1、c EIA ENGINEERING BULLETIN X-Ray Fluorescence for Measuring Plating Thickness CB13 September 1990 ELECTRONIC INDUSTRIES ASSOCIATION ENGINEERING DEPARTMENT i EIA CE13 90 3234b00 0000153 3 I NOTXCE i ELA Engineering Standards and Publications are designed to serve the public interest through eliminati
2、ng misunderstandings between manufacturers and purchasers, facilitating interchangeability and improvement of products, and assisting the purchaser in selecting and obtaining with minimum delay the proper product for his particular need. Existence of such Standards and Publications shall not in any
3、respect preclude any member or nonmember of EIA from manufacturing or selling products not conforming to such Standards and Publications, nor shall the existence of such Standards and Publications preclude their voluntary use by those other than EIA members, whether the standard is to be used either
4、 domestically or internationally. c Recommended Standards and Publications are adopted by EIA without regard to whether their adoption may involve patents on articles, materials, or processes. By such action, EU does not assume any liability to any patent owner, nor does it assume any obligation wha
5、tever to parties adopting the Recommended Standard or Publication. Technical Bulletins are distinguished from EIA Recommended Standards or Interim Standards, in that they contain a compilation of engineering data or information useful to the technical community, and represent approaches to good engi
6、neering practices that are suggested by the formulating committee. This Bulletin is not intended to preclude or discourage other approaches that similarly represent good engineering practice, or that may be acceptable to, or have been accepted by, appropriate bodies. Parties who wish to bring other
7、approaches to the attention of the formulating committee to be considered for inclusion in future revisions of this Bulletin are encouraged to do so. It is the intention of the formulating committee to revise and update practice, or government regulations, or for other appropriate reasons. this Bull
8、etin from time to time as may be occasioned by changes in technology, industry Published by ELIECIRONIC INDUSTRIES ASSOCIATION Engineering Department 2001 Pennsylvania Ave., N.W. Washington, D.C. 20006 PRICE Please refer to the current Catalog of ELA Le., measure the same product at various count ti
9、mes and evaluate the results for precision. Many of the units available today are equipped with software programs that will predict the percent uncertainty of a measurement at various count times. 7. DENSITY As with all mass-per-unit-area techniques, the x-ray measurement is based on a conversion of
10、 mass to thickness. Therefore, the density of the plate must be considered. For many electro-deposited coatings, such as nickel, the density is fairly straightforward and the value employed is consistent throughout the industry. Gold, however, is a different story with many densities reported. MZtar
11、y specifications for hard gold specify, by grade and type, acceptable purity and hardness - but not density. Densities ranging from l5.3 to 19.3 G/CC have been reported for gold, with hard gold electrodeposits at the lower end of the scale. The densiy of the plate will vary with bath chemistry and c
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ECACB131990XRAYFLUORESCENCEFORMEASURINGPLATINGTHICKNESS 镀层 厚度 测定 射线 荧光 PDF

链接地址:http://www.mydoc123.com/p-704138.html