REG NASA-LLIS-0542-1996 Lessons Learned Use of RGA Quadrupole Mass Spectrometer for Leak Detection.pdf
《REG NASA-LLIS-0542-1996 Lessons Learned Use of RGA Quadrupole Mass Spectrometer for Leak Detection.pdf》由会员分享,可在线阅读,更多相关《REG NASA-LLIS-0542-1996 Lessons Learned Use of RGA Quadrupole Mass Spectrometer for Leak Detection.pdf(2页珍藏版)》请在麦多课文档分享上搜索。
1、Lessons Learned Entry: 0542Lesson Info:a71 Lesson Number: 0542a71 Lesson Date: 1996-04-01a71 Submitting Organization: LARCa71 Submitted by: Ruth M. AmundsenSubject: Use of RGA Quadrupole Mass Spectrometer for Leak Detection Description of Driving Event: On the MIDAS project, leak detection for the v
2、acuum chamber was done both by helium “sniffing“ using the RGA, and also by checking the N2 to O2 ratio to see if there was a steady air leak present. In one case, this ratio did not indicate a leak, because the cryopump was preferentially pumping oxygen.Lesson(s) Learned: It is possible to not have
3、 a 5:1 ratio of N2 to O2 in a vacuum chamber that has a steady leak, when there is a cryopump running that will preferentially pump the O2.Recommendation(s): When using gas ratios based on a quadrupole mass spectrometer (RGA) to tell you if you have a leak (i.e., looking for 5:1 ratio between N2 and