ASTM F2790-2010(2014) Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性的标准实践规程》.pdf
《ASTM F2790-2010(2014) Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性的标准实践规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2790-2010(2014) Standard Practice for Static and Dynamic Characterization of Motion Preserving Lumbar Total Facet Prostheses《活动腰椎全椎间盘假体的静态和动态特性的标准实践规程》.pdf(8页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2790 10 (Reapproved 2014)Standard Practice forStatic and Dynamic Characterization of Motion PreservingLumbar Total Facet Prostheses1This standard is issued under the fixed designation F2790; the number immediately following the designation indicates the year oforiginal adoption or, in
2、the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This practice provides guidance for the static anddynamic testing of Lumbar Total Fa
3、cet Prostheses (FP). Theseimplants are intended to allow motion and lend support to oneor more functional spinal unit(s) through replacement of thenatural facets.1.2 These test methods are intended to provide a basis forthe mechanical comparison among past, present, and futurenon-biologic FP. These
4、test methods allow comparison ofdevices with different methods of application to the lumbarspine. These test methods are intended to enable the user tomechanically compare devices and do not purport to provideperformance standards for them.1.3 These test methods describe static and dynamic tests bys
5、pecifying load types and specific methods of applying theseloads.1.4 These test methods do not purport to address all clini-cally relevant failure modes for FP, some of which will bedevice specific. For example, these test methods do not addressimplant wear resistance under expected in vivo loads an
6、dmotions. In addition, the biologic response to wear debris is notaddressed in these test methods.1.5 Requirements are established for measuring displace-ments and evaluating the stiffness of FP.1.6 Some devices may not be testable in all test configura-tions.1.7 The values stated in SI units are to
7、 be regarded as thestandard with the exception of angular measurements, whichmay be reported in terms of either degrees or radians.1.8 The values stated in inch-pound units are to be regardedas standard. The values given in parentheses are mathematicalconversions to SI units that are provided for in
8、formation onlyand are not considered standard.2. Referenced Documents2.1 ASTM Standards:2D638 Test Method for Tensile Properties of PlasticsE4 Practices for Force Verification of Testing MachinesE6 Terminology Relating to Methods of Mechanical TestingE468 Practice for Presentation of Constant Amplit
9、ude Fa-tigue Test Results for Metallic MaterialsE739 Practice for StatisticalAnalysis of Linear or LinearizedStress-Life (S-N) and Strain-Life (-N) Fatigue DataF1582 Terminology Relating to Spinal Implants3. Terminology3.1 All functional and kinematic testing terminology isconsistent with the refere
10、nced standards (including Teminol-ogy E6 and Terminology F1582), unless otherwise stated.3.2 Definitions:3.2.1 coordinate systems/axesGlobal XYZ orthogonal axesare defined following a right-handed Cartesian coordinatesystem in which the XY plane is parallel to and co-planar withthe superior endplate
11、 of the inferior vertebral body. Alternativecoordinate systems may be used with justification. The globalaxes are fixed relative to the inferior vertebral body. Lowercase letters, xyz, denote a local moving orthogonal coordinatesystem attached to the superior vertebral body with directionsinitially
12、coincident with those of the global XYZ axes, respec-tively. The 3D motion of the superior relative to inferiorvertebra is specified and is to be measured in terms ofsequential Eulerian angular rotations about the xyz axes,respectively (z axial rotation, x lateral bend, and y flexion-extension).3.2.
13、1.1 origincenter of the global coordinate system thatis located at the posterior medial position on the superiorendplate of the inferior vertebral body.3.2.1.2 X-axispositive X-axis is to be directed anteriorlyrelative to the specimens initial unloaded position.3.2.1.3 Y-axispositive Y-axis is direc
14、ted laterally (towardthe left) relative to the specimens initial unloaded position.1This practice is under the jurisdiction ofASTM Committee F04 on Medical andSurgical Materials and Devices and is the direct responsibility of SubcommitteeF04.25 on Spinal Devices.Current edition approved Nov. 1, 2014
15、. Published November 2014. Originallyapproved in 2010. Last previous edition approved in 2010 as F2790-2010. DOI:10.1520/F279010R14.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume inform
16、ation, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States13.2.1.4 Z-axispositive Z-axis is to be directed superiorlyrelative to the specimens initial unloaded position.3.2.2
17、 failurefunctional failure or substantial mechanicalfailure.3.2.2.1 functional failurepermanent deformation resultingfrom fracture, plastic deformation, or loosening beyond theultimate displacement or loosening that renders the spinalimplant assembly ineffective or unable to adequately resistload.3.
18、2.2.2 mechanical failurefailure associated with a defectin the material (for example, fatigue crack) or of the bondingbetween materials that may or may not produce functionalfailure.3.2.3 fatigue lifethe number of cycles, N, that the FP cansustain at a particular load or moment before failure occurs
19、.3.2.4 intended method of applicationa FP may containdifferent types of features to stabilize the implant-tissue inter-face such as threads, spikes, and textured surfaces. Each typeof feature has an intended method of application or attachmentto the spine.3.2.5 insertion point of an anchorthe locati
20、on where theanchor is attached to the test block. The insertion points shownin Fig. 1 are to be adhered to if possible. In situations where thedesign of the spinal implant assembly or the manufacturerssurgical instructions for installation dictate otherwise, theattachment points may deviate from the
21、se dimensions.3.2.6 longitudinal directionthe initial spatial orientationbetween the insertion points in the superior test blocks and theinferior test blocks.3.2.7 maximum run-out load or momentthe maximumload or moment for a given test that can be applied to a FPwhere all of the tested constructs h
22、ave withstood 10 000 000cycles without failure.3.2.8 mechanical deteriorationdeterioration that is visibleto the naked eye and is associated with mechanical damage tothe device under test (for example, initiation of fatigue crack orsurface wear).3.2.9 permanent deformationthe remaining linear or an-
23、gular displacement (axialmm, angulardegrees or radians)relative to the initial unloaded condition of the FP after theapplied load or moment has been removed.3.2.10 radius of rotationthe distance between the centerof rotation and the functional position (for example, load-bearing contact point) of th
24、e FP, for a given motion (that is,flexion/extension, lateral bending, or axial rotation).3.2.11 spinal implant assemblya complete spinal implantconfiguration as intended for surgical use. A spinal implantassembly may contain anchors, interconnections, and longitu-dinal elements and may contain trans
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF279020102014STANDARDPRACTICEFORSTATICANDDYNAMICCHARACTERIZATIONOFMOTIONPRESERVINGLUMBARTOTALFACETPROSTHESES

链接地址:http://www.mydoc123.com/p-539475.html