UOP 983-2004 ACETYLENES IN 1 3-BUTADIENE BY GC.pdf
《UOP 983-2004 ACETYLENES IN 1 3-BUTADIENE BY GC.pdf》由会员分享,可在线阅读,更多相关《UOP 983-2004 ACETYLENES IN 1 3-BUTADIENE BY GC.pdf(9页珍藏版)》请在麦多课文档分享上搜索。
1、 IT IS THE USERS RESPONSIBILITY TO ESTABLISH APPROPRIATE PRECAUTIONARY PRACTICES AND TO DETERMINE THE APPLICABILITY OF REGULATORY LIMITATIONS PRIOR TO USE. EFFECTIVE HEALTH AND SAFETY PRACTICES ARE TO BE FOLLOWED WHEN UTILIZING THIS PROCEDURE. FAILURE TO UTILIZE THIS PROCEDURE IN THE MANNER PRESCRIB
2、ED HEREIN CAN BE HAZARDOUS. MATERIAL SAFETY DATA SHEETS (MSDS) OR EXPERIMENTAL MATERIAL SAFETY DATA SHEETS (EMSDS) FOR ALL OF THE MATERIALS USED IN THIS PROCEDURE SHOULD BE REVIEWED FOR SELECTION OF THE APPROPRIATE PERSONAL PROTECTION EQUIPMENT (PPE). COPYRIGHT 2004 UOP LLC ALL RIGHTS RESERVED uop,
3、UOP, and KLP are trademarks and/or service marks of UOP LLC. Marks of other proprietors may appear incidentally in this method for purposes such as product or service identification, but no claim is made to any other proprietors mark used. UOP Methods are available through ASTM International, 100 Ba
4、rr Harbor Drive, PO Box C700, West Conshohocken, PA 194428-2959, United States. The Methods may be obtained through the ASTM website, www.astm.org, or by contacting Customer Service at serviceastm.org, 610.832.9555 FAX, or 610.832.9585 PHONE. ACETYLENES IN 1,3-BUTADIENE BY GC uop Method 983-04 SCOPE
5、 This method is for determining low percent levels of acetylenes in crude butadiene feedstock and product, and is specifically applicable to KLP process streams. The following components and group classifications may be reported: methane, ethane, ethylene, propane, cyclopropane, propylene, isobutane
6、, acetylene, n-butane, propadiene, trans-2-butene, 1-butene, isobutylene, cis-2-butene, isopentane, n-pentane, 1,2-butadiene, methyl acetylene, 1,3-butadiene, 3-methyl-1-butene, trans-2-pentene, 2-methyl-1-butene, vinyl acetylene, ethyl acetylene, 2-methyl-2-butene, and total heavies. Components tha
7、t do not elute from the chromatographic system under the specified conditions are not determined. The range of quantitation for any compound is 0.01 to 99.9 mass-%. The analysis is applicable to laboratory or on-line use. REFERENCES ASTM Practice D 1265, “Sampling Liquefied Petroleum (LP) Gases (Man
8、ual Method),” www.astm.org Scanlon, J. T. and Willis, D. E., Journal of Chromatographic Science, 23, 333-340 (1985) UOP Method 516, “Sampling and Handling of Gasolines, Distillate Fuels, and C3-C4Fractions,” www.astm.org UOP Method 999, “Precision Statements in UOP Methods,” www.astm.org OUTLINE OF
9、METHOD The sample to be analyzed is injected as a liquefied petroleum gas (LPG) into a gas chromatograph equipped with an alumina PLOT (porous layer open tubular) capillary column, a capillary injection port, a flame ionization detector (FID), an LPG sampling valve, and a backflush valve. The PLOT c
10、olumn elutes individually measured components through 2-methyl-2-butene, after which the remainder of the sample is back-flushed to elute any remaining heavy materials. The composition of 2 of 9 983-04 the sample is obtained by the normalization technique, wherein the peak areas of the entire sample
11、 are first corrected for differences in response and then normalized to 100%. Because of the toxicity of 1,3-butadiene, follow safe venting and environmental practices. APPARATUS References to catalog numbers and suppliers are included as a convenience to the method user. Other suppliers may be used
12、 if equivalent performance can be obtained. Fittings and connectors are listed to connect to the specified apparatus. Additional fittings will be required for the specific sample cylinders, for instrument gases, for on-line connections, and to conform to good laboratory procedures and local safety r
