ASTM E1165-2004 Standard Test Method for Measurement of Focal Spots of Industrial X-Ray Tubes by Pinhole Imaging《用针孔成像法测量工业X射线管焦点的标准试验方法》.pdf
《ASTM E1165-2004 Standard Test Method for Measurement of Focal Spots of Industrial X-Ray Tubes by Pinhole Imaging《用针孔成像法测量工业X射线管焦点的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E1165-2004 Standard Test Method for Measurement of Focal Spots of Industrial X-Ray Tubes by Pinhole Imaging《用针孔成像法测量工业X射线管焦点的标准试验方法》.pdf(7页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E 1165 04Standard Test Method forMeasurement of Focal Spots of Industrial X-Ray Tubes byPinhole Imaging1This standard is issued under the fixed designation E 1165; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method provides instructions for determiningthe length and width dimensions of line focal spots in in
3、dustrialX-ray tubes (see Note 1). This determination is based on themeasurement of an image of a focal spot that has beenradiographically recorded with a “pinhole” projection/imagingtechnique.NOTE 1Line focal spots are associated with vacuum X-ray tubeswhose maximum voltage rating does not generally
4、 exceed 500 kV.1.2 This test method may not yield meaningful results onfocal spots whose nominal size is less than 0.3 mm 0.011 in.(See Note 2.)NOTE 2The X-ray tube manufacturer may be contacted for nominalfocal spot dimensions.1.3 This test method may also be used to determine thepresence or extent
5、 of focal spot damage or deterioration thatmay have occurred due to tube age, tube overloading, and thelike. This would entail the production of a focal spot radio-graph (with the pinhole method) and an evaluation of theresultant image for pitting, cracking, and the like.1.4 Values stated in SI unit
6、s are to be regarded as thestandard. Inch-pound units are provided for information only.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices
7、 and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2E 999 Guide for Controlling the Quality of Industrial Ra-diographic Film Processing3. Terminology3.1 Definitions of Terms Specific to This Standard:3.1.1 actual focal spotthe X-ray pro
8、ducing area of thetarget as viewed from a position perpendicular to the targetsurface (see Fig. 2).3.1.2 effective focal spotthe X-ray producing area of thetarget as viewed from a position perpendicular to the tube axisin the center of the X-ray beam (see Fig. 2).3.1.3 line focal spota focal spot wh
9、ose projected pinholeimage consists primarily of two curved lines (see Fig. 3).4. Significance and Use4.1 One of the factors affecting the quality of a radiographicimage is geometric unsharpness. The degree of geometricunsharpness is dependent upon the focal size of the radiationsource, the distance
10、 between the source and the object to beradiographed, and the distance between the object to beradiographed and the film. This test method allows the user todetermine the focal size of the X-ray source and to use thisresult to establish source to object and object to film distancesappropriate for ma
11、intaining the desired degree of geometricunsharpness.5. Apparatus5.1 Pinhole DiaphragmThe pinhole diaphragm shall con-form to the design and material requirements of Table 1 andFig. 1.1This test method is under the jurisdiction of ASTM Committee E07 onNondestructive Testing and is the direct respons
12、ibility of Subcommittee E07.01 onRadiology (X and Gamma) Method.Current edition approved January 1, 2004. Published February 2004. Originallyapproved in 1987. Last previous edition approved in 2002 as E 1165 92 (2002).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact AS
13、TM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.FIG. 1 Pinhole Diaphragm Design1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United Stat
14、es.5.2 CameraThe pinhole camera assembly consists of thepinhole diaphragm, the shielding material to which it is affixed,and any mechanism that is used to hold the shield/diaphragm inposition (jigs, fixtures, brackets, and the like; see Fig. 4).5.3 FilmIndustrial type extra fine grain. No intensifyi
15、ngscreens are to be used. The film shall be processed inaccordance with Guide E 999.5.4 Image Measurement ApparatusThis apparatus is usedto measure the size of the image of the focal spot. Theapparatus shall be an optical comparator with built-in graticulewith 0.1 mm or .001 in., or both divisions a
16、nd magnification of53 to 103 (or equivalent).6. Procedure6.1 If possible, use a standard 91.44 cm 36 in. focal spotto film plane distance (FFD) for all exposures. If machinegeometry or accessibility limitations will not permit the use ofa 91.44 cm 36 in. FFD, use the maximum attainable FFD (inthese
17、instances adjust the relative distances between focal spot,pinhole, and film accordingly to suit the image enlargementfactors specified in Table 2). The distance between the focalspot and the pinhole is based on the nominal size of the focalspot being measured and the desired degree of image enlarge
18、-ment (see Fig. 5). The specified focal spot to pinhole distance(FHD) for the different nominal focal spot size ranges isprovided in Table 2. Position the pinhole such that it is within61 of the central axis of the X-ray beam. Fig. 6 illustrates atypical focal spot exposure arrangement.NOTE 3The acc
19、uracy of the pinhole system is highly dependent uponthe relative distances between (and alignment of) the focal spot, thepinhole, and the film. Accordingly, specially designed apparatus may beFIG. 2 Actual/Effective Focal SpotNOTE 1During the production of X-rays the electrons are accelerated from t
20、he filament to the target in two separate paths (see Sketch 1). Electronsemitted at the front of the filament travel primarily along Path A, and electrons emitted at the backside of the filament travel primarily along Path B. Notethat these two paths intersect at a certain point; this is the point a
21、t which the target is positioned. As a result, the pinhole picture of the focal spot showstwo lines that correspond with the intersections of Paths A and B at the target (see Sketch 2).FIG. 3 Line Focal SpotE1165042necessary in order to assure compliance with the above requirements. Fig.7 provides a
22、n example of a special collimator that can be used to ensureconformance with the 61 alignment tolerance. Some other standardsimpose very stringent alignment requirements and express these require-ments in terms of radial tolerances. These documents do not, however,address any means for assuring comp
23、liance with such tolerances. In orderto simplify the focal spot radiography technique and to improve theoverall practicality of the procedure, it is considered that a workablealignment tolerance, and a means of assuring conformance with thattolerance, is appropriate. Accordingly, this standard addre
24、sses tolerancesin angular terms and provides a method for assuring compliance withthese tolerances. This provides a practical means of meeting the precisionand bias requirements of Section 9.6.2 Position the film as illustrated in Fig. 6. The exposureidentification appearing on the film (by radiogra
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME11652004STANDARDTESTMETHODFORMEASUREMENTOFFOCALSPOTSOFINDUSTRIALXRAYTUBESBYPINHOLEIMAGING 针孔 成像

链接地址:http://www.mydoc123.com/p-528059.html