UL SUBJECT 1363A-2010 OUTLINE OF INVESTIGATION FOR SPECIAL PURPOSE RELOCATABLE POWER TAPS (Issue 3)《专用重新配置的电源接头的调查大纲 (议题3)》.pdf
《UL SUBJECT 1363A-2010 OUTLINE OF INVESTIGATION FOR SPECIAL PURPOSE RELOCATABLE POWER TAPS (Issue 3)《专用重新配置的电源接头的调查大纲 (议题3)》.pdf》由会员分享,可在线阅读,更多相关《UL SUBJECT 1363A-2010 OUTLINE OF INVESTIGATION FOR SPECIAL PURPOSE RELOCATABLE POWER TAPS (Issue 3)《专用重新配置的电源接头的调查大纲 (议题3)》.pdf(10页珍藏版)》请在麦多课文档分享上搜索。
1、UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULJanuary 8, 20101SUBJECT 1363AOUTLINE OF INVESTIGATIONFORSPECIAL PURPOSE RELOCATABLE POWER TAPSIssue Number: 3January 8, 2010Summary of TopicsThis is the third issue of the Outline of Investigatio
2、n for Special PurposeRelocatable Power Taps, which covers Special Purpose RelocatablePower Taps that are intended for use with medical equipment. Medicalequipment is typically intended for use in General Patient Care Area orCritical Patient Care Area.The UL Foreword is no longer located within the U
3、L Outline of Investigation. Forinformation concerning the use and application of the requirements contained inthis Outline of Investigation, the current version of the UL Foreword is located onULStandardsInfoNet at: http:/ 2010 UNDERWRITERS LABORATORIES INC.UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR
4、 FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULJANUARY 8, 2010SUBJECT 1363A2No Text on This PageUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULCONTENTSINTRODUCTION1 Scope .42 General 42.1 Components .42.2 Units of measurement
5、.52.3 Undated references .53 Definitions 5CONSTRUCTION4 General 65 Enclosure 76 Protective Devices 7PERFORMANCE7 General 78 Limitation of Voltage and/or Energy .89 Protective Earthing, Functional Earthing and Potential Equalization .810 Leakage Current Test 811 Separation and Dielectric Voltage With
6、stand Test 8RATINGS12 Details 9MARKINGS13 Details .9INSTRUCTIONS14 Details .10JANUARY 8, 2010 SUBJECT 1363A 3UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULINTRODUCTION1 Scope1.1 These requirements must comply with all the requirements in the
7、 Standard for Relocatable PowerTaps, UL 1363 except as described in this outline.1.2 These requirements cover Special Purpose Relocatable Power Taps (SPRPT), rated 250 V AC orless, intended for indoor use only, with medical equipment, where the medical equipment is intended tobe used in General Pati
8、ent Care Areas or Critical Patient Care Areas as defined by Article 517 of theNational Electrical Code for Health Care Facilities, to supply power to plug-connected components of amovable equipment assemblies that is rack-, table-, or pedestal-mounted. The shall be an integral part ofthe equipment a
9、ssembly, permanently attached. The sum of the ampacity of all appliances connected tothe receptacles of the SPRPT shall not exceed 80 percent of the ampacity of the flexible cord supplyingthe power to the SPRPT receptacles. The ampacity of the SPRPT flexible cord shall be in accordance withthe Natio
10、nal Electrical Code, NFPA 70. The electrical and mechanical integrity of the assembly shall beregularly verified and documented through an ongoing maintenance program. These SPRPT do notinclude devices that incorporate isolating or any other types of transformers.1.3 They consist of a NEMA type atta
11、chment plug and a length of flexible cord terminated in an enclosurein which are mounted two or more receptacle outlets. A Special Purpose Relocatable Power Tap shall beprovided with Hospital Grade attachment plugs and Hospital Grade outlets (receptacles) and may also beprovided with a detachable po
12、wer supply cord and inlet fitting. A Special Purpose Relocatable Power Tapmay be provided with supplementary overcurrent protection, switches, and indicator lights singly or in anycombination.1.4 These SPRPT are rated 250 V AC or less and 20 AC or less. SPRPT are intended for indoor useonly to suppl
13、y power by flexible cord.2 General2.1 Components2.1.1 Except as indicated in 2.1.2, a component of a product covered by this outline shall comply with therequired for that component.2.1.2 A component is not required to comply with a specific requirement that:a) Involves a feature or characteristic n
14、ot required in the application of the component in theproduct covered by this outline, orb) Is superseded by a requirement in this outline.2.1.3 A component shall be used in accordance with its rating established for the intended conditions ofuse.JANUARY 8, 2010SUBJECT 1363A4UL COPYRIGHTED MATERIAL
15、NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM UL2.1.4 Specific components are incomplete in construction features or restricted in performancecapabilities. Such components are intended for use only under limited conditions, such as certaintemperatures not exceeding s
16、pecified limits, and shall be used only under those specific conditions.2.1.5 Capacitors shall comply with the applicable requirements in Connections of Capacitors, subclause56.4 of the Standard for Medical Electrical Equipment, Part 1: General Requirements for Safety, UL60601-1.2.2 Units of measure
17、ment2.2.1 Values stated without parentheses are the requirement. Values in parentheses are explanatory orapproximate information.2.2.2 Unless otherwise indicated, all voltage and current values mentioned in this outline areroot-mean-square (rms).2.3 Undated references2.3.1 Any undated reference to a
18、 code or standard appearing in the requirements of this outline shall beinterpreted as referring to the latest edition of that code or standard.3 Definitions3.1 For the purpose of this outline, the following definitions apply. Where not defined by Subject 1363A,the following terms as well as the glo
19、ssary terms found in the Standard for Relocatable Power Taps, UL1363, and the Standard for Medical Electrical Equipment, Part 1: General Requirements for Safety, UL60601-1, apply to this outline.3.1.1 BUILT IN POWER SUPPLY A Special Purpose Relocatable Power Tap may employ a suitablemedical grade po
20、wer supply for a dedicated output isolated low voltage, DC power.3.2 SPECIAL PURPOSE RELOCATABLE POWER TAP (SPRPT) A Special Purpose RelocatablePower Tap consists of a NEMA type attachment plug and a length of flexible cord terminated in anenclosure in which two or more receptacle outlets are mounte
21、d. A SPRPT shall be provided withHospital Grade attachment plugs and Hospital Grade outlets (receptacles) and may also be providedwith a detachable power supply cord and inlet fittings. A SPRPT may be provided with supplementaryovercurrent protection, switches, and indicator lights singly or in any
22、combination.3.3 SUPPLY MAINS Permanently installed power source which may also be used to supplyelectrical apparatus that is outside the scope of this outline.JANUARY 8, 2010 SUBJECT 1363A 5UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULCONST
23、RUCTION4 General4.1 In addition to the applicable construction requirements in Standard for Relocatable Power Taps, UL1363, a SPRPT shall comply with the construction requirements in Sections46ofthis outline.4.2 A SPRPT employing a non-detachable power supply cord shall be rated 12 A or less.4.3 A S
24、PRPT may be provided with non-NEMA type receptacle outlets.4.4 A SPRPT is required to be provided with only Hospital Grade type plugs and receptacle outlets.4.5 A detachable power supply cord employed in a SPRTP is not required to have a Hospital Grade orNEMA type cord connector.4.6 A SPRPT shall co
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ULSUBJECT1363A2010OUTLINEOFINVESTIGATIONFORSPECIALPURPOSERELOCATABLEPOWERTAPSISSUE3 专用 重新 配置 电源 接头 调查

链接地址:http://www.mydoc123.com/p-1069161.html