《ASTM C1428-2018 Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single&x2013 Standard Gas Source Multiple Collector Mass Spectrometer Meth.pdf》由会员分享,可在线阅读,更多相关《ASTM C1428-2018 Standard Test Method for Isotopic Analysis of Uranium Hexafluoride by Single&x2013 Standard Gas Source Multiple Collector Mass Spectrometer Meth.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: C1428 11C1428 18Standard Test Method forIsotopic Analysis of Uranium Hexafluoride bySingleStandard Gas Source Multiple Collector MassSpectrometer Method1This standard is issued under the fixed designation C1428; the number immediately following the designation indicates the year oforigi
2、nal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method is applicable to the isotopic analysis of urani
3、um hexafluoride (UF6) with 235U concentrations less than orequal to 5 % and 234U, 236U concentrations of 0.0002 to 0.1 %.1.2 This test method may be applicable to the analysis of the entire range of 235U isotopic compositions providing that adequateCertified Reference Materials (CRMs or traceable st
4、andards) are available.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibilityof the user of this standard to establish appropriate safety health safety, health, and environmental practices and determine theapplicability of
5、regulatory limitations prior to use.1.4 This international standard was developed in accordance with internationally recognized principles on standardizationestablished in the Decision on Principles for the Development of International Standards, Guides and Recommendations issuedby the World Trade O
6、rganization Technical Barriers to Trade (TBT) Committee.2. Referenced Documents2.1 ASTM Standards:2C787 Specification for Uranium Hexafluoride for EnrichmentC859 Terminology Relating to Nuclear MaterialsC996 Specification for Uranium Hexafluoride Enriched to Less Than 5 % 235U2.2 Other Document:USEC
7、-651 Uranium Hexafluoride: A Manual of Good Handling Practices33. Terminology3.1 For definitions of terms relating to the nuclear fuel cycle, refer to Terminology C859.4. Summary of Test Method4.1 The unknown sample and a CRM or traceable standard whose isotopic composition is close to that of the s
8、ample areintroduced in sequence into the mass spectrometer, and 234U, 235U, 236U, and 238U ions are focused through correspondingcollector slits to the four separate collectors. Measurements are made that are proportional to the ratios of 234U, 235U, or 236Uto 238U. With the known composition of the
9、 CRM or traceable standard, these ratios of molar ratios permit calculation of the 234U,235U, and 236U contents. Memory corrections are applied based on the periodic measurement of two CRMs or traceable standards.5. Significance and Use5.1 Uranium hexafluoride is a basic material used to produce nuc
10、lear reactor fuel. To be suitable for this purpose, the materialmust meet criteria for isotopic composition. This test method is designed to determine whether the material meets the requirementsdescribed in Specifications C787 and C996.1 This test method is under the jurisdiction of ASTM Committee C
11、26 on Nuclear Fuel Cycle and is the direct responsibility of Subcommittee C26.05 on Methods of Test.Current edition approved March 1, 2011Nov. 1, 2018. Published April 2011December 2018. Originally approved in 1999. Last previous edition approved in 20052011as C142899(2005).C1428 11. DOI: 10.1520/C1
12、428-11.10.1520/C1428-18.2 For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at serviceastm.org. For Annual Book of ASTM Standardsvolume information, refer to the standards Document Summary page on the ASTM website.3 Available from United States Enrich
13、ment Corporation, 6903 Rockledge Drive, Bethesda, MD 20817.This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Becauseit may not be technically possible to adequately depict all chang
14、es accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current versionof the standard as published by ASTM is to be considered the official document.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. Uni
15、ted States16. Apparatus6.1 A gas source multiple collector mass spectrometer with the following attributes:6.1.1 The resolving power of the mass spectrometer is not less than 500. The resolving power (R) is calculated from theregistered mass spectrum of both the 235UF5+ and 238UF5+ isotopes as follo
16、ws:R 5 aMbM (1)wherewhere:a = distance between centers of the 235UF5+ and 238UF5+ peaks,b = peak width of the 238UF5+ isotope (10 % valley),M = 333 mass(u) 238UF5+, andM = 3 = 333 330, 330 mass(u) 235UF5+.6.1.2 The abundance sensitivity of the mass spectrometer is specified as less than 1105 as cont
17、ribution of mass 333 (238UF5+)to mass 331 (236UF5+).6.1.3 The four collectors have collector slits adjusted for ions of masses 329, 330, 331, and 333. Ion currents are amplified byfour amplifiers, having noise level less than 0.5 mV.6.1.4 The ion beams are kept within the slits by an automatic beam
18、positioner circuit.6.1.5 The pumping system of the mass spectrometer analyser tube shall maintain a pressure less than 5108 torr with a sampleflowing into the ion source.6.1.6 The memory correction factor of the mass spectrometer as defined in 9.110.1 shall be consistent with the required accuracyan
19、d precision, and shall not exceed 1.005.6.1.7 The sample inlet system shall be equipped with a manifold, including adjustable leak, and valves for introducing thesample and CRM or traceable standard in sequence and for evacuating corresponding lines. The pumping system of the inletsystem must mainta
20、in a pressure less than 2102 torr by evacuating.7. Hazards7.1 Since UFUranium hexafluoride (UF6) is radioactive, toxic, and highly reactive,reactive especially with reducing substancesand moisture (see USEC-651), appropriate facilities and practices for analysis shall be provided.moisture. Appropria
21、te laboratoryfacilities, materials of construction, and techniques shall be utilized when handling UF6 (see for example USEC-651).7.2 When released to atmosphere, gaseous UF6 reacts with moisture to produce HF gas and toxic UO2F2 particulates. Usesufficient ventilation or respiratory protection to a
22、void breathing fumes. Use appropriate personal protective equipment such asgloves, eye, and face protection. Consult the Safety Data Sheet (SDS) for additional information.8. Calibration and Standardization8.1 Uranium Hexafluoride IsotopicIsotopic:8.1.1 Two CRMs or traceable standards are used to de
23、termine the memory correction factor. The 235U concentration ratio ofupper CRM or traceable standard C235SH ! to 235U of lower CRM or traceable standard C235SL ! shall not be more than three C235SH /C235SL #3).8.1.2 For memory correction factor determination for 234U and 236U isotopes, two CRMs or t
24、raceable standards are used whichdiffer in 234U (236U) concentration. In concentration range 0.0002 to 0.01,C234 236!SH /C234 236!SL #8; in concentration range 0.01 to0.1,C234 236!SH /C234 236!SL #6.8.1.3 The CRMs or traceable standards used for measurements may differ in 235U concentration from a s
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMC14282018STANDARDTESTMETHODFORISOTOPICANALYSISOFURANIUMHEXAFLUORIDEBYSINGLEX2013STANDARDGASSOURCEMULTIPLECOLLECTORMASSSPECTROMETERMETHPDF

链接地址:http://www.mydoc123.com/p-1243471.html