《ASTM E1277-2002 Standard Test Method for Chemical Analysis of Zinc-5% Aluminum-Mischmetal Alloys by ICP Emission Spectrometry《用ICP发射光谱测定法对锌-5%铝-铈合金作化学分析的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1277-2002 Standard Test Method for Chemical Analysis of Zinc-5% Aluminum-Mischmetal Alloys by ICP Emission Spectrometry《用ICP发射光谱测定法对锌-5%铝-铈合金作化学分析的标准试验方法》.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1277 02Standard Test Method forChemical Analysis of Zinc-5 % Aluminum-Mischmetal Alloysby ICP Emission Spectrometry1This standard is issued under the fixed designation E 1277; the number immediately following the designation indicates the year oforiginal adoption or, in the case of re
2、vision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the chemical analysis of zincalloys having chemical compositions within t
3、he followinglimits:Element Concentration Range, %Aluminum 3.08.0Antimony 0.002 maxCadmium 0.025 maxCerium 0.030.10Copper 0.10 maxIron 0.10 maxLanthanum 0.030.10Lead 0.026 maxMagnesium 0.05 maxSilicon 0.015 maxTin 0.002 maxTitanium 0.02 maxZirconium 0.02 max1.2 Included are procedures for elements in
4、 the followingconcentration ranges:Element Concentration Range, %Aluminum 3.08.0Cadmium 0.00160.025Cerium 0.0050.10Iron 0.00150.10Lanthanum 0.0090.10Lead 0.0020.0261.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the
5、 user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Specific safetyhazards statements are given in Section 7, 11.2, and 13.1.2. Referenced Documents2.1 ASTM Standards:D 1193 Specification for Reagent Wa
6、ter2E29 Practice for Using Significant Digits in Test Data toDetermine Conformance With Specifications3E50 Practices for Apparatus, Reagents, and Safety Precau-tions for Chemical Analysis of Metals4E55 Practice for Sampling Wrought Nonferrous Metals forDetermination of Chemical Composition4E88 Pract
7、ice for Sampling Nonferrous Metals and Alloysin Cast Form for Determination of Chemical Composition4E 173 Practice for Conducting Interlaboratory Studies ofMethods for Chemical Analysis of Metals4E 876 Practice for Use of Statistics in the Evaluation ofSpectrometric Data5E 1601 Practice for Conducti
8、ng an Interlaboratory Study toEvaluate the Performance of an Analytical Method53. Summary of Test Method3.1 The sample is dissolved in mixed acids. The samplesolution is introduced into the plasma source of an ICPspectrometer and the intensities at selected wavelengths fromthe plasma emission spectr
9、um are compared to the intensities atthe same wavelengths measured with calibration solutions.4. Significance and Use4.1 This test method for the chemical analysis of metals andalloys is primarily intended to test such materials for compli-ance with compositional specifications. It is assumed that a
10、llthose who use this test method will be trained analysts capableof performing common laboratory procedures skillfully andsafely. It is expected that work will be performed in a properlyequipped laboratory.5. Apparatus5.1 Inductively-Coupled Argon Plasma (ICP) Atomic Emis-sion SpectrometerThe instru
11、ment is equipped with an argon-plasma source, a sample transport system for introducing thetest sample and calibration solutions into the plasma. Themonochromator or polychromator must be capable of isolating1This test method is under the jurisdiction of ASTM Committee E01 onAnalytical Chemistry for
12、 Metals, Ores, and Related Materials and is the directresponsibility of Subcommittee E01.05 on Zn, Sn, Pb, Be, and Other Metals.Current edition approved October 10, 2002. Published December 2002. Origi-nally approved in 1991. Last previous edition in 1996 as E 1277 96.2Annual Book of ASTM Standards,
13、 Vol 11.01.3Annual Book of ASTM Standards, Vol 14.02.4Annual Book of ASTM Standards, Vol 3.05.5Annual Book of ASTM Standards, Vol 03.06.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.the required wavelengths shown in Table 1 for mea
14、surement oftheir intensities by a linear photometer. Multielement pro-grammed analysis including automatic data acquisition andcomputer-controlled calibration and analysis calculations maybe used if available, provided that, in addition to calculatedresults, the instrument records intensity readings
15、 each time atest sample or calibration solution is presented to the instru-ment.NOTE 1All elements (including aluminum) are calibrated as linearfunctions of intensity. If the instrument cannot be set to measurealuminum and ignore other elements in calibration solutions No. 1 and No.4, then a separat
16、e determination of aluminum must be made usingcalibration solutions No. 1, No. 2 and No. 4. The other elements can thenbe determined together in another run using only calibration solutions No.2 and No. 3. Use the calibration solutions prepared in 10.1 in determiningthe instrument settings for the e
17、lements in this matrix. Follow themanufacturers instructions to set the wavelengths and parameters toprovide as large a difference between the intensity readings for the highand low calibration concentrations as is consistent with stable instrumentreadings. If there is a question of linearity of the
18、 instruments responseover the range of solution concentrations given, a third standard, equidis-tant between the two listed standards, must be measured to verify linearity.6. Reagents6.1 Purity of ReagentsUnless otherwise indicated, allreagents used in this test method shall conform to the “ReagentG
19、rade” Specifications of the American Chemical Society.6Other chemicals may be used provided that it is first ascer-tained that the reagent used is of sufficiently high purity topermit its use without lessening the accuracy of the determi-nation.6.2 Purity of WaterUnless otherwise indicated, referenc
20、esto water shall be understood to mean reagent water as definedby Type II of Specification D 1193.6.3 Aluminum, Standard Solution (1 mL = 20.0 mg Al)Transfer 2.0000 g of aluminum (purity: 99.999 % min) to a250-mL beaker. Cover, add 50 mL of HCl (1 + 1) and a smallcrystal of mercuric nitrate. Heat ge
21、ntly to accelerate thereaction, but avoid temperatures high enough to cause anoticeable volume loss. If the reaction slows, add moremercuric salt as needed. A number of hours may be required tocomplete the dissolution (only a small droplet of mercury willremain undissolved). Transfer the solution to
22、 a 100-mL volu-metric flask, dilute to volume, and mix. Store in a polyethylenebottle.6.4 Cadmium, Standard Solution (1 mL = 1.00 mg Cd)Transfer 1.000 g of cadmium (purity: 99.95 % min) to a250-mL beaker. Cover and add 40 mL of HNO3(1 + 1) and 10mL of HCl. After dissolution is complete, heat to boil
23、ing toremove oxides of nitrogen. Cool, transfer to a 1-L volumetricflask, add 240 mL of HCl, dilute to volume, and mix. Store ina polyethylene bottle.6.5 Cerium, Standard Solution A (1 mL = 1.00 mg Ce)Dry ceric ammonium nitrate (NH4)2Ce(NO3)6, also known asammonium hexanitrato cerate) (purity: 99.95
24、 % min) for4hat85C and cool to room temperature in a desiccator. Dissolve3.913 g of dry ceric ammonium nitrate in 100 mL of HCl(1 + 9). Transfer to a 1-L volumetric flask, add 240 mL of HCland 20 mL of HNO3, dilute to volume, and mix. Store in apolyethylene bottle.6.6 Cerium, Standard Solution B (1
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME12772002STANDARDTESTMETHODFORCHEMICALANALYSISOFZINC5ALUMINUMMISCHMETALALLOYSBYICPEMISSIONSPECTROMETRY

链接地址:http://www.mydoc123.com/p-528324.html