreduced if the quantity of water-soluble extractables of thespecific membrane is previously determined. Certain mem-brane filter uses require specifications of maximum water-soluble extractable levels. This te
9、st method is intended to be arapid test to determine the loss of water-soluble compoundssuch as plasticizers or wetting agents from filtration mem-branes. This test method is not designed to predetermine theperformance of a filter, but is significant in determining thepercent extractables of membran
10、es from different sources andlot variations from a single source.6. Apparatus6.1 Beaker, borosilicate glass, 300-mL capacity.6.2 Forceps, stainless steel, unserrated tips.6.3 Desiccator.6.4 Drying Oven, with thermostatic control.6.5 Analytical Balance, sensitivity 0.01 mg.6.6 a-Emitting Polonium Sou
11、rce3(to discharge staticcharge in balance).6.7 Humidity-Controlled Room or Hood, for balance.6.8 Evaporating Dishes, glass, having sides or other meansto keep test filters on the dish while drying.7. Reagents and Materials7.1 Purity of Water Unless otherwise indicated, refer-ences to water shall be
12、understood to mean Type III reagentwater conforming to Specification D1193.7.2 Membrane Filters, 47-mm diameter (same type as thoseunder test).8. Preparation of Apparatus and Materials8.1 Preparation of Samples and Apparatus:8.1.1 Select 47-mm diameter sample (Note 1) membranefilters for the test, a
13、nd label with water-resistant ink.NOTE 1Square or rectangular membrane of equivalent area (17.4 cm2) may be used.1This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibility of Subcommittee D19.08 on Membranes and IonExchange Materials.Current edition ap
14、proved June 15, 1991. Published December 1991. Originallyapproved in 1979. Last previous edition approved in 1991 as D3861 91. DOI:10.1520/D3861-91R09.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStan
15、dards volume information, refer to the standards Document Summary page onthe ASTM website.3The “Staticmaster,” available from Nuclear Products Inc., El Monte, CA, hasbeen found suitable for this purpose.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-29
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD386119912011STANDARDTESTMETHODFORQUANTITYOFWATEREXTRACTABLEMATTERINMEMBRANEFILTERS 滤器 中水 萃取 物质 数量

链接地址:http://www.mydoc123.com/p-515660.html