ASTM E608 E608M-2013 7594 Standard Specification for Mineral-Insulated Metal-Sheathed Base Metal Thermocouples《压实的矿物绝缘 金属铠装 基底金属热电偶电缆的标准规范》.pdf
《ASTM E608 E608M-2013 7594 Standard Specification for Mineral-Insulated Metal-Sheathed Base Metal Thermocouples《压实的矿物绝缘 金属铠装 基底金属热电偶电缆的标准规范》.pdf》由会员分享,可在线阅读,更多相关《ASTM E608 E608M-2013 7594 Standard Specification for Mineral-Insulated Metal-Sheathed Base Metal Thermocouples《压实的矿物绝缘 金属铠装 基底金属热电偶电缆的标准规范》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E608/E608M 13Standard Specification forMineral-Insulated, Metal-Sheathed Base MetalThermocouples1This standard is issued under the fixed designation E608/E608M; the number immediately following the designation indicates the yearof original adoption or, in the case of revision, the year
2、of last revision. A number in parentheses indicates the year of last reapproval.A superscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers the requirements for mineral-insulated, metal-sheathed base metal thermocouples for indus
3、-trial or high-reliability applications. It applies specifically tothermocouples fabricated from sheathed thermocouple mate-rial in accordance with Specification E585/E585M. The speci-fication provides for the selection of thermoelements,insulation, sheath material, measuring junction configuration,
4、thermocouple assembly length, and the type of transition ortermination.1.2 This specification also includes provisions for eitherquality assurance or verification program requirements, orboth, as a purchasers option.1.3 The values stated in either SI units or inch-pound unitsare to be regarded separ
5、ately as standard. The values stated ineach system may not be exact equivalents; therefore, eachsystem shall be used independently of the other. Combiningvalues from the two systems may result in non-conformancewith the standard.1.4 This standard does not purport to address all of thesafety concerns
6、, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E165 Practice for Liquid Penetrant Examinat
7、ion for GeneralIndustryE220 Test Method for Calibration of Thermocouples ByComparison TechniquesE230 Specification and Temperature-Electromotive Force(EMF) Tables for Standardized ThermocouplesE344 Terminology Relating to Thermometry and Hydrom-etryE585/E585M Specification for Compacted Mineral-Insu
8、lated, Metal-Sheathed, Base Metal ThermocoupleCableE780 Test Method for Measuring the Insulation Resistanceof Mineral-Insulated, Metal-Sheathed Thermocouples andThermocouple Cable at Room TemperatureE839 Test Methods for Sheathed Thermocouples andSheathed Thermocouple Cable2.2 AWS Standards:3A5.9 Sp
9、ecification for Corrosion-Resisting Chromium andChromium-Nickel Steel Welding Rods and Bare Elec-trodesA5.14 Specification for Nickel and Nickel-Alloy Bare Weld-ing Rods and Electrodes3. Terminology3.1 DefinitionsThe definitions given in Terminology E344shall apply to this specification.3.2 Definiti
10、ons of Terms Specific to This Standard:3.2.1 common ungrounded junction, nmeasuring junc-tions within the same multi-pair thermocouple that are electri-cally isolated from the sheath but electrically connected to eachother.3.2.2 isolated ungrounded junction, nmeasuring junctionswithin the same multi
11、-pair thermocouple that are electricallyisolated from the sheath and electrically isolated from eachother.3.2.3 lot, na quantity of thermocouples manufacturedfrom the same continuous length of mineral-insulated, metal-sheathed thermocouple cable.3.2.4 sheathed thermocouple material, na combination o
12、ftwo or more continuous thermoelements embedded in ceramicinsulation compacted within a metal protecting sheath.1This specification is under the jurisdiction of ASTM Committee E20 onTemperature Measurement and is the direct responsibility of Subcommittee E20.04on Thermocouples.Current edition approv
13、ed Nov. 1, 2013. Published December 2013. Originallyapproved in 1977. Last previous edition approved in 2006 as E608/E608M - 06.DOI: 10.1520/E0608_E0608M-13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of AS
14、TMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from American Welding Society (AWS), 550 NW LeJeune Rd.,Miami, FL 33126, http:/www.aws.org.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. Un
15、ited States14. Significance and Use4.1 This specification describes the physical requirements ofmineral-insulated, metal-sheathed base metal thermocouplesand establishes suitable test criteria for their evaluation.4.2 Standardized dimensional requirements and acceptableallowances for manufacturing v
16、ariations, are given.4.3 Aset of standard tests to be applied to all thermocouplescovered by this specification are described, in addition tooptional tests from which applicable additional requirementsmay be selected.4.4 WarningUsers should be aware that certain charac-teristics of thermocouples mig
17、ht change with time and use.4.4.1 If a thermocouples designed shipping, storage,installation, or operating temperature has been exceeded, thethermocouples moisture seal may have been compromised andmay no longer prevent the deleterious intrusion of water vapor.Consequently, the thermocouples conditi
18、on established by testat the time of manufacture may not apply later after anextended period of use, and retesting may become necessary.4.4.2 In addition, inhomogeneities can develop in thermo-elements because of exposure to temperature, even in caseswhere the maximum exposure temperatures have been
19、 lowerthan the suggested upper temperature limits of Table 1. For thisreason, calibration of thermocouples destined for delivery to acustomer is not recommended (see S6.1).4.4.3 Furthermore, because the EMF indication of anythermocouple depends upon the condition of the thermoele-ments along their e
20、ntire length, as well as the temperatureprofile pattern in the region of any inhomogeneity, the EMFoutput of a used thermocouple will be unique to its installation.Because temperature profiles in calibration equipment areunlikely to duplicate those of the installation, removal of a usedthermocouple
21、to a separate apparatus for calibration is notrecommended. Instead, in-situ calibration by comparison to asimilar thermocouple of known accuracy is recommended.5. Ordering Information5.1 The purchase documents shall specify the following:5.1.1 The type and quantity of thermoelement pairs, and thetol
22、erances, if other than standard (see 6.4 and 6.5).5.1.2 The nominal sheath diameter. See Table 1 or Specifi-cation E585/E585M for preferred nominal diameters.5.1.3 The insulating material (either MgO or Al2O3) and itsrespective type. See Specification E585/E585M.5.1.4 The sheath material. See Specif
23、ication E585/E585M.5.1.5 The style of measuring junction: Style G (grounded)or Style U (ungrounded). See Figs. 1 and 2. If more than onepair of thermoelements is specified, Style U is further subdi-vided into Style CU (common ungrounded) and Style IU(isolated ungrounded).45.1.6 The quantity, length,
24、 and length tolerance of eachthermocouple. See Figs. 3-6 for examples.5.1.7 The type and configuration of connection head,connector, transition piece, or termination, and moisture sealrequired on the end opposite the measuring junction. See Figs.3-6 for examples. The minimum and maximum intendedoper
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME608E608M20137594STANDARDSPECIFICATIONFORMINERALINSULATEDMETALSHEATHEDBASEMETALTHERMOCOUPLES 矿物 绝缘

链接地址:http://www.mydoc123.com/p-533144.html