ASTM D5011-1992(2009) Standard Practices for Calibration of Ozone Monitors Using Transfer Standards《使用传送标准对臭氧检测器的标准实施规程》.pdf
《ASTM D5011-1992(2009) Standard Practices for Calibration of Ozone Monitors Using Transfer Standards《使用传送标准对臭氧检测器的标准实施规程》.pdf》由会员分享,可在线阅读,更多相关《ASTM D5011-1992(2009) Standard Practices for Calibration of Ozone Monitors Using Transfer Standards《使用传送标准对臭氧检测器的标准实施规程》.pdf(28页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D 5011 92 (Reapproved 2009)Standard Practices forCalibration of Ozone Monitors Using Transfer Standards1This standard is issued under the fixed designation D 5011; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 These practices describe means for calibrating ambient,workplace or indoor ozone monitors, using transfer standa
3、rds.1.2 These practices describe five types of transfer standards:(A) Analytical instruments(B) Boric acid potassium iodide (BAKI) manual analyticalprocedure(C) Gas phase titration with excess nitric oxide(D) Gas phase titration with excess ozone(E) Ozone generator device.1.3 These practices describ
4、e procedures to establish theauthority of transfer standards: qualification, certification, andperiodic recertification.1.4 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-pr
5、iate safety and health practices and determine the applica-bility of regulatory limitations prior to use. See Section 8 forspecific precautionary statements.2. Referenced Documents2.1 ASTM Standards:2D 1071 Test Methods for Volumetric Measurement of Gas-eous Fuel SamplesD 1193 Specification for Reag
6、ent WaterD 1356 Terminology Relating to Sampling and Analysis ofAtmospheresD 3195 Practice for Rotameter CalibrationD 3249 Practice for General Ambient Air Analyzer Proce-duresD 3631 Test Methods for Measuring Surface AtmosphericPressureD5110 Practice for Calibration of Ozone Monitors andCertificati
7、on of Ozone Transfer Standards Using Ultravio-let PhotometryE 591 Practice for Safety and Health Requirements Relatingto Occupational Exposure to Ozone32.2 Other Documents:40 CFR Part 50, Environmental Protection Agency Regula-tions on Ambient Air Monitoring Reference Methods43. Terminology3.1 For d
8、efinitions of terms used in this standard, seeTerminology D 1356.3.2 Definitions of Terms Specific to This Standard:3.2.1 primary standarda standard directly defined andestablished by some authority, against which all secondarystandards are compared.3.2.2 secondary standarda standard used as a means
9、 ofcomparison, but checked against a primary standard.3.2.3 standardan accepted reference sample or deviceused for establishing measurement of a physical quantity.3.2.4 transfer standarda type of secondary standard. It isa transportable device or apparatus, which, together withoperational procedures
10、, is capable of reproducing pollutantconcentration or producing acceptable assays of pollutantconcentrations.3.2.5 zero airpurified air that does not contain ozone anddoes not contain any other component that may interfere withthe measurement. See 7.1.3.3 Symbols:b = Spectrophotometer cell path leng
11、th, cm. SeeAnnex A2.davg= Average of discrete single point compari-sons. See Annex A1.di= Single point comparison. See Annex A1.FD= Diluent air flow, mL/min.FD8 = New diluent air flow, mL/min.FNO= NO flow, mL/min.FO= Flow through the O3generator, mL/min.FR= Flowrate corrected to reference conditions
12、(25C and 101.3 kPa), mL/min. See AnnexA2.1These practices are under the jurisdiction of ASTM Committee D22 on AirQuality and are the direct responsibility of Subcommittee D22.03 on AmbientAtmospheres and Source Emissions.Current edition approved March 1, 2009. Published March 2009. Originallyapprove
13、d in 1989. Last previous edition approved in 2003 as D 5011 92 (2003).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM
14、 website.3Withdrawn4Available from U.S. Government Printing Office Superintendent of Documents,732 N. Capitol St., NW, Mail Stop: SDE, Washington, DC 20401, http:/www.access.gpo.gov.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.FS=
15、 Flowrate at sampling conditions, mL/min.See Annex A2.FT= The total flow required at the output mani-fold (monitors demand plus 10 to 50 %excess), mL/min.I = The intensity of light which passes throughthe photometer absorption cell and is sensedby the detector when the cell contains an O3sample. See
16、 Annex A4.I2i= Concentration of each I2standard, mol I2/L.See Annex A2.Iavg= Average intercept. See Annex A1.Ii= Individual intercepts. See Annex A1.IO= The intensity of light which passes throughthe photometer absorption cell and is sensedby the detector when the cell contains zeroair. See Annex A4
17、.mavg= Average slope. See Annex A1.mi= Individual slopes. See Annex A1.mol I2=I2released, mols. See Annex A2.NKIO3= Normality of KIO3, equivalent/L. SeeAnnexA2.NO = Diluted NO concentration, ppm. See AnnexA4.NOORIG= Original NO concentration, ppm. See AnnexA3.NOOUT= Highest NO concentration required
18、 at theoutput manifold, ppm. It is approximatelyequal to 90 % of the upper range limit of theO3concentration to be determined. See An-nex A3.NORC= NO concentration (approximate) in the reac-tion chamber, ppm. See Annex A3.NOREM= NO concentration remaining after addition ofO3, ppm. See Annex A3.NOSTD
19、= Concentration of the undiluted NO standard,ppm.n = Number of comparisons. See Eq 4O3CERT= Certified O3concentration, ppm.O3CERT8= Diluted certified O3concentration, ppm.O3GEN=O3concentration produced by the O3genera-tor, ppm. See Annex A4.O3OUT= Indicated O3concentration, ppm. See AnnexA2.O3OUT8=
20、Diluted O3concentration, ppm.O3RC=O3concentration (approximate) at the outputmanifold, ppm.PH2O= Vapor pressure of H2OatTS, kPa, wetvolume standard. (For a dry standard,PH2O= 0.) (See Test Method D 4230 fortables of saturation vapor pressure of water.)See Annex A2.PR= Dynamic specification, determin
21、ed empiri-cally, to ensure complete reaction of O3orNO, ppm/min.PS= Barometric pressure at sampling conditions,kPa. See Annex A2.Sc= Slope of KI calibration curve, mL/mol/cm.See Annex A2.sd= Standard deviation of single point compari-sons. See Annex A1.si= Relative standard deviation of the six inte
22、r-cepts. See Annex A1.sm= Relative standard deviation of the six slopes.See Annex A1.tR= Residence time in reaction chamber, min.ts= Sampling time, min. See Annex A2.TS= Temperature at sampling conditions, C. SeeAnnex A2URL = Upper range limit of O3or NO monitor, ppm.Vi= Volume of I2solution, mL. Se
23、e Annex A2VO3= Volume of O3absorbed, L. See Annex A2.VR= Volume of air sampled, corrected to 25Cand 101.3 kPa (1 atm), mL. See Annex A2.VRC= Volume of the reaction chamber, mL.yi=O3concentration indicated by the transferstandard, ppm. See 10.6.2.Z = Recorder response with zero air, % scale.4. Summar
24、y of Practices4.1 These practices describe the procedures necessary toestablish the authority of ozone transfer standards: qualifica-tion, certification, and periodic recertification. Qualificationconsists of demonstrating that a candidate transfer standard issufficiently stable (repeatable) to be u
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD501119922009STANDARDPRACTICESFORCALIBRATIONOFOZONEMONITORSUSINGTRANSFERSTANDARDS 使用 传送 标准 臭氧 检测器

链接地址:http://www.mydoc123.com/p-518650.html