ASHRAE 4704-2004 Analysis and Validation of a Psychrometric Apparatus《分析和验证某状态器具》.pdf
《ASHRAE 4704-2004 Analysis and Validation of a Psychrometric Apparatus《分析和验证某状态器具》.pdf》由会员分享,可在线阅读,更多相关《ASHRAE 4704-2004 Analysis and Validation of a Psychrometric Apparatus《分析和验证某状态器具》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、4704 Analysis and Validation of a Psychrometric Apparatus Joo Batista Dias ABSTRACT This article uses the analysis ofvariance as a tool to vali- date a calibration apparatus for temperature and relative humidity sensors. The validation procedure establishes the operational ranges for measurement of
2、the dry-bulb (tdb) and wet-bdb (twb) temperatures andofthe relative humidity of air (RH) for this particular apparatus. Analyses are carried out testing twb as afunction of air speed ranging from I. 25 to 5.00 m/s and twb as afunction of RHand also compares two envi- ronments to calibrate tdb: eithe
3、r in thermostatic bath or in the airjlow of a test section. The results show that air speed on wet covered semors within the tested range of I. 25 io 5. O m/s does not change the measured twb, thus widening the range suggested in ASHRAEStandard41.1-1974. Howevecfor rela- tive humidity of air within
4、the studied range, significant differ- ences are found, and these are close to the 2-3% indicated range. The use of different calibrated sensorsfor tdb result in different calculated RHc values in a controlled environment. INTRODUCTION Data presented by the manufacturers of some electronic temperatu
5、re sensors show that some ofthem have high thermal dissipation. Some semiconductor sensors present a power dissipation of approximately 4.4 mW 15 V. The use of a sensor covered with a cotton sock to obtain relative humidity of air (RH), which provides the measure of the wet-bulb temperature (twb), i
6、ntroduces a superficial thermal resistance in the sensor, and this may interfere in the process of heat transfer between the air and the sensor. This fact indicates that the sensor must be calibrated again. Therefore, it was found that a calibration apparatus should be built in order to test this co
7、nditions. Paulo Smith Schneider, Ph.D. Member ASHRAE This article presents a calibration apparatus, based on ASHRAE standard41 .l-1974, using the method of analysis of variance (ANOVA) as a validation tool (Dias 2001). CALIBRATION APPARATUS The apparatus basically consists of three subsystems, as sh
8、own in Figure I. Subsystem I is the airflow circuit. Air temperature and humidity are controlled by a conditioning chamber subsystem 21. Subsystem 3 is the test section, where calibration by comparison is carried out. Description of the Apparatus Subsystems Closed Ventilation Circuit-Subsystem 1. Th
9、is consists of piping, a fan, and a flow variator. Piping is made of conventional 100 rnm diameter PVC plastic, which is used in buildings and thermally isolated by a sheet of aluminum- coated polystyrene foam, which is commonly used to cover building roofs. The centrifugal fan is the same as those
10、used in Figure 1 General vision of the calibration apparatus. Joo Batista Dias is a doctoral student and Paulo Smith Schneider is a mechanical engineer at the Federal University of Rio Grande, Porto Alegre, Rio Grande do Sul, Brazil. 02004 ASHRAE. 117 RAC air-conditioners, and the flow variator was
11、built with a plate restricting the the flow cross-sectional area. The subsystem is provided with a micro-manometer with a pitot tube, as shown in Figure 2. Air-conditioning Chamber-Subsystem 2. The function of this subsystem is to supply air at controlled temperature and humidity in order to carry o
12、ut the tests. The chamber consists of a sensitive heat exchange circuit and a vaporizer. Air temperature control is obtained only by heating, depending on environmental conditions. The sensitive heat exchange device consists of an automatic thermostatic bath, which supplies heated water to a heat ex
13、changer placed inside the chamber. The exchanger is a room air conditioner (RAC) evaporator with 2.2 kW (7500 Bhdh) capacity. A water vaporizer is also placed inside the chamber. Vapor injection into the chamber is controlled by a PPC (programmable process controller), connected to a capacitive sens
14、or, whose only function is to measure relative humidity levels and therefore supply control signal information. The structure of the chamber consists of an expanded polystyrene box, internalIy covered with a galva- nized steel plate aimed at protecting the chamber if it is exposed to high air temper
15、ature and humidity. Test Section-Subsystem 3. The test section shown in Figure 3 is part of the piping, where the sensor elements are placed for testing. It consists of reference and test temperature sensor elements, a sock humidifier, a distilled water reservoir, and other accessories. The sensor e
16、lements are placed on the testing section transverse to airflow. The calibration procedure employs comparisons, where the different tested sensors are compared to a reference thermometer. The testing section ensures that the air that flows throughout the system is under the same conditions. The posi
17、tions ofthe reference and the test sensors follow the recommendations of ASHRAE Standard 41.1-1 974. Also according to this standard, the sensors must penetrate the piping up to its circumference center. The sensors for measuring wet-bulb temperature must have a cotton fabric sock or other hygroscop
18、ic material on the sensi- tive part. The sock must cover the sensor element, which must be at a minimal distance of 25.4 mm (1.0 in.) of the levei of water of the sock humidifier, considering that the diameter of air outflow piping is 100 mm (3.93 in.). Taking into consideration that minimum and max
19、imum air speed within the pipe is 1.25 mis and 5.00 mis (246.06 and 984.25 ftlmin), respectively, the Reynolds number varies between 8600 and 34400, assuming an air density of 1.225 kg/ m3 (0.07647425 lb,/ft3), air absolute viscosity of 1.78110 kg/ms (7.18110 Ib,/fimin), and temperature of 15C (59F)
20、. These results, which are higher than the critical Reynolds number (Re,=2300), indicate turbulent flow. There- fore, the position of the sensors must be within a minimal distance of 10 diameters from any accessory, such as curves or restrictions (Hansen 1974). The diameter of the piping used was O.
21、 10 m (3.93 in.), which leads to a minimal distance of 1 .O rn (39.36 in:). thmnostic bah- 1.63 m (49.21 in) Re sensor flow vuiator W air conditioning- chamber ZJ vaporizer Figure 2 Design of the calibration apparatus. Figure 3 Testing section. FUNDAMENTALS Psychrometry A psychrometer is an instrume
22、nt capable of determining relative air humidity from dry and wet-bulb temperatures. When a wet bulb is exposed to an airflow, water evaporates from the tissue, determining an equilibrium temperature called wet-bulb temperature (twb). This process is not an adiabatic saturation process, which defines
23、 the thermody- namic wet-bulb temperature, but rather simultaneous heat and mass transfers in the wet bulb (ASHRAE 1994). In order to avoid saturation of the environment surrounding the sensor, a small fan is added on the wet sensor to promote the aspiration of air flowing over it. Calculated relati
24、ve humidity of air (RHc) in YO is defined by Equation 1, obtained from the sequence of calculations proposed by ASHRAE (1 997). 118 ASHRAE Transactions: Research wherep is total atmospheric pressure (Pa);pws/db is the partial pressure of water vapor in saturation at tdb (Pa), and p is the degree of
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASHRAE47042004ANALYSISANDVALIDATIONOFAPSYCHROMETRICAPPARATUS 分析 验证 状态 器具 PDF

链接地址:http://www.mydoc123.com/p-454287.html