《ASTM D7184-2012 0678 Standard Test Method for Ultra Low Nitrogen in Aromatic Hydrocarbons by Oxidative Combustion and Reduced Pressure Chemiluminescence Detection《用氧化燃烧和减压化学发光检测法测.pdf》由会员分享,可在线阅读,更多相关《ASTM D7184-2012 0678 Standard Test Method for Ultra Low Nitrogen in Aromatic Hydrocarbons by Oxidative Combustion and Reduced Pressure Chemiluminescence Detection《用氧化燃烧和减压化学发光检测法测.pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D7184 12Standard Test Method forUltra Low Nitrogen in Aromatic Hydrocarbons by OxidativeCombustion and Reduced Pressure ChemiluminescenceDetection1This standard is issued under the fixed designation D7184; the number immediately following the designation indicates the year oforiginal ad
2、option or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope*1.1 This test method covers the determination of totalnitrogen in aromati
3、c hydrocarbons, such as benzene, toluene,and xylene.1.2 This test method is applicable for samples containingnitrogen from 0.38 to 1.2 mg N/kg. For higher nitrogenconcentrations refer to Test Method D6069 or D4629.1.3 In determining the conformance of the test results usingthis method to applicable
4、specifications; results shall berounded off in accordance with the rounding-off method ofPractice E29.1.4 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.1.5 This standard does not purport to address all of thesafety concerns, i
5、f any associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. For specific hazardstatements, see Section 9.2. Referenced Documents2.1 ASTM Standards:2D
6、3437 Practice for Sampling and Handling Liquid CyclicProductsD4629 Test Method for Trace Nitrogen in Liquid PetroleumHydrocarbons by Syringe/Inlet Oxidative Combustion andChemiluminescence DetectionD6069 Test Method for Trace Nitrogen in Aromatic Hydro-carbons by Oxidative Combustion and Reduced Pre
7、ssureChemiluminescence DetectionD6809 Guide for Quality Control and Quality AssuranceProcedures for Aromatic Hydrocarbons and Related Ma-terialsE29 Practice for Using Significant Digits in Test Data toDetermine Conformance with SpecificationsE691 Practice for Conducting an Interlaboratory Study toDe
8、termine the Precision of a Test Method2.2 Other Documents:OSHA Regulations, 29 CFR paragraphs 1910. 1000 and1910.120033. Terminology3.1 Definitions:3.1.1 oxidative pyrolysis, na process in which a sampleundergoes combustion in an oxygen rich environment attemperatures greater than 900C.3.1.2 pyrolyt
9、ic decomposition, ncombusting a compoundto decompose it to carbon dioxide, water and elemental oxides.3.1.3 reduced pressure chemiluminescence, na chemicalreaction at pressure less than 760 mm mercury (Hg) in whichlight is emitted.4. Summary of Test Method4.1 A specimen is introduced into a carrier
10、gas stream, at acontrolled rate, and incorporated into a high temperaturefurnace (900 to 1150C) where an excess of oxygen is added.Pyrolysis converts hydrocarbons in the specimen to carbondioxide and water. Organic nitrogen and inorganic nitrogencompounds, present in the specimen, are converted to n
11、itricoxide (NO). Nitric oxide reacts with ozone in the detectorproducing nitrogen dioxide molecules in an excited state. Asthe excited nitrogen dioxide molecules relax to ground state,light is emitted. This light is detected by a photomultiplier tubeor by a photodiode with the resulting signal propo
12、rtional to theconcentration of nitrogen in the sample. Operating the detectorat a reduced pressure lowers the probability of the excited1This test method is under the jurisdiction of ASTM Committee D16 onAromatic Hydrocarbons and Related Chemicals and is the direct responsibility ofSubcommittee D16.
13、04 on Instrumental Analysis.Current edition approved Dec. 1, 2012. Published January 2013. Originallyapproved in 2007. Last previous edition approved in 2007 as D7184 07. DOI:10.1520/D7184-12.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at ser
14、viceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3Available from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:/www.access.gpo.gov.*A Sum
15、mary of Changes section appears at the end of this standardCopyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States1nitrogen dioxide molecule colliding with other moleculesbefore it undergoes chemiluminescence. Thus, reduced pres-sure provides
16、 improved sensitivity and lower noise.5. Significance and Use5.1 This test method is useful to detect and quantifynitrogen-containing compounds at a concentration of 0.03 to 1mg N/kg in light aromatic hydrocarbons used or produced inmanufacturing process. These nitrogen-containing compoundsare undes
17、irable in finished aromatic products and may be usedin setting specification for determining the total nitrogencontent in aromatic hydrocarbons.5.2 This test method requires the use of reduced pressure atthe detector. Loss of vacuum or pressure fluctuations impactthe sensitivity of the detector and
18、the ability to determinenitrogen concentrations less than 1 mg N/kg.6. Interferences6.1 Chlorides, bromides, and iodides can interfere if any oneor all of these elements are present in a sample in concentra-tions greater than 10 % by total weight of halogen in thesample.6.2 Moisture in the sample pr
19、oduced during the combustionstep can interfere if not removed prior to the gas entering thedetector cell.7. Apparatus7.1 Pyrolysis FurnaceA furnace capable of maintaining atemperature sufficient to volatilize and pyrolyse the sample andoxidize organically bound nitrogen to NO. The actual tempera-tur
20、e should be recommended by the specific instrument manu-facturer.7.2 Quartz Pyrolysis TubeCapable of withstanding 900 to1200C.7.3 Chemiluminescence DetectorCapable of operation atreduced pressure (less than 760 mm mercury) and able tomeasure light emitted from the reaction between NO andozone.7.4 Mi
21、croliter Syringe5 to 250 L or as recommended byinstrument manufacturer.7.5 Constant Rate Injector SystemIf the sample is to beintroduced into the pyrolysis furnace via syringe, use aconstant rate injector or a liquid introduction module or asrecommended by instrument manufacturer.7.6 Liquid Auto-Sam
22、plerCapable of injecting 5 to 250 Lof sample or as recommended by instrument manufacturer.7.7 Boat Inlet System (Optional)If the instrument isequipped with a boat inlet system, care must be taken to ensurethe boat is sufficiently cooled between analyses to prevent thesample from vaporizing as it is
23、injected into the boat. Thesample should start vaporizing as it enters the furnace. It iscritical that the sample vaporize at a constant and reproduciblerate.7.8 Automatic Boat Drive SystemIf the instrument isequipped with a boat inlet system the boat should be introducedinto the furnace at a contro
24、lled rate.7.9 Membrane DryerRemoves moisture of combustionbefore the detector.8. Reagents48.1 Purity of ReagentsReagent grade chemicals shall beused in all tests. It is intended that all reagents shall conform tothe specifications of the Committee on Analytical Reagents ofthe American Chemical Socie
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD718420120678STANDARDTESTMETHODFORULTRALOWNITROGENINAROMATICHYDROCARBONSBYOXIDATIVECOMBUSTIONANDREDUCEDPRESSURECHEMILUMINESCENCEDETECTION

链接地址:http://www.mydoc123.com/p-524744.html