ASTM E70-2007(2015) 7831 Standard Test Method for pH of Aqueous Solutions With the Glass Electrode《带玻璃电极水溶液的pH值的标准试验方法》.pdf
《ASTM E70-2007(2015) 7831 Standard Test Method for pH of Aqueous Solutions With the Glass Electrode《带玻璃电极水溶液的pH值的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM E70-2007(2015) 7831 Standard Test Method for pH of Aqueous Solutions With the Glass Electrode《带玻璃电极水溶液的pH值的标准试验方法》.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: E70 07 (Reapproved 2015)Standard Test Method forpH of Aqueous Solutions With the Glass Electrode1This standard is issued under the fixed designation E70; the number immediately following the designation indicates the year of originaladoption or, in the case of revision, the year of last
2、 revision.Anumber in parentheses indicates the year of last reapproval.Asuperscriptepsilon () indicates an editorial change since the last revision or reapproval.This standard has been approved for use by agencies of the U.S. Department of Defense.1. Scope*1.1 This test method specifies the apparatu
3、s and proceduresfor the electrometric measurement of pH values of aqueoussolutions with the glass electrode. It does not deal with themanner in which the solutions are prepared. pH measurementsof good precision can be made in aqueous solutions containinghigh concentrations of electrolytes or water-s
4、oluble organiccompounds, or both. It should be understood, however, that pHmeasurements in such solutions are only a semiquantitativeindication of hydrogen ion concentration or activity. Themeasured pH will yield an accurate result for these quantitiesonly when the composition of the medium matches
5、approxi-mately that of the standard reference solutions. In general, thistest method will not give an accurate measure of hydrogen ionactivity unless the pH lies between 2 and 12 and the concen-tration of neither electrolytes nor nonelectrolytes exceeds 0.1mol/L (M).1.2 The values stated in SI units
6、 are to be regarded asstandard. The values in parentheses are for information only.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and
7、determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM Standards:2D1193 Specification for Reagent WaterE180 Practice for Determining the Precision of ASTMMethods for Analysis and Testing of Industrial and Spe-cialty Chemicals (Withdrawn 2009)3E691 Practic
8、e for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method3. Terminology3.1 Definitions:3.1.1 pHdefined formally as the negative logarithm to thebase 10 of the conventional hydrogen ion activity. See Appen-dix X1.3.2 Definitions of Terms Specific to This Standard:3.2.1 For
9、the purpose of this test method, the term “meter”shall apply to the instrument used for the measurement ofpotential (either in millivolts or in terms of pH units), the term“electrodes” to the glass electrode and the reference electrode,and the term “assembly” to the combination of the meter andthe e
10、lectrodes. The performance of the meter shall be differ-entiated from that of the electrodes.4. Significance and Use4.1 pH is, within the limits described in 1.1, an accuratemeasurement of the hydrogen ion concentration and thus iswidely used for the characterization of aqueous solutions.4.2 pH meas
11、urement is one of the main process controlvariables in the chemical industry and has a prominent place inpollution control.5. Apparatus5.1 pH metersMany excellent pH meters are availablefrom commercial sources. To some extent, the choice of meterwill depend on the desired precision of measurement. T
12、hemeter may operate on a null-detection principle or may utilizedigital readout or a direct deflection meter with a large scale.Power may be supplied by batteries or a-c operation may beprovided. The maximum grid current drawn from the glasselectrode during measurement shall not exceed 2 1012A.Autom
13、atic or manual adjustment shall allow for changes inF/(RT ln 10) when the temperature of the assembly is altered.For referee work, or in case of dispute, meters capable ofdiscriminating changes of pH to 0.01 unit (0.6 mV) or less shallbe used.5.2 Reference Electrodes and Glass Electrodes:5.2.1 The s
14、aturated calomel electrode and the 3.5 mol/L(M)calomel electrode are suitable as reference electrodes in pH1This test method is under the jurisdiction of ASTM Committee E15 onIndustrial and Specialty Chemicals and is the direct responsibility of SubcommitteeE15.01 on General Standards.Current editio
15、n approved June 1, 2015. Published June 2015. Originallyapproved in 1952. Last previous edition approved in 2007 as E70 07. DOI:10.1520/E0070-07R15.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandar
16、ds volume information, refer to the standards Document Summary page onthe ASTM website.3The last approved version of this historical standard is referenced onwww.astm.org.*A Summary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700
17、, West Conshohocken, PA 19428-2959. United States1assemblies (Note 1). If the saturated electrode is used, a fewcrystals of solid potassium chloride shall be present in thechamber surrounding the electrode element at each tempera-ture. The design of the electrode shall permit a fresh liquidjunction
18、between the solution of potassium chloride and thebuffer or test solution to be formed for each test and shall allowtraces of solution to be readily removed by washing.NOTE 1Other reference electrodes of constant potential may be used,provided no difficulty is experienced in standardizing the assemb
19、ly asdescribed in Section 8.5.2.2 The silver-silver chloride electrode also is used widelyas a reference electrode.5.2.3 Commercial glass electrodes are designed for certainspecific ranges of pH and temperature; consequently, the pHand temperature of the test solutions shall be considered inselectin
20、g the glass electrode for use. The pH response shallconform with the requirements set forth in Section 7. The leadsshall be shielded from the effects of body capacitance.5.2.4 If the assembly is in intermittent use, the ends of theelectrodes shall be immersed in distilled water between mea-surements
21、. The high-alkalinity type of glass electrode shall bestored in the borax buffer solution. For prolonged storage, glasselectrodes may be allowed to become dry, and referenceelectrodes shall be capped to prevent undue evaporation.NOTE 2New glass electrodes and those that have been stored dry shallbe
22、conditioned as recommended by the manufacturer. Requirements forthe physical dimensions and shape of the electrodes and the compositionof the internal reference solution are not considered part of this testmethod.6. Reagents and Materials6.1 The pH(S) of six recommended standard solutions atseveral
23、temperatures is listed in Table 1. The buffer solutionsshall be prepared from highly purified materials sold specifi-cally as pH standards (Note 3). Potassium hydrogen phthalateand the two phosphate salts shall be dried at 110C for 1 hbefore use, but borax and sodium bicarbonate shall not beheated a
24、bove room temperature. Potassium dihydrogen citrateshall be dried for1hat80C, and sodium carbonate shall beignited for1hat270C before use. The standard solutionsshall be prepared as described in 6.4 6.9. They shall bepreserved in bottles of chemically resistant glass or polyethyl-ene and shall be re
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTME70200720157831STANDARDTESTMETHODFORPHOFAQUEOUSSOLUTIONSWITHTHEGLASSELECTRODE 玻璃 电极 水溶液 PH 标准 试验

链接地址:http://www.mydoc123.com/p-527610.html