13、egulations. Chromatographic column, 50 m of 0.32-mm ID fused silica capillary, internally coated to a film thickness of 5-m with potassium chloride washed aluminum oxide (Al2O3/KCl), Chrompack Cat. No. 7515, Varian Inc. Fittings, bulkhead, 1.6-mm (1/16-inch) zero dead volume union with internal bore
14、 of 0.25 mm used as connection between fused silica and stainless steel tubing. Valco Instruments, Cat. No. ZBU1C. See Figure 1. Gas chromatograph, temperature programmable, capable of constant flow, built for capillary column chromatography, utilizing a split injection system, having a glass inject
15、ion port insert, a heated valve box with valve actuators, and equipped with an FID that will give a minimum peak height response of five times the background noise for 0.01 mass-% of 1-butene when operated at the recommended conditions, Agilent Technologies, Model 6890 Injection apparatus, LPG, for
16、injection of LPG samples in the liquid phase. See Figure 1. Fitting, internal union, for connecting injection port to 4-port sampling valve, Valco Instruments, Cat. No. ZU1T Fitting, reducing union, 3.2-mm (1/8-inch) to 1.6-mm (1/16-inch), Swagelok, Cat. No. SS-200-6-1 Tubing, stainless steel, for c
17、onnecting the sample cylinder to the 4-port sampling valve, 1.6-mm (1/16-inch) OD, 0.76-mm (0.30-inch) ID, Alltech Associates, Cat. No. 300010 Tubing, translucent, for connecting the 4-port sampling valve to the vent shut-off valve, FEP Teflon, 3.2-mm (1/8-inch) OD, 1.6-mm (0.062-inch) ID, 3450 kPa
18、(500 psig), Alltech Associates, Cat. No. 45740 Valve, sample injection, 4-port, stainless steel, 0.5-L internal sample volume, 1.6-mm (1/16-inch) fittings, Valco Instruments, Cat. No. CI4UWE.5, with air actuator, Valco Instruments, Cat. No. A-904, controlled by GC Valve, vent shut-off, Swagelok, Cat
19、. No. SS-ORS2 Integrator, or data system, electronic, for obtaining peak areas. This device must integrate areas at a sufficiently fast rate so that narrow peaks typically obtained from a capillary column can be accurately measured. The integrator must have programmable parameters for controlling ba
20、seline events, and have graphic capabilities. ChemStation, Agilent Technologies 3 of 9 983-04 Leak detector, gas, Alltech Associates, Cat. No. 21-250 Regulator, air, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-590, two required Regulator, heli
21、um, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Regulator, hydrogen, two-stage, high purity, delivery pressure range 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-350 Regulator, nitrogen, two-stage, high purity, delivery pressure ra
22、nge 30-700 kPa (4-100 psi), Matheson Tri-Gas, Model 3122-580 Tubing, fused silica, 5-m x 0.32-mm ID or smaller, deactivated fused silica tubing. If deactivated tubing is not available, a thin film poly-dimethyl siloxane column may be substituted. Tubing, stainless steel, 6-m (20-ft) x 1.6-mm (1/16-i
23、nch) OD, with internal diameter 0.5-mm (0.02-inch) or smaller, used as connection between bulkhead unions and backflush valve, and between GC and remote injection valve. Valve, backflush, Valco Instruments, Cat. No. C4UWE REAGENTS AND MATERIALS References to catalog numbers and suppliers are include
24、d as a convenience to the method user. Other suppliers may be used. Air, for valve actuation Air, zero gas, total hydrocarbons less than 2.0 ppm as methane Gas purifier, helium, Mat/Sen, Cat. No. P100-1 Helium, zero gas, total hydrocarbons less than 0.5 ppm as methane Hydrogen, zero gas, 99.95% mini
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- UOP9832004ACETYLENESIN13BUTADIENEBYGCPDF