ASTM D7725-2012 5625 Standard Test Method for the Continuous Measurement of Turbidity Above 1 Turbidity Unit (TU)《混浊度大于1浊度单位(TU)连续测量的标准试验方法 n》.pdf
《ASTM D7725-2012 5625 Standard Test Method for the Continuous Measurement of Turbidity Above 1 Turbidity Unit (TU)《混浊度大于1浊度单位(TU)连续测量的标准试验方法 n》.pdf》由会员分享,可在线阅读,更多相关《ASTM D7725-2012 5625 Standard Test Method for the Continuous Measurement of Turbidity Above 1 Turbidity Unit (TU)《混浊度大于1浊度单位(TU)连续测量的标准试验方法 n》.pdf(24页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7725 12Standard Test Method for theContinuous Measurement of Turbidity Above 1 TurbidityUnit (TU)1This standard is issued under the fixed designation D7725; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of
2、last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers the online and inline determi-nation of high-level turbidity in water that is greater than 1.0
3、turbidity units (TU) in municipal, industrial and environmentalusage.1.2 In principle there are three basic applications for on-linemeasurement set ups. This first is the slipstream (bypass)sample technique. For the slipstream sample technique aportion of sample is transported out of the process and
4、 throughthe measurement apparatus. It is then either transported back tothe process or to waste. The second is the in-line measurementwhere the sensor is brought directly into the process (seeFigure 8). The third basic method is for in-situ monitoring ofsample waters. This principle is based on the
5、insertion of asensor into the sample itself as the sample is being processed.The in-situ use in this method is intended for the monitoring ofwater during any step within a processing train, includingimmediately before or after the process itself.1.3 This test method is applicable to the measurement
6、ofturbidities greater than 1.0 turbidity unit (TU). The absoluterange is dictated by the technology that is employed.1.4 The upper end of the measurement range is left unde-fined because different technologies described in this methodcan cover very different ranges of turbidity.1.5 Many of the turbi
7、dity units and instrument designscovered in this method are numerically equivalent in calibra-tion when a common calibration standard is applied acrossthose designs listed in Table 1. Measurement of a commoncalibration standard of a defined value will also produceequivalent results across these tech
8、nologies. This methodprescribes the assignment of a determined turbidity values tothe technology used to determine those values. Numericalequivalence to turbidity standards is observed between differ-ent technologies but is not expected across a common sample.Improved traceability beyond the scope o
9、f this method may bepracticed and would include the listing of the make and modelnumber of the instrument used to determine the turbidityvalues.1.5.1 In this method, calibration standards are often definedin NTU values, but the other assigned turbidity units, such asthose in Table 1 are equivalent.
10、For example,a1NTUformazin standard is alsoa1FNU,a1FAU,a1BU,andsoforth.1.6 This standard does not purport to cover all availabletechnologies for high-level turbidity measurement.1.7 This test method was tested on different waters, and withstandards that will serve as surrogates to samples. It is theu
11、sers responsibility to ensure the validity of this test methodfor waters of untested matrices.1.8 Those samples with the highest particle densities typi-cally prove to be the most difficult to measure. In these cases,the process monitoring method can be considered with ad-equate measurement protocol
12、s installed.1.9 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. Re
13、fer to the MSDSsfor all chemicals used in this procedure.2. Referenced Documents2.1 ASTM Standards:2D1129 Terminology Relating to WaterD2777 Practice for Determination of Precision and Bias ofApplicable Test Methods of Committee D19 on WaterD3370 Practices for Sampling Water from Closed ConduitsD386
14、4 Guide for Continual On-Line Monitoring Systemsfor Water AnalysisD6698 Test Method for On-Line Measurement of TurbidityBelow 5 NTU in WaterD7315 Test Method for Determination of TurbidityAbove 11This test method is under the jurisdiction of ASTM Committee D19 on Waterand is the direct responsibilit
15、y of Subcommittee D19.03 on Sampling Water andWater-Formed Deposits, Analysis of Water for Power Generation and Process Use,On-Line Water Analysis, and Surveillance of Water.Current edition approved Jan. 1, 2012. Published June 2012. DOI: 10.1520/D772512.2For referenced ASTM standards, visit the AST
16、M website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, Unite
17、d States.Turbidity Unit (TU) in Static Mode3. Terminology3.1 DefinitionsFor definitions of terms used in thismethod refer to Terminology D1129.3.2 Definitions of Terms Specific to This Standard:3.2.1 Calibration turbidity standard, nA turbidity stan-dard that is traceable and equivalent to the refer
18、ence turbiditystandard to within statistical errors; calibration turbidity stan-dards include commercially prepared 4000 NTU Formazin,stabilized formazin, and styrenedivinylbenzene (SDVB). Dis-cussion - these standards may be used to calibrate the instru-ment. Calibration turbidity standards may be
19、instrument spe-cific.3.2.2 calibration-verification standards, nDefined stan-dards used to verify the accuracy of a calibration in theTABLE 1 Technologies for Measuring Turbidity Greater Than 1 TU that can be used for In-Line or On-Line ApplicationsDesign and Reporting Unit Prominent Application Key
20、 Design Features Typical Instrument Range Suggested ApplicationNephelometric non- ratio(NTU)White light turbidimetersComply with EPA 180.1 for lowlevel turbidity monitoring.Detector centered at 90 degreesrelative to the incident lightbeam. Uses a white light spectralsource.0.012 to 40 NTU Regulatory
21、 reporting of cleanwaterRatio White Light tur bidimeters(NTRU)Complies with ISWTRregulations and Standard Method2130B. Can be used for bothlow and high level measurement.Used a white light spectralsource. Primary detectorcentered at 90. Other detectorslocated at other angles. Aninstrument algorithm
22、uses acombination of detector readingsto generate the turbidity reading.0.012 - 10,000 NTRU Regulatory Reporting of cleanwaterFormazin Nephelometric, near-IR turbidimeters, non-ratiometric (FNU)Complies with ISO 7027. Thewavelengthis less susceptible tocolor interferences. Applicablefor samples with
23、 color and goodfor low level monitoring.Detector centered at 90 degreesrelative to the incident lightbeam. Uses a near-IR (780-900nm) monochromatic light source.0.012 - 1,000 FNU 0 - 40 FNU ISO 7027Regulatory reportingFormazin Nephelometric near-IRturbidimeters, ratiometric(FNRU)Complies with ISO 70
24、27.Applicable for samples with highlevels of color and for monitoringto high turbidity levels.Uses a near-IR monochromaticlight source (780-900 nm).Primary detector centered at 90.Other detectors located at otherangles. An instrument algorithmuses a combination of detectorreadings to generate the tu
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD772520125625STANDARDTESTMETHODFORTHECONTINUOUSMEASUREMENTOFTURBIDITYABOVE1TURBIDITYUNITTU 混浊 大于

链接地址:http://www.mydoc123.com/p-526192.html