《AIR FORCE MIL-PRF-32133-2003 X-RAY APPARATUS SET RADIOGRAPHIC INDUSTRIAL PORTABLE 160KVCP [Superseded AIR FORCE A-A-58077]《160KVCP型便携式工业X光线照相术X光设备设置 [代替 空军 A-A-58077]》.pdf》由会员分享,可在线阅读,更多相关《AIR FORCE MIL-PRF-32133-2003 X-RAY APPARATUS SET RADIOGRAPHIC INDUSTRIAL PORTABLE 160KVCP [Superseded AIR FORCE A-A-58077]《160KVCP型便携式工业X光线照相术X光设备设置 [代替 空军 A-A-58077]》.pdf(19页珍藏版)》请在麦多课文档分享上搜索。
1、 MIL-PRF-32133 14 November 2003 SUPERSEDING A-A-58077 6 August 1996 PERFORMANCE SPECIFICATION X-RAY APPARATUS SET, RADIOGRAPHIC, INDUSTRIAL, PORTABLE, 160KVCP This specification is approved for use by all Departments and Agencies of the Department of Defense. 1. SCOPE 1.1 Scope. This specification c
2、ontains the technical and performance requirements for a light weight, portable, industrial x-ray apparatus set. These x-ray apparatus sets are intended for use in a military aircraft maintenance environment, which may be at fixed military bases, deployed to austere forward operating locations, or n
3、aval vessels afloat. This specification provides for first article testing described herein. 1.2 Classification. The x-ray apparatus set should conform to the following types: Type I - X-ray apparatus set with a panoramic, liquid-cooled tubehead. Type II - X-ray apparatus set with a directional, liq
4、uid-cooled tubehead. Type III - X-ray apparatus set with a directional, air-cooled tubehead. Type IV - X-ray apparatus set with a directional, hybrid liquid/air-cooled tubehead Comments, suggestions, or questions on this document should be addressed to: AFRL/MLS-OL, 4750 Staff Drive, Tinker AFB OK 7
5、3145-3317. Since contact information can change, you may want to verify the currency of this address information using the ASSIST Online database at www.dodssp.daps.mil. AMSC: N/A FSC 6635 DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited. INCH-POUND Provided by IHSNot
6、 for ResaleNo reproduction or networking permitted without license from IHS-,-,-MIL-PRF-32133 22. APPLICABLE DOCUMENTS 2.1 General. The documents listed in this section are specified in sections 3 and 4 of this specification. This section does not include documents cited in other sections of this sp
7、ecification or recommended for additional information or as examples. While every effort has been made to ensure the completeness of this list, document users are cautioned that they must meet all specified requirements documents cited in sections 3 and 4 of this specification, whether or not they a
8、re listed. 2.2 Government documents. 2.2.1 Specifications, standards, and handbooks. The following specifications, standards, and handbooks form a part of this document to the extent specified herein. Unless otherwise specified, the issues of these documents are those listed in the issue of the Depa
9、rtment of Defense Index of Specifications and Standards (DoDISS) and supplements thereto, cited in the solicitation (see 6.2). SPECIFICATIONS DEPARTMENT OF DEFENSE MIL-PRF-28800 - Test Equipment for use with Electrical and Electronic Equipment, General Specification For (Unless otherwise indicated,
10、copies of the above specifications, standards, and handbooks are available from the Standardization Document Order Desk, 700 Robbins Avenue, Building 4D, Philadelphia, PA 19111-5094.) 2.2.2 Other government documents, drawings, and publications. The following government documents, drawings, and publ
11、ications form a part of this document to the extent specified herein. Unless otherwise specified, the issues are those specified in the solicitation. PUBLICATIONS Environmental Protection Agency (EPA) List of Pollutants/Toxics (Replacement to SD-14, Listing of Toxic Chemicals, Hazardous Substances,
12、and Ozone-Depleting Chemicals). See EPA website: http:/www.epa.gov/ebtpages/pollutants.html Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-MIL-PRF-32133 3Code of Federal Regulations 1040.10 Laser Products. See Food and Drug Administration website: h
13、ttp:/www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfCFR/showCFR.cfm?CFRPart=1040tubehead current actual and set; exposure time elapsed/remaining and set. 4.5.4.3 Test control unit to ensure continuously adjustable over the range of 20-160kVcp in 5kV steps minimum and 0.1-5.0 mA in 0.1mA steps. 20kV, h
14、ighest achievable mA: _ 160kV, highest achievable mA: _ 0.1mA, lowest achievable kV: _ 5.0mA, highest achievable kV: _ 4.5.4.4 Using ASTM E 1165, pinhole camera method, measure the effective total focal spot size for directional tubeheads, to ensure it is no greater than 0.060 square inches. Actual
15、size: _ 4.5.4.5 Using a ratemeter with a probe installed in a 2.5 inch lead tube, place the probe 19.6 inches from the center of the anode, energize the tubehead to 160kVcp/5mA, and ensure the radiation output is a minimum of 230 Roentgens per minute for directional tubeheads or 14 Roentgens per min
16、ute for panoramic tubeheads. Actual output at 160kVcp/5mA: _ Note: Contact the governments cognizant engineer for acceptable ratemeters. 4.5.4.6 Change tubehead types for each control unit and re-accomplish paragraph 4.5.4.5 to ensure radiation output does not vary by more than +/- 3 percent. Actual
17、 output at 160kVcp/5mA: _ Provided by IHSNot for ResaleNo reproduction or networking permitted without license from IHS-,-,-MIL-PRF-32133 114.5.4.7 Repeat paragraph 4.5.4.5 with 0.50 inch thick aluminum plate between probe and tubehead and ensure the radiation output is a minimum of 13 Roentgens per
18、 minute for directional tubeheads or 6 Roentgens per minute for panoramic tubeheads. Actual output at 160kVcp/5mA: _ 4.5.4.8 Using the ratemeter with a probe, place the directional tubehead (N/A for panoramic) in a horizontal position with the beryllium window facing down and blocked with 25 half-va
19、lue-layers of lead. Using a mechanical probe drive, plot the leakage rate in polar coordinates at 39.37 inches away from the tubehead. Ensure the leakage radiation is less than 2.0 roentgens per hour. Actual minimum leakage radiation: _ Actual maximum leakage radiation: _ 4.5.4.9 Ensure the x-ray ap
20、paratus shall produce a H&D density of not less than 2.0 through 2.0 inches of 1000-series aluminum at a source to film distance of 36 inches in 3 minutes or less using Kodak type M100 or equivalent film at 160kVcp/5mA. Actual film density: _ 4.5.4.10 Ensure the x-ray apparatus shall be capable of s
21、howing a 0.010 inch hole in a 0.010 inch thick penetrameter placed on 1.0 inch thick block of 1000-series aluminum at a source to film distance of 36 inches on Kodak type M100 or equivalent film with less than a 10 minute exposure. Procedure kVcp: _ Procedure mA: _ Procedure time: _ 4.5.4.11 Ensure
22、the directional (N/A for Type I) x-ray apparatus shall be capable of showing a density difference of 2.5 H&D units between aluminum foil 0.003 inches and 0.030 inches thick at a source to film distance of 36 inches on Kodak type M100 or equivalent film with less than a 10 minute exposure. Procedure
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- AIRFORCEMILPRF321332003XRAYAPPARATUSSETRADIOGRAPHICINDUSTRIALPORTABLE160KVCPSUPERSEDEDAIRFORCEAA58077160KVCP

链接地址:http://www.mydoc123.com/p-428216.html