UL SUBJECT 1365-2018 UL Outline for Investigation Corrugated Medical Tubing (CMT) Systems (Issue 5).pdf
《UL SUBJECT 1365-2018 UL Outline for Investigation Corrugated Medical Tubing (CMT) Systems (Issue 5).pdf》由会员分享,可在线阅读,更多相关《UL SUBJECT 1365-2018 UL Outline for Investigation Corrugated Medical Tubing (CMT) Systems (Issue 5).pdf(16页珍藏版)》请在麦多课文档分享上搜索。
1、UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULJANUARY 29, 20181UL 1365Outline of Investigation for Corrugated Medical Tubing (CMT) SystemsIssue Number: 3January 29, 2018Summary of TopicsThis third issue of the Outline of Investigation for Co
2、rrugated MedicalTubing (CMT) Systems, UL 1365, which covers corrugated medical tubingsystems intended for use in hospitals and health centers for distributionof nonflammable medical gas, is being issued to address correctionsto Section 18, Mechanical Tube Fitting Performance.ULs Outlines of Investig
3、ation are copyrighted by UL LLC. Neither a printed norelectronic copy of an Outline of Investigation should be altered in any way. All ofULs Outlines of Investigation and all copyrights, ownerships, and rights regardingthose Outlines of Investigation shall remain the sole and exclusive property of U
4、LLLC.COPYRIGHT 2018 UL LLCUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULJANUARY 29, 2018UL 13652No Text on This PageUL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULCONTENTSINTRODUCTION
5、1 Scope .42 Units of Measurement .43 Undated References 44 Components 4CONSTRUCTION5 General 5PERFORMANCE6 General 57 External Leakage Test .68 Hydrostatic-Strength Test 69 Impact Test 610 Axial Tension Test .711 Torsion Test .712 Temperature Test 713 Flame Test 814 Compression Test .915 Bending Tes
6、t 916 Electrical Resistance Test .917 Jacket Burning Characteristics .1118 Mechanical Tube Fitting Performance .1119 System Piping Purge Test .1220 System Piping Purity Test .12MARKINGS21 General 14INSTRUCTIONS22 General 1423 Installation and Operating Instructions .15JANUARY 29, 2018 UL 1365 3UL CO
7、PYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULINTRODUCTION1 Scope1.1 This outline covers corrugated medical tubing systems intended for use in hospitals and healthcenters for distribution of nonflammable medical gas in accordance with the Standar
8、d for Health CareFacilities, NFPA 99.1.2 CMT systems are intended for inside aboveground and in slab, when installed in approvednonmetallic conduit, applications only.1.3 This outline covers medical tubing systems up to 2 inches (50.8 mm), based on the nominal insidediameter of the tubing.1.4 These
9、products have not been evaluated for use after fires, natural disasters such as earthquakes, orexcessive physical damage beyond the expected assembly, installation and uses as identified in theserequirements.1.5 These products are intended to be periodically inspected and maintained for continued se
10、rvice, ortaken out of service if necessary by qualified persons in accordance with industry recommended practicesand/or the manufacturers instructions.2 Units of Measurement2.1 Values stated without parentheses are the requirement. Values in parentheses are explanatory orapproximate information.3 Un
11、dated References3.1 Any undated reference to a code or standard appearing in the requirements of this outline shall beinterpreted as referring to the latest edition of that code or standard.4 Components4.1 Except as indicated in 4.2, a component of a product covered by this standard shall comply wit
12、h therequirements for that component.4.2 A component is not required to comply with a specific requirement that:a) Involves a feature or characteristic not required in the application of the component in theproduct covered by this outline, orb) Is superseded by a requirement in this outline.4.3 A co
13、mponent shall be used in accordance with its rating established for the intended conditions ofuse.JANUARY 29, 2018UL 13654UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM UL4.4 Specific components are incomplete in construction features or restri
14、cted in performance capabilities.Such components are intended for use only under limited conditions, such as certain temperatures notexceeding specified limits, and shall be used only under those specific conditions.CONSTRUCTION5 General5.1 A product shall employ materials found by investigation and
15、 allowed by NFPA 99 to be acceptable forthe intended application.5.2 Corrugated medical tubing systems shall be constructed entirely of new and unused parts andmaterials.5.3 A product shall be assembled so that it will not be adversely affected by the vibration of normaloperation.PERFORMANCE6 Genera
16、l6.1 Representative samples of each size and specific construction of the CMT are to be subjected to thetests described in these requirements.6.2 CMT systems are to be tested on the following basis consistent with the intended use of the productand Table 6.1:a) Size or diameter;b) Minimum radius of
17、bend (per manufacturers instructions); andc) Maximum rated pressure (per manufacturers instructions).Table 6.1CMT systemSize (in.)OAL (in.) hydrostatic,vibration, torsion tests Length over diameter ratio Torsional stress (in-lb)3/8 17 37 1501/2 21 34 15034 26 28 2001 26 26 2501-12 36 24 3502 45 23 4
18、506.3 For tests requiring pressure a calibrated pressure-indicating device is to be mounted to indicatepressures developed within the vessel. The pressure-indicating device shall comply with one of thefollowing:a) An analog gauge having a pressure range of at least 150 percent of the anticipatedmaxi
19、mum working pressure;JANUARY 29, 2018 UL 1365 5UL COPYRIGHTED MATERIAL NOT AUTHORIZED FOR FURTHER REPRODUCTION ORDISTRIBUTION WITHOUT PERMISSION FROM ULb) A digital pressure transducer, or other digital gauge, that is calibrated over a range ofpressure that includes the test pressure; orc) Other dev
20、ice that is equivalent to the devices in (a) or (b).7 External Leakage Test7.1 A CMT system shall not leak when tested as described below.7.2 One sample of each diameter, not less than 3 feet (914 m) in length, shall be installed with one endconnected to an air supply source and the other end plugge
21、d. The pressure is slowly increased to 1-1/2times rated pressure and maintained at that pressure for at least 1 minute.8 Hydrostatic-Strength Test8.1 A medical tubing system shall not leak nor rupture when tested as described below.8.2 One sample of each diameter per Table 6.1 having a minimum lengt
22、h of 24 inches (610 mm) shall beused in this test. The sample is to be installed in a straight line, with one end attached to the pressuresource and other end plugged and free. Each sample is to be filled with water so as to exclude all air.8.3 The pressure is to be uniformly raised to 1-1/2 times r
23、ated pressure and maintained at that pressurefor at least 1 minute. The tubing is then to be bent with a radius of curvature as specified in Table 6.1 andthe pressure maintained for one additional minute.8.4 With the tubing maintained in the bent position, the pressure is then to be increased to 3.5
24、 timesmaximum rated at least pressure for 1 minute.9 Impact Test9.1 The medical tubing system shall withstand an impact of 30 ft-lb (40.67 Nm), using a 10 lb.(4.54 kg)weight falling 3 ft. (0.914 m) through a vertical guide tube, without external leakage, per Section 7, beforeand after having been te
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ULSUBJECT13652018ULOUTLINEFORINVESTIGATIONCORRUGATEDMEDICALTUBINGCMTSYSTEMSISSUE5PDF

链接地址:http://www.mydoc123.com/p-1069163.html