1、Designation: D 4298 04Standard Guide forIntercomparing Permeation Tubes to Establish Traceability1This standard is issued under the fixed designation D 4298; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the year of last revision
2、. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.1. Scope1.1 This guide covers two procedures for establishing thepermeation rate of a permeation tube and defining the uncer-tainty of the rat
3、e by comparison to National Institute ofStandards and Technologys Standard Reference Materials(SRM).1.2 Procedure A consists of a direct comparison of thepermeation rate of the device undergoing calibration with thatof an SRM.1.3 Procedure B consists of a gravimetric calibration pro-cess in which a
4、certified permeation tube is used as a qualitycontrol for the measurements.1.4 Both procedures are limited to the case where a suitablecertified permeation device is available.1.5 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsib
5、ility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to use. (See 8.2 on SafetyPrecautions.)2. Referenced Documents2.1 ASTM Standards:2D 1356 Terminology Relating to Sampling and Analysis ofAtmospher
6、esD 3249 Practice for General Ambient Air Analyzer Proce-duresD 3609 Practice for Calibration Techniques Using Perme-ation TubesD 3631 Test Methods for Measuring Surface AtmosphericPressureE 1 Specification for ASTM ThermometersE 319 Practice for the Evaluation of Single-Pan MechanicalBalancesE 617
7、Specification for Laboratory Weights and PrecisionMass Standards3. Terminology3.1 Definitions:3.1.1 For definitions of terms used in this guide, refer toTerminology D 1356.3.1.2 Definitions of Terms Specific to This Standard:3.1.2.1 working standarda standard used in the laboratoryor field for perio
8、dic standardization of a measuring instrument.4. Summary of Guide4.1 Procedure AA certified SRM permeation source, ob-tained from the National Institute of Standards and Technologyis used to calibrate a continuous analyzer. The analyzer is thenused to measure the concentration of a gaseous mixturege
9、nerated from the permeation tube under calibration. Equa-tions are provided that permit calibration of the permeation rateof the latter from the test data.4.2 Procedure BThe permeation source is calibrated,gravimetrically, using temperature and mass standards trace-able to NIST standards. The validi
10、ty of the calibration isconfirmed by concurrently measuring the permeation rate of aCertified Reference Material (CRM).5. Significance and Use5.1 The accuracy of air pollution measurements is directlydependent upon accurate calibrations.5.2 Such measurements gain accuracy and can be intercom-pared w
11、hen the measurement procedures are traceable tonational measurement standards.5.3 This guide describes procedures for enhancing theaccuracy of air pollution measurements which may be specifiedby those organizations requiring traceability to national stan-dards.6. Apparatus6.1 For apparatus used in t
12、he calibration of permeationdevices, refer to Practice D 3609.6.1.1 The thermometers used shall conform to SpecificationE1and shall have calibration certificates traceable to the NIST.Measurement uncertainty should be 0.1C or less.1This guide is under the jurisdiction of ASTM Committee D22 on Sampli
13、ng andAnalysis of Atmospheres and is the direct responsibility of Subcommittee D22.01on Quality Control.Current edition approved October 1, 2004. Published October 2004. Originallyapproved in 1983. Last previous edition approved in 1999 as D 4298 - 95 (1999).2For referenced ASTM standards, visit the
14、 ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, U
15、nited States.6.1.2 The mercury barometer shall conform to Test MethodsD 3631.6.2 Apparatus for Procedure A:6.2.1 Analytical InstrumentsAn analytical instrument re-sponsive to the permeant with the following minimum perfor-mance specifications:Noise 1 % of full-scaleZero drift 64 % of full-scale per
16、daySpan drift 63 % of full-scale per dayRange 0 to 0.5 ppm (or appropriate for source strength)6.2.2 Continuous strip chart recorder with the followingminimum performance specifications:Uncertaintycomponent0.33 3 (0.25 % full-scaledeflection)Chart width no less than 6 in.Time for full-scale travel 1
17、 sNOTE 1ISO GUM3points out that with approximately normal plottedpoints, the above maximum 3 3 standard deviations is equivalent to99 % of plotted points lying within 6(0.25 % full-scale deflection) oftrue values.6.3 Apparatus for Procedure B:6.3.1 Analytical balance, meeting the requirements of Pra
18、c-tices E 319 and D 3609.6.3.2 Analytical weights meeting the requirements of Speci-fication E 617 and having a calibration certificate traceable tothe NIST.7. Materials7.1 Refer to Practice D 3609.7.2 CRM Permeation Device.48. Precautions8.1 Procedural Precautions:8.1.1 The procedural precautions d
19、escribed in PracticeD 3609 are applicable to the present guide.8.1.2 When possible, the permeation device should becompared to the CRM using the same system with identicalflow and temperature conditions. Unpredictable errors may beintroduced if permeation devices are compared at widelydifferent temp
20、eratures and flow rates (pertains to Procedure A).Intercomparisons are valid only at temperatures for which theCRM tube is calibrated.8.1.3 Equilibration of the permeation device, the calibrationequipment, and the analytical system must be assured, prior touse. During storage, avoid exposing tubes t
21、o high humidities orwide variations in temperature, that may permanently alter thepermeation rate.8.2 Safety Precautions:8.2.1 For precautions concerning the use of analytical in-struments and of cylinders containing pressurized gases, seePractice D 3249.9. Calibration and Standardization9.1 Procedu
22、re A:9.1.1 Set up a gas generation system using a CertifiedReference Material (CRM) permeation tube and apparatus andprocedure such as described in Practice D 3609. Equilibrate atthe desired temperature of calibration.9.1.2 Optimize the performance of the analytical instrumentaccording to the manufa
23、cturers instructions.9.1.3 Using dry air or nitrogen, set the zero point of theinstrument.9.1.4 Use the gas generation system to provide gas concen-trations corresponding to 20, 40, 60, and 80 % of full-scalereadings. Record the concentration and respective readings.Repeat the measurements in random
24、 order.9.1.5 Plot concentration versus instrument readings anddraw the line of best fit, or alternatively fit by linear leastsquares regression. Calculate the slope (ppm(v)/scale reading)and standard deviation estimate scal, expressed as g/min.9.1.5.1 If any point deviates by more than 61 % from the
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMD42982004STANDARDGUIDEFORINTERCOMPARINGPERMEATIONTUBESTOESTABLISHTRACEABILITY 相互 比较 渗透 确定 能力 标准 指南

链接地址:http://www.mydoc123.com/p-516753.html