1、Designation: E1292 94 (Reapproved 2016)Standard Specification forGravity Convection and Forced Ventilation Incubators1This standard is issued under the fixed designation E1292; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This specification covers the performance requirementsfor general purpose air incubators ordinarily used for inc
3、ubat-ing procedures, which have an incubating chamber up to 0.6m3(25 ft3) in volume. It is applicable to gravity and forcedventilation incubators designed to operate over all or part of thetemperature range from 5C above ambient to 75C.1.2 This specification does not include any requirements forthe
4、safe handling of harmful or disease bearing organisms.1.3 The following precautionary caveat pertains only to thetest method portions, Sections 4, 5, and 6, of this specification.This standard does not purport to address all of the safetyconcerns, if any, associated with its use. It is the responsib
5、ilityof the user of this standard to establish appropriate safety andhealth practices and determine the applicability of regulatorylimitations prior to use.2. Types2.1 This specification covers the following four types of airincubators:2.1.1 Type 1AAn incubator ventilated by gravity convec-tion havi
6、ng a uniformity of temperature within 3 % of thedifferential between incubator and ambient temperature,2.1.2 Type 1BAn incubator ventilated by gravity convec-tion having a uniformity of temperature within 6 % of thedifferential between incubator and ambient temperature,2.1.3 Type 2AAn incubator vent
7、ilated by forced ventila-tion having a uniformity of temperature within 2 % of thedifferential between incubator and ambient temperature,2.1.4 Type 2BAn incubator ventilated by forced ventila-tion having a uniformity of temperature within 5 % of thedifferential between incubator and ambient temperat
8、ure.3. Performance Requirements3.1 The temperature within the incubating chamber shall becontrolled by an automatic device, and shall be uniform withinthe tolerances given in Table 1, when tested in accordance withSection 4.3.2 The rate of ventilation of the incubating chamber shallconform to the re
9、quirements specified in Table 1 for theparticular type of incubator when measured in accordance withthe procedure given in Section 5.3.3 The recovery time of the incubating chamber shallconform to the requirements specified in Table 1 for theparticular type of incubator, when tested in accordance wi
10、th theprocedure specified in Section 6.4. Test Methods4.1 Temperature Uniformity:4.1.1 Place nine calibrated thermocouples (Note 1) madefrom iron-constantan or copper-constantan wires approxi-mately 0.5 mm in diameter (No. 24 gauge) and having ajunction size of not more than 2 mm (0.08 in.) in the e
11、mptyincubating chamber with shelves in place, and vents open.Locate one thermocouple in each of the eight corners of theincubator approximately 5 cm (2 in.) from each wall, and placethe ninth thermocouple within 2.5 cm (1 in.) of the geometriccenter of the chamber. A minimum length of 30 cm (12 in.)
12、 ofwire for each thermocouple shall be inside the incubator tominimize the conduction of heat from the thermocouple.NOTE 1If calibrated thermocouples are not available, nine thermo-couples made from the same spool of wire may be used, provided theygive the same value for temperature when placed adja
13、cent to one anotherin the incubating chamber at the temperature of the test.4.1.2 Bring the incubator to the specified temperature andallow it to reach a steady state (Note 2). Record the tempera-ture of the nine thermocouples for a period of at least 12 h anddetermine from the record the maximum de
14、viation of eachpoint from the desired temperature. The ambient room tem-perature shall vary by not more than a total of 5C and the linevoltage shall not vary by more than a total of 5 % during thetest.NOTE 2Some incubators may require as much as 12 h to reach asteady state. If a steady state does no
15、t exist, there is a drift in thetemperature toward the steady state condition.5. Rate of Ventilation5.1 Seal the ventilation ports, doors, and all apertures of theincubator with adhesive tape, or by other means to prevent anyair from passing through the incubator (Note 3). Connect a Whmeter with the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME129219942016STANDARDSPECIFICATIONFORGRAVITYCONVECTIONANDFORCEDVENTILATIONINCUBATORS 重力 对流 强制 通风

链接地址:http://www.mydoc123.com/p-528341.html