DOD A-A-54830-1993 TABLE AUTOPSY《尸体解剖药片》.pdf
《DOD A-A-54830-1993 TABLE AUTOPSY《尸体解剖药片》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-54830-1993 TABLE AUTOPSY《尸体解剖药片》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、A-A-54830 9999974 OL422bb 009 a e l I I I I I l I I l l I 1 INCH-POUND 1 A-A-54 830 19 March 1993 COMMERCIAL ITEM DESCRIPTION TABLE, AUTOPSY The General Services Administration has authorized the use of this Commercial Item Description This Commercial Item Description covers an autopsy table that sh
2、all provide a stable working surface for postmortem examinations. Salient Characteristics: The autopsy table shall be constructed of stainless steel or corrosion-resistant material that has smooth edges and surfaces that are free of screws or rivets, as well as finished welds that are smooth and fre
3、e of crevices, to facilitate cleaning and prevent the accumulation of micro- organisms. The autopsy table shall corisist of the following: a. b. C. d. e. f. g- h. i. 5 One removable slat for support of the cadavers head. A pedestal type table with a trough-shaped surface that slopes toward the sink
4、for rapid and positive drainage. A one-gallon aspirator bottle. Waterproof electrical outlets. A perimeter flushing system. A disposal unit with a 1/2 or 3/4 hp motor. One aspirator that includes a vacuum breaker with a variable control valve to adjust the amount of suction. A minimum 8 feet long su
5、ction hose and nozzle shall be provided. elbow-operated valves for hot and cold water and a minimum 8 feet long spray hose and nozzle. One irrigator unit that includes a vacuum breaker with One knee-operated waste control valve. One rubber headrest. AMSC N/A FSC 6530 DISTRIBUTION STATEMENT A: Approv
6、ed for public release; distribution is unlimited. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5q830 9999974 OLY2267 T45 = A-A-54 8 3 O k. One dissecting board 12 inches plus or minus 1/2 inches square and 3/4 inch plus or minus 1/8 inch thick
7、 with the cutting surface covered in neoprene cork. capacity, approximately 10 inches high and 12 inches in diameter. Included shall be a grid plate, supported by the bottom of the container, that facilitates thorough rinsing. Overflow holes shall be provided around the entire top periphery of the c
8、ontainer. A minimum of 8 feet of transparent tubing and a locking cover shall also be included. 1. One organ rinse container of at least 4 gallons m. Five removable slats for support of the body. n. One dissecting tray, 25 inches plus or minus 1 inch long, 21 inches plus or minus 1 inch wide and 9 i
9、nches plus or minus 1/2 inch in depth. The tray shall have an elevated dissecting area, a removable perforated grid plate, a permanently attached instrument tray, and shall provide positive drainage. o. One large, single compartment, integral sink at the end of the table, approximately 11 inches dee
10、p, 12 inches wide, and 20 inches long. The autopsy table shall have overall dimensions of 34 inches plus or minus 1 inch high by 32-1/2 inches plus or minus 1 inch wide by 101 inches plus or minus 3 inches long and shall weigh 385 pounds plus or minus 15 pounds. The electrical components of the auto
11、psy table shall operate on single phase, alternating current, at 60 Hertz and 120 volts. The autopsy table leakage current shall comply with Underwriters Laboratories, UL 544, Standard for Medical and Dental equipment, Type C, non-patient equipment. Identification markinq. permanently marked with th
12、e following: Each table shall be legibly and National Stock Number Manufacturers name Contract number ODeration and service manuals. The contractor shall furnish manuals, handbooks, and/or brochures containing complete operation, installation and service/maintenance instructions with each table. 2 P
13、rovided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-54830 m 9999974 OL422b8 981 m A-A-5 4 8 3 O Workmanship. The autopsy table shall be free from defects which detract from its appearance or impair its serviceability. Unit. Each (EA). One autopsy tabl
14、e, as specified, constitutes one unit. Qualitv Assurance Provisions. Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the
15、 contract or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where s
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。