1、Designation: D 1092 05An American National StandardStandard Test Method forMeasuring Apparent Viscosity of Lubricating Greases1This standard is issued under the fixed designation D 1092; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revisi
2、on, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the Department of Defense.1. Scope1.1 This test method cov
3、ers measurement, in poises, of theapparent viscosity of lubricating greases in the temperaturerange from 54 to 38C (65 to 100F). Measurements arelimited to the range from 25 to 100 000 P at 0.1 s1and1to100 P at 15 000 s1.NOTE 1At very low temperatures the shear rate range may be reducedbecause of th
4、e great force required to force grease through the smallercapillaries. Precision has not been established below 10 s1.1.2 This standard uses inch-pound units as well as SI(acceptable metric) units. The values stated first are to beregarded as standard. The values given in parentheses are forinformat
5、ion only. The capillary dimensions in SI units in Fig.A1.1 and Fig. A1.2 are standard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices a
6、nd determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D88 Test Method for Saybolt ViscosityD 217 Test Methods for Cone Penetration of LubricatingGreaseD 3244 Practice for Utilization of Test Data to DetermineConformance with Specifications3
7、. Terminology3.1 Definitions:3.1.1 apparent viscosity, nof a lubricating grease is theratio of shear stress to shear rate calculated from Poiseuillesequation, and is measured in poises (see 10.1).3.1.2 capillary, nFor the purpose of this method, a capil-lary is any right cylindrical tube having a le
8、ngth to diameterratio of 40 to 1.3.1.3 shear rate, nthe rate at which a series of adjacentlayers of grease move with respect to each other; proportionalto the linear velocity of flow divided by the capillary radius,and is thus expressed as reciprocal seconds.4. Summary of Test Method4.1 The sample i
9、s forced through a capillary by means of afloating piston actuated by the hydraulic system. From thepredetermined flow rate and the force developed in the system,the apparent viscosity is calculated by means of Poiseuillesequation.Aseries of eight capillaries and two pump speeds areused to determine
10、 the apparent viscosity at sixteen shear rates.The results are expressed as a log-log plot of apparent viscosityversus shear rate.5. Significance and Use5.1 Apparent viscosity versus shear rate information can beuseful in predicting pressure drops in grease distributionsystems under steady-state flo
11、w conditions at constant tem-perature.6. Apparatus6.1 The assembled pressure viscometer consists of fourmajor divisions, the power system, the hydraulic system, thegrease system (described in the annex and shown in Fig. 1),and a bath of optional design. Fig. 2 is a photograph of the firstthree divis
12、ions as commonly used at room temperature. Thisform of the apparatus can be used with a cylindrical insulatedtank 178 mm (7 in.) in diameter and 508 mm (20 in.) deep. Thebath medium may be kerosene or alcohol cooled manually withdry ice. Alternatively the grease system, the grease andhydraulic syste
13、m, or all three major divisions can be built intoany liquid or air bath that will cover the temperature range andmaintain the grease at test temperature 60.25 C (60.5F).1This test method is under the jurisdiction of ASTM Committee D02 onPetroleum Products and Lubricants and is the direct responsibil
14、ity of SubcommitteeD02.G0 on Lubricating Grease.Current edition approved Nov. 1, 2005. Published November 2005. Originallyapproved in 1950. Last previous edition approved in 1999 as D 109299.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serv
15、iceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.7. Sampling7.1 A single filling of the grease cylinder
16、requires about0.223 kg (12 lb) of grease which is the minimum size sample.NOTE 2It is possible for an experienced operator to complete the 16single determinations with a single filling. However, some samples reachthe equilibrium pressure slowly, making it advisable to have a sample ofseveral pounds
17、available.7.2 Generally no special preparation of the sample is nec-essary.NOTE 3The apparatus works the samples to some extent as they passthrough the capillary. Somewhat better precision is obtained if they arepreviously worked as described in Test Methods D 217. Working of somegreases may cause a
18、eration.NOTE 4It is desirable to filter some greases through a 60-mesh screento prevent plugging the No. 8 capillary. Follow prudent laboratory practiceto keep equipment cleaned and flushed before use.8. Calibration and Standardization8.1 To calibrate the hydraulic system, remove the greasecylinder
19、and replace it with a needle valve. Select a hydraulicoil of about 2000 cSt (2000 mm2/s) viscosity at the testtemperature. Fill the system with hydraulic oil and circulate theoil until it is free of air bubbles. At atmospheric pressure,quickly place a 60-mL Saybolt receiving flask (Test MethodD88),
20、under the outlet and start a timer. Determine the deliverytime for 60 mL and calculate the flow rate in cubic centimetresper second assuming 1 mL equal to 1 cm3. Repeat thisobservation at 500, 1000, 1500 psi (3.45, 6.89, 10.4 MPa) andat sufficient pressures above 1500 psi to develop a calibrationcur
21、ve of the type as shown in Fig. 3. The developed curve ofthe type is used to correct flow rates when grease is dispensed.Repeat the calibration at intervals to determine if wear ischanging the pump flow.8.2 An alternative procedure for the calibration of thehydraulic system is the measurement of the
22、 rate of flow of thetest grease. To cover the desired range of shear rates, flow ratesFIG. 1 Schematic Drawing of ApparatusFIG. 2 Photograph of ApparatusD1092052over an approximate range of pressure are determined. Anysuitable means of measuring the rate of grease flow may beused.9. Procedure9.1 Cha
23、rge the sample so as to reduce inclusion of air to aminimum. Soft greases may be poured into the cylinder ordrawn up by vacuum; heavy samples must be hand packed.When filling the cylinder by vacuum, remove the capillary endcap and place the piston flush with the open end and then insertinto the samp
24、le. Apply vacuum to the opposite end of thecylinder until the cylinder is fully charged with grease. Thismust be facilitated by tapping with a wooden block. Replacethe capillary end cap and fill the upper end of the cylinderabove the piston with hydraulic oil.9.2 Fill the entire hydraulic system wit
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD10922005STANDARDTESTMETHODFORMEASURINGAPPARENTVISCOSITYOFLUBRICATINGGREASES 润滑脂 测量 表观 粘度 标准 试验 方法

链接地址:http://www.mydoc123.com/p-509249.html