ASTM E1897-1997(2002)e1 Standard Test Methods for Measuring and Compensating for Transmittance of an Attenuating Medium Using Infrared Imaging Radiometers《用红外成像辐射仪测量和均衡衰减介质的传输性的标准试.pdf
《ASTM E1897-1997(2002)e1 Standard Test Methods for Measuring and Compensating for Transmittance of an Attenuating Medium Using Infrared Imaging Radiometers《用红外成像辐射仪测量和均衡衰减介质的传输性的标准试.pdf》由会员分享,可在线阅读,更多相关《ASTM E1897-1997(2002)e1 Standard Test Methods for Measuring and Compensating for Transmittance of an Attenuating Medium Using Infrared Imaging Radiometers《用红外成像辐射仪测量和均衡衰减介质的传输性的标准试.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1897 97 (Reapproved 2002)e1Standard Test Methods forMeasuring and Compensating for Transmittance of anAttenuating Medium Using Infrared Imaging Radiometers1This standard is issued under the fixed designation E 1897; the number immediately following the designation indicates the year o
2、foriginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTERemoved Footnote 2 in August 2002 since the organization is no
3、longer in business.1. Scope1.1 These test methods cover procedures for measuring andcompensating for transmittance when using an infrared imag-ing radiometer to measure the temperature of a specimenthrough an attenuating medium, such as a window, filter oratmosphere.1.2 The values stated in SI units
4、 are to be regarded as thestandard.1.3 These test methods may involve use of equipment andmaterials in the presence of heated or electrically-energizedequipment, or both.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility
5、of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:E 1316 Terminology for Nondestructive Examinations23. Terminology3.1 Definitions of Terms Specific to T
6、his Standard:3.1.1 attenuating mediuma semi-transparent solid, liquidor gas, such as a window, filter, external optics and/or anatmosphere that attenuates radiation.3.1.2 blackbody simulatora device with an emissivityclose to 1.00 that can be heated or cooled to a stabletemperature.3.1.3 filtera sem
7、i-transparent material that attenuates cer-tain wavelengths of radiation.3.1.4 infrared thermographerthe person using an infraredimaging radiometer.3.1.5 reflected temperaturethe temperature of the energyincident upon and reflected by the measurement surface of thespecimen.3.1.6 windowa semi-transpa
8、rent material that separatesconditioned and unconditioned atmospheres and attenuatescertain wavelengths of radiation.3.2 See also Terminology E 1316.4. Summary of Test Method4.1 Using the computer built into an infrared imagingradiometer, a method is given for measuring the transmittanceof an attenu
9、ating medium.4.2 Using the computer built into an infrared imagingradiometer, a method is given for compensating for errorswhen measuring the temperature of a specimen through anattenuating medium when the emissivity of the specimen andthe transmittance of the attenuating medium are known.4.3 Using
10、the computer built into an infrared imagingradiometer, a method is given for measuring and compensatingfor unknown transmittance and emissivity errors when thespecimen temperature is known.5. Significance and Use5.1 The transmittance of an attenuating medium can causeerrors for an infrared thermogra
11、pher using an infrared imagingradiometer to measure the temperature of a specimen throughthe medium. Three test methods are given for measuring andcompensating for this error source.5.1.1 A test method is given for measuring the transmittanceof an attenuating medium.5.1.2 A test method is given for
12、compensating for errorswhen measuring the temperature of a specimen having aknown emissivity through an attenuating medium with aknown transmittance.5.1.3 A test method is given for measuring and compensat-ing for transmittance and emissivity errors when the specimentemperature is known.5.2 These te
13、st methods can be used in the field or laboratoryusing commonly available materials.1These test methods are under the jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct responsibility of Subcommittee E07.10 onEmerging NDT Methods.Current edition approved July 10, 1997. Pub
14、lished December 1997.2Annual Book of ASTM Standards, Vol 03.03.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.5.3 These test methods can be used with any infraredradiometers that have the required computer capabilities.6. Interferen
15、ces6.1 Test Method for Measuring the Transmittance of anAttenuating Medium:6.1.1 This test method requires a blackbody simulator withan emissivity of 0.95 or greater that is at least 20C warmerthan ambient temperature. Potential errors can be minimizedby ensuring the stability of the temperature dif
16、ference betweenthe source and the ambient temperature during the test. Also,the transmittance measurement accuracy can be increased byincreasing this temperature difference.6.1.2 Errors can be minimized by ensuring that the tempera-ture of the attenuating medium is as close as possible to thereflect
17、ed temperature incident upon the specimen.6.1.3 The compositions and thicknesses of attenuating me-dia can vary within the same specimen. Errors can be mini-mized by measuring through the same portion of the specimenevery time.6.2 Test Method for Compensating for Unknown ErrorsWhen Measuring the Tem
18、perature of a Specimen With a KnownEmissivity Through an Attenuating Medium Having a KnownTransmittance:6.2.1 Errors can be minimized by ensuring that the tempera-ture of the attenuating medium is as close as possible to thereflected temperature incident upon the specimen.6.3 The transmittance of an
19、 attenuating medium may bespecific to the temperature of the medium and the spectralwaveband of the radiometer used to make the measurement.Therefore, the temperature of the measured specimen and thespectral waveband of the radiometer used should be noted withthe measured transmittance value.6.4 The
20、 use of these test methods to compensate for thetransmittance of an attenuating medium is valid only whenmeasuring the temperatures of specimens that are opaque in thewaveband of the infrared imaging radiometer.7. Apparatus7.1 A calibrated infrared imaging radiometer with a built-incomputer that all
21、ows the infrared thermographer to inputreflected temperatures and emissivity values.7.2 The test method for measuring the transmittance of anattenuating medium requires a tripod or device to support theinfrared imaging radiometer.7.3 The test method for measuring the transmittance of anattenuating m
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME189719972002E1STANDARDTESTMETHODSFORMEASURINGANDCOMPENSATINGFORTRANSMITTANCEOFANATTENUATINGMEDIUMUSINGINFRAREDIMAGINGRADIOMETERS

链接地址:http://www.mydoc123.com/p-529751.html