ASTM D4298-2004(2010) Standard Guide for Intercomparing Permeation Tubes to Establish Traceability《相互比较渗透管以确定示踪能力的标准指南》.pdf
《ASTM D4298-2004(2010) Standard Guide for Intercomparing Permeation Tubes to Establish Traceability《相互比较渗透管以确定示踪能力的标准指南》.pdf》由会员分享,可在线阅读,更多相关《ASTM D4298-2004(2010) Standard Guide for Intercomparing Permeation Tubes to Establish Traceability《相互比较渗透管以确定示踪能力的标准指南》.pdf(3页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: D4298 04 (Reapproved 2010)Standard Guide forIntercomparing Permeation Tubes to Establish Traceability1This standard is issued under the fixed designation D4298; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year
2、of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers two procedures for establishing thepermeation rate of a permeation tube and defining the uncer-ta
3、inty of the rate by comparison to National Institute ofStandards and Technologys Standard Reference Materials(SRM).1.2 Procedure A consists of a direct comparison of thepermeation rate of the device undergoing calibration with thatof an SRM.1.3 Procedure B consists of a gravimetric calibration pro-c
4、ess in which a certified permeation tube is used as a qualitycontrol for the measurements.1.4 Both procedures are limited to the case where a suitablecertified permeation device is available.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It
5、is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. (See 8.2 on SafetyPrecautions.)2. Referenced Documents2.1 ASTM Standards:2D1356 Terminology Relating to Sampling and Analys
6、is ofAtmospheresD3249 Practice for General Ambient Air Analyzer Proce-duresD3609 Practice for Calibration Techniques Using Perme-ation TubesD3631 Test Methods for Measuring Surface AtmosphericPressureE1 Specification for ASTM Liquid-in-Glass ThermometersE319 Practice for the Evaluation of Single-Pan
7、 MechanicalBalancesE617 Specification for Laboratory Weights and PrecisionMass Standards3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this guide, refer toTerminology D1356.3.1.2 Definitions of Terms Specific to This Standard:3.1.2.1 working standarda standard used in the labor
8、atoryor field for periodic standardization of a measuring instrument.4. Summary of Guide4.1 Procedure AA certified SRM permeation source, ob-tained from the National Institute of Standards and Technologyis used to calibrate a continuous analyzer. The analyzer is thenused to measure the concentration
9、 of a gaseous mixturegenerated from the permeation tube under calibration. Equa-tions are provided that permit calibration of the permeation rateof the latter from the test data.4.2 Procedure BThe permeation source is calibrated,gravimetrically, using temperature and mass standards trace-able to NIS
10、T standards. The validity of the calibration isconfirmed by concurrently measuring the permeation rate of aCertified Reference Material (CRM).5. Significance and Use5.1 The accuracy of air pollution measurements is directlydependent upon accurate calibrations.5.2 Such measurements gain accuracy and
11、can be intercom-pared when the measurement procedures are traceable tonational measurement standards.5.3 This guide describes procedures for enhancing theaccuracy of air pollution measurements which may be specifiedby those organizations requiring traceability to national stan-dards.6. Apparatus6.1
12、For apparatus used in the calibration of permeationdevices, refer to Practice D3609.1This guide is under the jurisdiction of ASTM Committee D22 on Air Qualityand is the direct responsibility of Subcommittee D22.01 on Quality Control.Current edition approved July 1, 2010. Published July 2010. Origina
13、lly approvedin 1983. Last previous edition approved in 2004 as D4298 - 04. DOI: 10.1520/D4298-04R10.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Do
14、cument Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.6.1.1 The thermometers used shall conform to SpecificationE1 and shall have calibration certificates traceable to the NIST.Measurement uncertainty
15、 should be 0.1C or less.6.1.2 The mercury barometer shall conform to Test MethodsD3631.6.2 Apparatus for Procedure A:6.2.1 Analytical InstrumentsAn analytical instrument re-sponsive to the permeant with the following minimum perfor-mance specifications:Noise 1 % of full-scaleZero drift 64 % of full-
16、scale per daySpan drift 63 % of full-scale per dayRange 0 to 0.5 ppm (or appropriate for source strength)6.2.2 Continuous strip chart recorder with the followingminimum performance specifications:Uncertaintycomponent0.33 3 (0.25 % full-scaledeflection)Chart width no less than 6 in.Time for full-scal
17、e travel 1 sNOTE 1ISO GUM3points out that with approximately normal plottedpoints, the above maximum 3 3 standard deviations is equivalent to99 % of plotted points lying within 6(0.25 % full-scale deflection) oftrue values.6.3 Apparatus for Procedure B:6.3.1 Analytical balance, meeting the requireme
18、nts of Prac-tices E319 and D3609.6.3.2 Analytical weights meeting the requirements of Speci-fication E617 and having a calibration certificate traceable tothe NIST.7. Materials7.1 Refer to Practice D3609.7.2 CRM Permeation Device.48. Precautions8.1 Procedural Precautions:8.1.1 The procedural precaut
19、ions described in PracticeD3609 are applicable to the present guide.8.1.2 When possible, the permeation device should becompared to the CRM using the same system with identicalflow and temperature conditions. Unpredictable errors may beintroduced if permeation devices are compared at widelydifferent
20、 temperatures and flow rates (pertains to Procedure A).Intercomparisons are valid only at temperatures for which theCRM tube is calibrated.8.1.3 Equilibration of the permeation device, the calibrationequipment, and the analytical system must be assured, prior touse. During storage, avoid exposing tu
21、bes to high humidities orwide variations in temperature, that may permanently alter thepermeation rate.8.2 Safety Precautions:8.2.1 For precautions concerning the use of analytical in-struments and of cylinders containing pressurized gases, seePractice D3249.9. Calibration and Standardization9.1 Pro
22、cedure A:9.1.1 Set up a gas generation system using a CertifiedReference Material (CRM) permeation tube and apparatus andprocedure such as described in Practice D3609. Equilibrate atthe desired temperature of calibration.9.1.2 Optimize the performance of the analytical instrumentaccording to the man
23、ufacturers instructions.9.1.3 Using dry air or nitrogen, set the zero point of theinstrument.9.1.4 Use the gas generation system to provide gas concen-trations corresponding to 20, 40, 60, and 80 % of full-scalereadings. Record the concentration and respective readings.Repeat the measurements in ran
24、dom order.9.1.5 Plot concentration versus instrument readings anddraw the line of best fit, or alternatively fit by linear leastsquares regression. Calculate the slope (ppm(v)/scale reading)and standard deviation estimate scal, expressed as g/min.9.1.5.1 If any point deviates by more than 61 % from
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD429820042010STANDARDGUIDEFORINTERCOMPARINGPERMEATIONTUBESTOESTABLISHTRACEABILITY 相互 比较 渗透 确定 能力

链接地址:http://www.mydoc123.com/p-516754.html