1、Designation: E2719 09 (Reapproved 2014)Standard Guide forFluorescenceInstrument Calibration and Qualification1This standard is issued under the fixed designation E2719; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of la
2、st revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide (1)2lists the available materials and methodsfor each type of calibration or correction for fluorescenceinstr
3、uments (spectral emission correction, wavelengthaccuracy, and so forth) with a general description, the level ofquality, precision and accuracy attainable, limitations, anduseful references given for each entry.1.2 The listed materials and methods are intended for thequalification of fluorometers as
4、 part of complying with regu-latory and other quality assurance/quality control (QA/QC)requirements.1.3 Precision and accuracy or uncertainty are given at a 1 confidence level and are approximated in cases where thesevalues have not been well established.31.4 The values stated in SI units are to be
5、regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and det
6、ermine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:4E131 Terminology Relating to Molecular SpectroscopyE388 Test Method for Wavelength Accuracy and SpectralBandwidth of Fluorescence SpectrometersE578 Test Method for Linearity of Fluorescence Me
7、asuringSystemsE579 Test Method for Limit of Detection of Fluorescence ofQuinine Sulfate in Solution3. Terminology3.1 Definitions(2):3.1.1 absorption coeffcient (), na measure of absorptionof radiant energy from an incident beam as it traverses anabsorbing medium according to Bouguers law, I/Io= e-b,
8、where I and Ioare the transmitted and incident intensities,respectively, and b is the path length of the beam through thesample. E1313.1.1.1 DiscussionNote that transmittance T = I/Ioandabsorbance A = log T.3.1.2 absorptivity (a), nthe absorbance divided by theproduct of the concentration of the sub
9、stance and the samplepathlength, a = A/bc. E1313.1.3 Beer-Lambert law, nrelates the dependence of theabsorbance (A) of a sample on its path length (see absorptioncoeffcient, ) and concentration (c), such that A = abc.3.1.3.1 DiscussionAlso called Beers law or Beer-Lambert-Bouquer law. E1313.1.4 cali
10、brated detector (CD), noptical radiation detectorwhose responsivity as a function of wavelength has beendetermined along with corresponding uncertainties (3).3.1.5 calibrated diffuse reflector (CR), nLambertian re-flector whose reflectance as a function of wavelength has beendetermined along with co
11、rresponding uncertainties (4).3.1.6 calibrated optical radiation source (CS), nopticalradiation source whose radiance as a function of wavelengthhas been determined along with corresponding uncertainties (5,6).3.1.7 calibration, nset of procedures that establishes therelationship between quantities
12、measured on an instrument andthe corresponding values realized by standards.3.1.8 certified reference material (CRM), nmaterial withproperties of interest whose values and corresponding uncer-tainties have been certified by a standardizing group ororganization. E1313.1.9 certified value, nvalue for
13、which the certifying bodyhas the highest confidence in its accuracy in that all known or1This guide is under the jurisdiction of ASTM Committee E13 on MolecularSpectroscopy and Separation Science and is the direct responsibility of Subcom-mittee E13.01 on Ultra-Violet, Visible, and Luminescence Spec
14、troscopy.Current edition approved May 1, 2014. Published June 2014. Originallyapproved in 2009. Last previous edition approved in 2009 as E271909. DOI:10.1520/E2719-09R14.2The boldface numbers in parentheses refer to the list of references at the end ofthis standard.3Certain commercial equipment, in
15、struments, or materials are identified in thisguide to foster understanding. Such identification does not imply recommendationor endorsement by ASTM International nor does it imply that the materials orequipment identified are necessarily the best available for the purpose.4For referenced ASTM stand
16、ards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA
17、 19428-2959. United States1suspected sources of bias have been investigated or accountedfor by the certifying body (7).3.1.10 diffuse scatterer, nmaterial that scatters opticalradiation in multiple directions; this includes diffuse reflectors,which are often Lambertian, and scattering solutions, whi
18、ch arenot Lambertian.3.1.11 fluorescence anisotropy (r), nmeasure of the degreeof polarization of fluorescence, defined as r =(Ill I)/(Ill+2I), where Illand Iare the observed fluorescence intensitieswhen the fluorometers emission polarizer is oriented paralleland perpendicular, respectively, to the
19、direction of the polar-ized excitation.3.1.12 fluorescence band, nregion of a fluorescence spec-trum in which the intensity passes through a maximum, usuallycorresponding to a discrete electronic transition.3.1.13 fluorescence lifetime, nparameter describing thetime decay of the fluorescence intensi
20、ty of a sample compo-nent; if a sample decays by first-order kinetics, this is the timerequired for its fluorescence intensity and corresponding ex-cited state population to decrease to 1/e of its initial value.3.1.14 fluorescence quantum effciency, nratio of the num-ber of fluorescence photons leav
21、ing an emitter to the number ofphotons absorbed.3.1.15 fluorescence quantum yield (), nprobability that amolecule or species will fluoresce once it has absorbed aphoton.3.1.15.1 DiscussionThis quantity is an innate property ofthe species and is typically calculated for a sample as the ratioof the nu
22、mber of molecules that fluoresce to the number ofmolecules that absorbed.3.1.16 flux (or radiant flux or radiant power), nrate ofpropagation of radiant energy typically expressed in Watts.3.1.17 grating equation, nrelationship between the angleof diffraction and wavelength of radiation incident on a
23、 grating,that is, m = d(sin + sin), where d is the groove spacing onthe grating; and are the angles of the incident and diffractedwavefronts, respectively, relative to the grating normal; and mis the diffraction order, which is an integer (8).3.1.18 inner filter effects, ndecrease in the measuredqua
24、ntum efficiency of a sample as a result of significantabsorption of the excitation beam, reabsorption of the emissionof the sample by itself, or both, and this causes the measuredquantum efficiency to be dependent on the absorbance,concentration, and excitation and emission path lengths of thesample
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME271920092014STANDARDGUIDEFORFLUORESCENCEINSTRUMENTCALIBRATIONANDQUALIFICATION 荧光 仪器 校准 鉴定 标准 指南

链接地址:http://www.mydoc123.com/p-531760.html