1、Designation: E 676 02Standard Specification forInterchangeable Taper-Ground Joints1This standard is issued under the fixed designation E 676; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in pa
2、rentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers designations, dimensions, andtolerances for only the ground surfaces of interchangeabletaper-ground joints for laborator
3、y and industrial equipment infull-length, medium-length, short-length, and international-length (ISO K-6 series) grindings.NOTE 1The dimensions pertaining to full-length, medium-length, andshort-length joints, other than for the tolerance on length of grind, weretaken directly from the now obsolete
4、Commercial Standard CS 21.NOTE 2Tolerances on the diameters of the international-length jointsdo not agree with those given in ISO Recommendation R383. Thesetolerances have been established to agree with those of CS 21 in order toprevent the negative projection (large end of inner joint being smalle
5、r thanlarge end of outer joint) permitted by ISO.NOTE 3Although glass is the most commonly used material forground joints, other materials may be used as specified. Taper-groundjoints constructed from glass should be of suitable chemical and thermalproperties in accordance with Specification E 438.
6、Joints should be as freeas possible of visible defects and conform to Specification E 671.1.2 The following precautionary caveat pertains only to theTest Method portion, Section 4, of this specification. Thisstandard does not purport to address all of the safety concerns,if any, associated with its
7、use. It is the responsibility of the userof this standard to establish appropriate safety and healthpractices and determine the applicability of regulatory limita-tions prior to use.2. Referenced Documents2.1 ASTM Standards:E 438 Specification for Glasses in Laboratory Apparatus2E 671 Specification
8、for Maximum Permissible Thermal Re-sidual Stress in Annealed Glass Laboratory Apparatus2E 920 Specification for Commercially Packaged LaboratoryApparatus2E 921 Specification for Export Packaged Laboratory Appa-ratus2E 1133 Practice for Performance Testing of Packaged Labo-ratory Apparatus United Sta
9、tes Government Procurement2E 1157 Specification for Sampling and Testing ReusableLaboratory Glassware22.2 U.S. Department of Commerce Standard:CS 21 Interchangeable Taper-Ground Joints, Stopcocks,Stoppers, and Spherical-Ground Joints32.3 ISO Standard:R 383 Interchangeable Conical Ground Glass Joints
10、43. Requirements3.1 Designations, dimensions, and tolerances shall be asshown in Fig. 1 and in Table 1, Table 2, Table 3, and Table 4.3.2 Surface FinishThe roughness average (Ra)oftheground surface shall not exceed 1 m (40uim).3.3 TaperAll taper-ground joints shall have a taper of 160.006 mm of diam
11、eter per 10 mm of length (1 to 10).3.4 LeakageUsing the leak test apparatus illustrated inFig. 2, or one functionally similar to it, with the averagedifferential pressure during the test kept above 600 mm Hg, theleak rate calculated from the test data (see 4.9) shall not exceed0.05 torrlitres/s (or
12、Palitres/s) when the inner and outer jointsare assembled dry.4. Test Method4.1 Clean both joints to be tested with distilled water, andthen dry them with a clean paper towel. Next, wipe the groundzones with a clean chamois to remove any remaining particles.4.2 Tightly close the unground end of the j
13、oint to beconnected to the manometer system, using a rubber stoppercontaining a concentric hole (or rubber tubing, depending uponthe joint size under test) sized to fit snugly on the manometerconnection, and mount the joint on the connection in an uprightposition.4.3 Firmly plug the unground end of
14、the other type of jointwith a solid rubber stopper.4.4 Carefully mate the ground surfaces of the joints withoutexerting any pressure on them.4.5 Gently apply a 225-g weight on the upper joint forseating pressure, then rotate the upper joint 90 and remove theweight.4.6 After ensuring that the stopcoc
15、k to the vacuum pump isin the closed position, turn on the pump and adjust the stopcock1This specification is under the jurisdiction of ASTM Committee E41 onLaboratory Apparatus and is the direct responsibility of Subcommittee E 41.01 onApparatus.Current edition approved Oct. 10, 2002. Published Nov
16、ember 2002. Originallypublished as E 676 79. Last previous edition E 676 84 (1998).2Annual Book of ASTM Standards, Vol 14.04.3Discontinued 1979U.S. Department of Commerce, Washington, DC 20234.4Available from American National Standards Institute, 1430 Broadway, NewYork, NY 10018.1Copyright ASTM Int
17、ernational, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.to the pump so that the mercury in the manometer rises to alevel such that the average differential pressure will remainabove 600 mm at the conclusion of the test.4.7 Air leakage is now allowed to occur t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 1人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME6762002STANDARDSPECIFICATIONFORINTERCHANGEABLETAPERGROUNDJOINTS 互换 圆锥形 接头 标准规范 PDF

链接地址:http://www.mydoc123.com/p-533292.html