ASTM D6526-2003e1 Standard Test Method for Analysis of Toluene by Capillary Column Gas Chromatography《毛细管柱气象色谱法分析甲苯的标准试验方法》.pdf
《ASTM D6526-2003e1 Standard Test Method for Analysis of Toluene by Capillary Column Gas Chromatography《毛细管柱气象色谱法分析甲苯的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM D6526-2003e1 Standard Test Method for Analysis of Toluene by Capillary Column Gas Chromatography《毛细管柱气象色谱法分析甲苯的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 6526 03e1Standard Test Method forAnalysis of Toluene by Capillary Column GasChromatography1This standard is issued under the fixed designation D 6526; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last
2、revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEEditorial change was made in 14.1.2 in June 2003.1. Scope*1.1 This test method covers the determination of hydrocar-bon impuriti
3、es typically found in, and the purity of, samplescontaining 98 wt % and greater toluene. This test method isapplicable to impurity concentrations in the range of 0.0005 to1.6 wt %.1.2 Monocyclic aromatic hydrocarbons containing 6through 8 carbon atoms, cumene, 1,4dioxane, and nonaro-matic aliphatic
4、hydrocarbons containing up to 12 carbon atomscan be detected by this test method. The nonaromatic com-pounds are determined as a composite.1.3 The following applies to all specified limits in this testmethod: for purposes of determining conformance with thistest method, an observed value or a calcul
5、ated value shall berounded off “to the nearest unit” in the last right-hand digitused in expressing the specification limit, in accordance withthe rounding-off method of Practice E29.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is there
6、sponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, see Section 9.2. Referenced Documents2.1 ASTM Standards:D 3437 Practice for Sampling and Handling Li
7、quid CyclicProducts2D 4790 Terminology of Aromatic Hydrocarbons and Re-lated Chemicals2E29 Practice for Using Significant Digits in Test Data toDetermine Conformance with Specifications3E 355 Practice for Gas Chromatography Terms and Rela-tionships4E 691 Practice for Conducting an Interlaboratory St
8、udy toDetermine the Precision of a Test Method3E 1510 Practice for Installing Fused Silica Open TubularCapillary Columns in Gas Chromatographs42.2 Other Document:OSHA Regulations, 29 CFR, paragraphs 1910.1000 and1910.120053. Terminology3.1 See Terminology D 4790 for definitions of terms used inthis
9、test method.4. Summary of Test Method4.1 A portion of the sample is injected into a gas chromato-graph using a microlitre syringe at the specified conditions ofthe test method. The toluene and other components areseparated as they are transported through the column by aninert carrier gas. The compon
10、ents in the effluent are measuredby a flame ionization detector (FID). The area of the impuritypeaks and toluene are electronically integrated. The peak areasare corrected with effective carbon number (ECN)6responsefactors and normalized to 100.0000 %.5. Significance and Use5.1 This test method is s
11、uitable for determining the concen-trations of known impurities in refined toluene and for use as1This test method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and Related Chemicals and is the direct responsibility ofSubcommittee D16.01 on Benzene, Toluene, Xylenes, Cycloh
12、exane, and TheirDerivatives.Current edition approved Jan. 10, 2003. Published March 2003. Originallyapproved in 2000. Last previous edition approved in 2000 as D 6526 - 00.2Annual Book of ASTM Standards, Vol 06.04.3Annual Book of ASTM Standards, Vol 14.02.4Annual Book of ASTM Standards, Vol 03.06.5A
13、vailable from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401.6Scanlon, J. T., and Willis, D. E., “Calculation of Ionization Detector RelativeResponse Factors Using the Effective Carbon Number Concept,” Journal ofChromatographi
14、c Science, Vol 35, August, 1985, pp. 333-339.1*A Summary of Changes section appears at the end of this standard.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.an integral quality control tool where toluene is produced orused in manuf
15、acturing.5.2 Toluene purity is reported, but a chromatographic analy-sis cannot determine absolute purity if unknown or undetectedcomponents are present in the sample.6. Interferences6.1 If present, nonaromatic hydrocarbons of 13 carbons orgreater, alcohols, ethers, and other similar organic compoun
16、dscan interfere with this test method by co-eluting with thearomatic hydrocarbons.6.2 Compounds not detected by a FID are not determined bythis test method.6.3 Nonvolatile material is not determined.7. Apparatus7.1 Gas Chromatograph (GC)any GC built for capillarycolumn chromatography. The system sha
17、ll have sufficientsensitivity, linearity, and range to obtain a minimum peakheight response for 0.0010 wt % impurity of twice the heightof the signal background noise, while not exceeding the fullscale of either the detector or the electronic integration for themajor component. It shall have a split
18、 injection system that willnot discriminate over the boiling range of the samples ana-lyzed. The system should be capable of operating at conditionsgiven in Table 1.7.2 Recorderelectronic integration is recommended.7.3 Capillary Columnfused silica capillary column with1,2,3-tris-2-cyano-ethoxypropan
19、e (TCEP) phase is recom-mended. Polyethylene glycol (PEG) columns have been suc-cessfully used. Other columns may be used after it has beenestablished that such a column is capable of separating allmajor impurities under operating conditions appropriate for thecolumn.7.4 Microsyringecapable of deliv
20、ering 1 L of sample.8. Reagents8.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. It is intended that all reagents shall conform tothe specifications of the Committee on Analytical Reagents oftheAmerican Chemical Society7, where such specifications areavailable, unless otherwis
21、e indicated. Other grades may beused, provided it is first ascertained that the reagent is ofsufficiently high purity to permit its use without lessening theaccuracy of the determination.8.2 Carrier GasHelium, 99.99 mole % minimum, isrecommended.8.3 FID Detector Gases:8.3.1 Hydrogen99.99 mole % mini
22、mum.8.3.2 Airless than 10 ppm each of total hydrocarbons andwater.9. Hazards9.1 Consult current OSHA regulations, suppliers MaterialSafety Data Sheets, and local regulations for all materials usedin this test method.10. Sample Handling10.1 Collect the samples in accordance with PracticeD 3437.10.2 T
23、o preserve sample integrity (consistency) and preventthe loss of volatile components, which may be in somesamples, do not uncover samples any longer than necessary.11. Preparation of Apparatus11.1 Follow the manufacturers instructions for mountingand conditioning the column in the chromatograph.11.2
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD65262003E1STANDARDTESTMETHODFORANALYSISOFTOLUENEBYCAPILLARYCOLUMNGASCHROMATOGRAPHY 毛细管 气象 色谱 分析

链接地址:http://www.mydoc123.com/p-522766.html