《ASTM D7883-2013 6250 Standard Test Method for Determination of 4-Carboxybenzaldehyde and p-Toluic Acid in Purified Terephthalic Acid by Weak Anion Exchange High Performance Liquid .pdf》由会员分享,可在线阅读,更多相关《ASTM D7883-2013 6250 Standard Test Method for Determination of 4-Carboxybenzaldehyde and p-Toluic Acid in Purified Terephthalic Acid by Weak Anion Exchange High Performance Liquid .pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7883 13Standard Test Method forDetermination of 4-Carboxybenzaldehyde and p-Toluic Acidin Purified Terephthalic Acid by Weak Anion Exchange HighPerformance Liquid Chromatography1This standard is issued under the fixed designation D7883; the number immediately following the designation
2、indicates the year oforiginal adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method2covers the determinat
3、ion of the4-Carboxybenzaldehyde (4-CBA) and p-Toluic acid (p-TOL)in purified terephthalic acid (PTA) by weak anion exchangehigh performance liquid chromatography (HPLC). Thismethod is applicable for 4-CBA from 2 to 500 mg/kg and forp-TOL from 10 to 500 mg/kg, respectively.1.2 In determining the conf
4、ormance of the test results usingthis method to applicable specification, results shall be roundedoff in accordance with the rounding-off method of PracticeE29.1.3 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.4 This standar
5、d does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use.2. Referenced Documents2.1 ASTM
6、 Standards:3D1193 Specification for Reagent WaterD4790 Terminology of Aromatic Hydrocarbons and RelatedChemicalsD6809 Guide for Quality Control and Quality AssuranceProcedures for Aromatic Hydrocarbons and Related Ma-terialsE29 Practice for Using Significant Digits in Test Data toDetermine Conforman
7、ce with SpecificationsE177 Practice for Use of the Terms Precision and Bias inASTM Test MethodsE682 Practice for Liquid Chromatography Terms and Rela-tionshipsE691 Practice for Conducting an Interlaboratory Study toDetermine the Precision of a Test Method2.2 ISO Document:4EN ISO 8213 Chemical produc
8、ts for industrial useSampling techniquesSolid chemical products in theform of particles varying from powders to coarse lumps2.3 Other Document:5OSHA Regulations, 29 CFR paragraphs 1910.1000 and1910.12003. Summary of Test Method3.1 Weak Anion Exchange HPLC MethodPTA sample isdissolved in ammonium hyd
9、roxide solution. After pHadjustment, a fixed volume of this solution is injected into ahigh performance liquid chromatograph equipped with a UVdetector. An anion-exchange column is used to separate theimpurities 4-CBA and p-TOL from PTA. The external standardcalibration is used for quantification.4.
10、 Significance and Use4.1 The presence of 4-CBA and p-TOL in PTA used for theproduction of polyester is undesirable because they can slowdown the polymerization process, and 4-CBA is also impartingcoloration to the polymer due to thermal instability.4.2 Determining the amount of 4-CBA and p-TOL remai
11、n-ing from the manufacture of PTA is often required. This testmethod is suitable for setting specifications and could be usedas an internal quality control tool where these products areproduced or are used.1This test method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and
12、Related Chemicals and is the direct responsibility ofSubcommittee D16.02 on Oxygenated Aromatics.Current edition approved Aug. 1, 2013. Published October 2013. DOI: 10.1520/D7883-13.2This standard is based on SH/T 1612.7-1995 Purified terephthalic acid forindustrial usedetermination of p-Toluic Acid
13、, 4-Carboxybenzaldehyde-HPLC,copyright SINOPEC, 22 Chaoyangmen North Street, Chaoyang District, Beijing,China 100728. A copy of SH/T 1612.7-1995 may be obtained from ChinaPetrochemical Press, www.sinopec-.3For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer S
14、ervice at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.4Available from American National Standards Institute (ANSI), 25 W. 43rd St.,4th Floor, New York, NY 10036, http:/www.ansi.org.5Available from U.S. Governme
15、nt Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:/www.access.gpo.gov.Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15. Apparatus5.1 High Performance Liquid Chromatograp
16、h (HPLC)any HPLC capable of pumping the mobile phase at flow ratesbetween 0.1 and 2.0 mL/min, with a pressure between 0 and 40MPa and a pulsation of less than 1 % full scale deflection underthe test conditions described in Table 1. The S/N (signal tonoise) ratio should be 3:1 or greater for 2 mg/kg
17、4-CBAand 10mg/kg p-TOL.5.2 Sample Injection Systemcapable of injecting 1 to 25L, using either partial or full loop mode, with a repeatabilityof 61%.5.3 Detector, Variable Wavelength Ultraviolet PhotometricDetector (VWD), Multi-wavelength Detector, or PhotometricDiode Array Detector (PDA), capable of
18、 operating at 236 and258 nm.5.4 Column Ovenany suitable HPLC column oven (blockheating or air circulating) capable of maintaining a constanttemperature of 61C within the range of 20 to 70C.5.5 Chromatography Data System.5.6 HPLC Columns:5.6.1 Guard Columna stainless steel column placed infront of th
19、e analytical column is recommended. A column,packed with the same stationary phase as the analyticalcolumn, 3 to 5 mm ID and 50 to 100 mm length, has beenfound to be satisfactory. Other hydrophilic chemically-bondedsilica stationary phases can also be used.5.6.2 Analytical Columna stainless steel HP
20、LC columnpacked with amino-bonded silica stationary phase is suitable.See Table 1 for recommended operating conditions.5.7 Analytical Balancereadable to 60.0001 g.5.8 pH Meter.5.9 Sample Filtera disposable syringe filter made ofcellulose acetate, with a pore size between 0.22 and 0.45 m,and is chemi
21、cally inert to aqueous solutions, is recommendedfor the removal of particulate matter from the sample solution.5.10 Vacuum Filtercapable of filtering mobile phase.6. Reagents and Materials6.1 Purity of ReagentsUnless otherwise indicated, it isintended that all reagents shall conform to the reagent g
22、radespecification for analytical reagents of the American ChemicalSociety, where such specifications are available.6Other gradesmay be used, provided it is first ascertained that the reagent isof sufficiently high purity to permit its use without lesseningthe performance or accuracy of the determina
23、tion. Reagentchemicals shall be used for all tests.NOTE 1Calibration and detection limits of this test method can bebiased by the purity of the reagents.6.2 Ammonium Dihydrogen Phosphate(WarningAmmonium dihydrogen phosphate may cause irritation withonly minor residual injury.)6.3 Ammonium Hydroxide,
24、 25 to 28 % (wt).6.4 Phosphoric Acid, 82 % (wt).6.5 AcetonitrileHPLC grade. (WarningAcetonitrile isflammable and hazardous in case of skin and eye contact,ingestion or inhalation.)6.6 MethanolHPLC grade. (WarningMethanol ishighly flammable and toxic by inhalation, ingestion or skincontact.)6.7 Water
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD788320136250STANDARDTESTMETHODFORDETERMINATIONOF4CARBOXYBENZALDEHYDEANDPTOLUICACIDINPURIFIEDTEREPHTHALICACIDBYWEAKANIONEXCHANGEHIGHPERFORMANCELIQUIDPDF

链接地址:http://www.mydoc123.com/p-526475.html