UOP 988-2011 Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion with Ultraviolet Fluorescence Detection.pdf
《UOP 988-2011 Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion with Ultraviolet Fluorescence Detection.pdf》由会员分享,可在线阅读,更多相关《UOP 988-2011 Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion with Ultraviolet Fluorescence Detection.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2011 UOP LLC. All rights reserved. Non
3、confidential UOP Methods are available from ASTM International, 100 Barr Harbor Drive, P.O. Box C700, West Conshohocken, PA 19428-2959, USA. The UOP Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585
4、 PHONE. Low Trace Sulfur in LPG and Gaseous Hydrocarbons by Oxidative Combustion with Ultraviolet Fluorescence Detection UOP Method 988-11 Scope This method is for determining sulfur in liquefied petroleum gas (LPG) and gaseous hydrocarbons at concentrations ranging from 10 to 1400 ng/g (0.01 to 1.4
5、0 mass-ppm). A trap approximately twice that amount for a 3.2-L bag). For gas samples, fill until the bag expands, but is not completely full. 4. Close the valve on the gas sampling bag and remove from the cylinder. 5. Analyze the sample within 3 hours of transferring to the gas sampling bag. Prepar
6、ation of Apparatus 1. Set up the instrument according to the manufacturers instructions for the trap and release unit. Connect the membrane drier in series between the combustion tube and the detector. Allow the instrument to warm up and the baseline to stabilize before injecting samples. Suggested
7、Operating Conditions for the Mitsubishi TS-100V analyzer are listed in Table 1. The same trap time must be used for the calibration and the samples use the longest trap time if calibrating for multiple ranges. A 25-mL syringe is installed in the GI-220 gas injector. Table 1 Operating Conditions for
8、Mitsubishi TS-100V/SD-100/TRU-100 Upper temperaturea 900C Lower temperatureb 1000C Argon main 170 mL/min Oxygen main 150 mL/min Argon auxiliary 50 mL/min Oxygen auxiliary 400 mL/min GI-220 carrier argon 80 mL/min Gain Ultra Normal end Off Minimum area 40000 Base line 75% GI-220 Syringe 25 mL Absorpt
9、ion rate 50 mL/min Injection rate 20 mL/min aCombustion tube, upper portion bCombustion tube, lower portion Table 2 Recommended Sample Size and Instrument Timing Concentration Range (ng/g) Sample Volume (mL) Trap Time (s) Timer Start (s) Timer End (s) 10 - 200 80 532 535 835 200 - 500 40 296 310 610
10、 500 - 1400 10 250 270 570 Calibration Calibrate weekly when in use. Check the calibration daily when in use, by analyzing one of the calibration standards or a reference material. 1. Transfer the calibration standard to a gas sampling bag, and then connect to the GI-220 gas injector module. The sam
11、e gas sampling bag can be re-used for the standard. Empty the bag before refilling. 5 of 8 988-11 2. Set up the sample table for multiple sample volumes and replicate injections of the sulfur standard. Three or four injections are recommended for volumes below 10 mL and two injections for volumes ab
12、ove 10 mL. Use the trap time and timer start/end from Table 2. If calibrating for multiple ranges, use the trap and integration times for the longest method that will be used for all standards and samples. For a low level calibration, use 1-, 2-, 3-, 5-, and 10-mL volumes. For a mid or high level ca
13、libration, use 5-, 10-, 20-, and 30-mL volumes. Do not exceed 30 mL of the standard, or the trap may be saturated. See Note 4. If run as a “calibration” method, the “START” button will need to be pressed for each new sample volume. If run as a “sample” method, the instrument will run all measurement
14、s on the same material automatically, and the S content and response areas can be entered in manually. See Note 5. Relative standard deviation (RSD), as calculated by the instrument for the standards, should be within 15% for the 1- or 2-mL injections and within 10% for 3-mL or larger injections. 3.
15、 Analyze the standard according to the instrument manufacturers instructions. 4. Create a regression line using the instrument software using the appropriate volumes of the standards. Set the regression line to “y=bx+c” for this calibration. Sample Analysis 1. Determine the average molecular weight
16、of the sample, see the Appendix. 2. Attach the gas sampling bag with the first sample to the GI-220 gas injector module. 3. Add the sample to the sample table. If unsure of concentration, start with a small volume such as 5 mL; otherwise, select the volume based on Table 2. The amount of sulfur inje
17、cted should not exceed 50 ng per measurement. Analyze samples above 100 ng/g in duplicate. For samples below 100 ng/g, 3 or more measurements should be made. Enter the sample average molecular weight to calculate the results as mass-ppm. 4. Analyze the sample replicates according to the instrument m
18、anufacturers instructions. 5. Repeat Steps 1 through 4 for each additional sample. If the detector becomes contaminated (trace off scale), disconnect the gas sampling bag and analyze air blanks until the response stabilizes. Then, confirm that the sensitivity has not changed by analyzing a calibrati
19、on standard or a control standard. Integration and calculations are done automatically. The average of the replicate injections is calculated by the instrument software. Calculations All calculations are performed by the software, and results are displayed and printed in mass-ppm (mg/kg) or mass-ppb
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9882011LOWTRACESULFURINLPGANDGASEOUSHYDROCARBONSBYOXIDATIVECOMBUSTIONWITHULTRAVIOLETFLUORESCENCEDETECTIONPDF

链接地址:http://www.mydoc123.com/p-1074603.html