ASTM E2603-2008 Standard Practice for Calibration of Fixed-Cell Differential Scanning Calorimeters《固定化细胞差式扫描量热仪校准的标准实施规程》.pdf
《ASTM E2603-2008 Standard Practice for Calibration of Fixed-Cell Differential Scanning Calorimeters《固定化细胞差式扫描量热仪校准的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM E2603-2008 Standard Practice for Calibration of Fixed-Cell Differential Scanning Calorimeters《固定化细胞差式扫描量热仪校准的标准实施规程》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 2603 08Standard Practice forCalibration of Fixed-Cell Differential Scanning Calorimeters1This standard is issued under the fixed designation E 2603; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last re
2、vision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 1.1 This practice covers the calibration of fixed-celldifferential scanning calorimeters over the temperature rangefrom 10 to
3、 +120C.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establi
4、sh appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific precau-tionary statements are given in Section 7.2. Referenced Documents2.1 ASTM Standards:2E 473 Terminology Relating to Thermal Analysis and Rhe-ologyE 967 Test Method for
5、Temperature Calibration of Differ-ential Scanning Calorimeters and Differential ThermalAnalyzersE 968 Practice for Heat Flow Calibration of DifferentialScanning CalorimetersE 1142 Terminology Relating to Thermophysical Properties3. Terminology3.1 Specific technical terms used in this practice are de
6、finedin Terminologies E 473 and E 1142.4. Summary of Practice4.1 This practice covers calibration of fixed-cell differentialscanning calorimeters. These calorimeters differ from anothercategory of differential scanning calorimeter in that the formerhave generally larger sample volumes, slower maximu
7、m tem-perature scan rate capabilities, provision for electrical calibra-tion of heat flow, and a smaller range of temperature overwhich they operate. The larger sample cells, and their lack ofdisposability, make inapplicable the calibration methods ofPractices E 967 and E 968.4.2 This practice consi
8、sts of heating the calibration mate-rials in aqueous solution at a controlled rate through a region ofknown thermal transition. The difference in heat flow betweenthe calibration material and a reference material, both relativeto a heat reservoir, is monitored and continuously recorded. Atransition
9、is marked by the absorption or release of energy bythe specimen resulting in a corresponding peak in the resultingcurve.4.3 The fixed-cell calorimeters typically, if not always, haveelectrical heating facilities for calibration of the heat-flow axis.Despite the use of resistance heating for calibrat
10、ion, a chemicalcalibration serves to verify the correct operation of the calibra-tion mechanism and the calorimeter. The thermal denaturationof chicken egg white lysozyme is used in this practice forverification of the proper functioning of the instrumentssystems. The accuracy with which the denatur
11、ation enthalpy ofchicken egg white lysozyme is currently known, 65%, is suchthat it should be rare that a calorimeter provides a value outsidethat established in the literature for this reference material.5. Significance and Use5.1 Fixed-cell differential scanning calorimeters are used todetermine t
12、he transition temperatures and energetics of mate-rials in solution. For this information to be accepted withconfidence in an absolute sense, temperature and heat calibra-tion of the apparatus or comparison of the resulting data to thatof known standard materials is required.5.2 This practice is use
13、ful in calibrating the temperature andheat flow axes of fixed-cell differential scanning calorimeters.6. Apparatus6.1 Apparatus shall be:6.1.1 Differential Scanning Calorimeter (DSC), capable ofheating a test specimen and a reference material at a controlledrate and of automatically recording the di
14、fferential heat flowbetween the sample and the reference material to the requiredsensitivity and precision.1This practice is under the jurisdiction of ASTM Committee E37 on ThermalMeasurements and is the direct responsibility of Subcommittee E37.09 on Biologi-cal Calorimetry.Current edition approved
15、 Sept. 1, 2008. Published October 2008.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM Intern
16、ational, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.1.2 DSC Test Chamber, composed of:6.1.2.1 A device(s) to provide uniform controlled heating orcooling of a specimen and reference to a constant temperatureor at a constant rate within the applicable temper
17、ature range ofthis method.6.1.2.2 A temperature sensor to provide an indication of thespecimen temperature to 60.01 K.6.1.2.3 Differential sensors to detect a heat flow (power)difference between the specimen and reference with a sensi-tivity of 60.1 W.6.1.3 A temperature controller, capable of execu
18、ting aspecific temperature program by operating the furnace(s)between selected temperature limits at a rate of temperaturechange of 0.01 K/min to 1 K/min constant to 60.001 K/min orat an isothermal temperature constant to 60.001 K.6.1.4 A data collection device, to provide a means ofacquiring, stori
19、ng, and displaying measured or calculatedsignals, or both. The minimum output signals required for DSCare heat flow, temperature, and time.6.1.5 Containers, that are inert to the specimen and refer-ence materials and that are of suitable structural shape andintegrity to contain the specimen and refe
20、rence in accordancewith the specific requirements of this test method. Thesecontainers are not designed as consumables. They are either anintegral part of the instrument, whether or not user-removablefor replacement or, in some implementations, are removableand reusable. Container volumes generally
21、range from 0.1 mlto 1 ml, depending on the instruments manufacture.6.2 Analytical Balance, capable of weighing to the nearest0.1 mg, for preparation of solutions.6.3 UV spectrophotometer or UV/Vis spectrophotometer,capable of scanning the UV spectrum in a region about 280 nm.6.4 Reagents:6.4.1 Phosp
22、hatidylcholines, 1,2-ditridecanoyl-sn-glycero-3-phosphocholine (DTPC) CAS Number 71242-28-9 and 1,2-ditetracosanoyl-sn-glycero-3-phosphocholine (DLPC) CASNumber 91742-11-9 are the minimum required.6.4.2 Aqueous buffer solutions, 0.01 Molar, pH 7 aqueoussolution of Na2HPO4 NaH2PO4 and 0.1 Molar, pH (
23、2.4 60.1) aqueous solution of HCl + glycine.6.4.3 Chicken egg white lysozyme .7. Precautions7.1 This practice assumes linear temperature indication.Care must be taken in the application of this practice to ensurethat calibration points are taken sufficiently close together sothat linear temperature
24、indication may be approximated.8. Calibration Materials8.1 Phosphatidylcholines: 1,2-ditridecanoyl-sn-glycero-3-phosphocholine (DTPC) CAS Number 71242-28-9; and 1,2-ditetracosanoyl-sn-glycero-3-phosphocholine (DLPC) CASNumber 91742-11-9. Purities are to be 0.99 or better. Addi-tional calibration mat
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME26032008STANDARDPRACTICEFORCALIBRATIONOFFIXEDCELLDIFFERENTIALSCANNINGCALORIMETERS 固定 细胞 扫描 量热仪 校准

链接地址:http://www.mydoc123.com/p-531527.html