ASTM E986-04(2017) Standard Practice for Scanning Electron Microscope Beam Size Characterization.pdf
《ASTM E986-04(2017) Standard Practice for Scanning Electron Microscope Beam Size Characterization.pdf》由会员分享,可在线阅读,更多相关《ASTM E986-04(2017) Standard Practice for Scanning Electron Microscope Beam Size Characterization.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E986 04 (Reapproved 2017)Standard Practice forScanning Electron Microscope Beam Size Characterization1This standard is issued under the fixed designation E986; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year o
2、f last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides a reproducible means by whichone aspect of the performance of a scanning electron micro-scope
3、 (SEM) may be characterized. The resolution of an SEMdepends on many factors, some of which are electron beamvoltage and current, lens aberrations, contrast in the specimen,and operator-instrument-material interaction. However, theresolution for any set of conditions is limited by the size of theele
4、ctron beam. This size can be quantified through the mea-surement of an effective apparent edge sharpness for a numberof materials, two of which are suggested. This practice requiresan SEM with the capability to perform line-scan traces, forexample, Y-deflection waveform generation, for the suggested
5、materials. The range of SEM magnification at which thispractice is of utility is from 1000 to 50 000 . Highermagnifications may be attempted, but difficulty in makingprecise measurements can be expected.1.2 This standard does not purport to address all of thesafety concerns, if any, associated with
6、its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.1.3 This international standard was developed in accor-dance with internationally recognized principles on stand
7、ard-ization established in the Decision on Principles for theDevelopment of International Standards, Guides and Recom-mendations issued by the World Trade Organization TechnicalBarriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2E7 Terminology Relating to MetallographyE766 P
8、ractice for Calibrating the Magnification of a Scan-ning Electron Microscope3. Terminology3.1 Definitions: For definitions of terms used in thispractice, see Terminology E7.3.2 Definitions of Terms Specific to This Standard:3.2.1 Y-deflection waveformthe trace on a CRT resultingfrom modulating the C
9、RT with the output of the electrondetector. Contrast in the electron signal is displayed as achange in Y (vertical) rather than brightness on the screen. Thisoperating method is often called Y-modulation.4. Significance and Use4.1 The traditional resolution test of the SEM requires, as afirst step,
10、a photomicrograph of a fine particulate sample takenat a high magnification. The operator is required to measure adistance on the photomicrograph between two adjacent, butseparate edges. These edges are usually less than one millime-tre apart. Their image quality is often less than optimumlimited by
11、 the S/N ratio of a beam with such a small diameterand low current. Operator judgment is dependent on theindividual acuity of the person making the measurement andcan vary significantly.4.2 Use of this practice results in SEM electron beam sizecharacterization which is significantly more reproducibl
12、e thanthe traditional resolution test using a fine particulate sample.5. Suggested Materials5.1 SEM resolution performance as measured using theprocedure specified in this practice will depend on the materialused; hence, only comparisons using the same material havemeaning. There are a number of cri
13、teria for a suitable materialto be used in this practice. Through an evaluation of thesecriteria, two samples have been suggested. These samples arenonmagnetic; no surface preparation or coating is required;thus, the samples have long-term structural stability. Thesample-electron beam interaction sh
14、ould produce a sharplyrising signal without inflections as the beam scans across theedge. Two such samples are:5.1.1 Carbon fibers, NISTSRM 2069B.35.1.2 Fracture edge of a thin silicon wafer, cleaved on a(111) plane.1This practice is under the jurisdiction of ASTM Committee E04 on Metallog-raphy and
15、 is the direct responsibility of Subcommittee E04.11 on X-Ray andElectron Metallography.Current edition approved June 1, 2017. Published June 2017. Originallyapproved in 1984. Last previous edition approved in 2010 as E986 04(2010). DOI:10.1520/E0986-04R17.2For referenced ASTM standards, visit the A
16、STM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from National Institute of Standards and Technology (NIST), 100Bureau Dr., Stop 1070, Gaither
17、sburg, MD 20899-1070, http:/www.nist.gov.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United StatesThis international standard was developed in accordance with internationally recognized principles on standardization established in the Decision
18、on Principles for theDevelopment of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.16. Procedure6.1 Inspect the specimen for cleanliness. If the specimenappears contaminated, a new sample is recommended as anycle
19、aning may adversely affect the quality of the specimen edge.6.2 Ensure good electrical contact with the specimen byusing a conductive cement to hold the specimen on a SEMstub, or by clamping the specimen on the stage of the SEM.Mount the specimen rigidly in the SEM to minimize any imagedegradation c
20、aused by vibration.6.3 Verify magnification calibration for both X and Y direc-tions. This can be accomplished by using Practice E766.6.4 Use a clean vacuum of 1.33 by 102Pa (104mm Hg)or better to minimize specimen contamination resulting fromelectron beam and residual hydrocarbons interacting durin
21、gexamination. The presence of a contamination layer has adeleterious effect on image-edge quality.6.5 Allow a minimum of 30 min for stabilization of elec-tronic components, vacuum stability, and thermal equilibriumfor the electron gun and lenses. The selection of optimum SEMparameters is at the disc
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTM E986 04 2017 Standard Practice for Scanning Electron Microscope Beam Size Characterization

链接地址:http://www.mydoc123.com/p-287144.html