DOD A-A-53837-1989 ELECTROSURGICAL APPARATUS《电外科学的仪器》.pdf
《DOD A-A-53837-1989 ELECTROSURGICAL APPARATUS《电外科学的仪器》.pdf》由会员分享,可在线阅读,更多相关《DOD A-A-53837-1989 ELECTROSURGICAL APPARATUS《电外科学的仪器》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、A-A-53B37 65 W 7797779 0032342 7 W A-A- 53 83 7 30 May 1989 COMMERCIAL ITEM DESCRIPTION ELECTROSURGICAL APPARATUS The General Services Administration has authorized the use of this commercial item description. This commercial item description covers a lightweight, portable, AC, solid state electrosu
2、rgical apparatus. Salient characteristics: The electrosurgical apparatus shall be compact, solid state, lightweight and shall not exceed 45 pounds. The unit shall be portable, suitable for performing urological, and general surgical procedures. The outside dimensions of the unit shall not exceed 18
3、inches by 14 inches by 7 inches. The unit shall be equipped with an appropriately rated three-conductor line cord terminated in accordance with Underwriters Laboratories, Inc. (UL). The unit shall be equipped with approved hospital-grade 3-prong plug. The unit shall be suitable for operation on nomi
4、nal 115 volt, 50/60 Hz. electric power (230 volt, 50/60 Hz. when so specified in the procurement document). The output power for the unit shall not exceed 1500 VA for full operation. The unit shall have a minimum of the following connections for accessories: a. A connector or jack for the patient re
5、turn electrode. b. Connectors or jacks for simultaneous attachment of one standard type active electrode, one forceps type coagulating electrode with built-in control switch for selection of cut or coagulation modes. - a msc N/A FSC 6515 DISTRIBUTION STATEMENT A. Approved for public release; distrib
6、ution is unlimited. THIS DOCUMENT _- CONTAINS PAGES. Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-53837 65 W 9999974 0032343 O A-A-53 8 3 7 c. A connector or jack for attaching an explosion-proof or intrinsically safe foot switch to enable sel
7、ection of the cut or coagulation modes. d. A common return. The unit shall be supplied with the fol.lowing accessories: a. b. d. e. f. 4- h. i. C. A 1/16 inch ball electrode. A flat cutting electrode. A straight 1/2 inch blade Two reusable active electrode chuck handles with ten foot cords. A 45 deg
8、ree angle 1/2 inch cutting electrode. A straight needle point electrode. An explosion-proof or intrinsically safe (low voltage current limited) foot switch. A dummy load. Two patient plates with cords. The unit shall have a minimum of the following modes of operation with controls: a. b. C. output r
9、emain Cut mode: Cutting mode power shall be in range, which allows precise spee and depth of cutting effect. Coaguation mode: Coagulation motor power shall be in range, which allows exact hemostatic effect with minimum tissue destruction and charring. Modified cut mode: The modified cut mode shall c
10、ause the application of intermittent radio frequency (IRF) power to the active electrode to allow control of the degree of hemostasis. * 1 power, as measured at the active electrode, shall constant within 10 percent, with typical 105-130 volt line voltage fluctuations. The unit shall have isolated o
11、utput circuits., i.e., output not referenced to ground. b 2 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-53837 65 9777774 0032344 2 m . Unit shall meet NFPA 493, UL 540, and AAMI HF-18 Electrical Safety Standards. Al1 items shall conform to th
12、e applicable requirements of nationally recognized standards for the safety of rfiedcal equipment of the issue in effect on the date of solicitation. All items shall be approved as conforming to such a standard by an independent, nationally recognized testing laboratory. A-A-53837 The unit shall gen
13、erate niinimal radio frequency interference (RFI)? both radiated and conducted, through the power line. The active cable of the ECU shall be properly shielded to limit radiai-ed losses and prevent capacitive coupling into nearby wires and metal equipment. The unit shall be provided with a positive o
14、n/off switch. An indicator light shall signal that the ESU has been turned on. Separate visual indicators shall be incorporated to indicate when the ESU output has been activated. In addition, separate distinct audible tones shall indicate cut and coag output activation. An adjustable volume control
15、 which cannot reduce the sound level below 45 dB shall be provided. Overcurrent prokection for the ESU, preferably in the form of a circuit breaker, is required. If a fuse is used? a spare must be provided and clearly labeled as to function and rating. impinging against it. Circuit breakers niust be
16、 protected from other objects Mininium length of the cords and cables supplied shall be: Cable Type Line Cord Act j.ve Dispers ive Foot; s w i t-, c h - Leng t 11 3.0 meters (9.8 feet) 2.5 meters (8.2 feet) 2.4 meters (7.9 feet) 3.0 meters (9.8 feet) -.- Workmanship. defects which detract from its a
17、ppearance OK impair its serviceability. The electrosurgical apparatus shall be free from 3 Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-A-A-5 3 8 3 7 Unit. Each (ea). One electrosurgical apparatus along with accessories, as specified, constitutes
18、one unit. Quality Assurance Provisions. Responsibility for inspection. Unless otherwise specified the contract or Durchase order, the contractor is responsible in for the performance Qf all inspection requirements as specified herein. Except as otherwise specified in the contract or purchase order,
19、the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed ne
20、eessary to assure supplies and services conform to prescribed requirements. Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and rnade available to the Government upon the Governments request, at any time, or from time to time, durin
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- DODAA538371989ELECTROSURGICALAPPARATUS 外科学 仪器 PDF

链接地址:http://www.mydoc123.com/p-701686.html