FORD FLTM AP 151-01-2016 Determination of Cyclical Volatile Siloxane Compounds in Seals Using Thermal Extraction and Gas Chromatography-Mass Spectrometry (GC-MS).pdf
《FORD FLTM AP 151-01-2016 Determination of Cyclical Volatile Siloxane Compounds in Seals Using Thermal Extraction and Gas Chromatography-Mass Spectrometry (GC-MS).pdf》由会员分享,可在线阅读,更多相关《FORD FLTM AP 151-01-2016 Determination of Cyclical Volatile Siloxane Compounds in Seals Using Thermal Extraction and Gas Chromatography-Mass Spectrometry (GC-MS).pdf(5页珍藏版)》请在麦多课文档分享上搜索。
1、 FORD LABORATORY TEST METHOD AP 151-01 Date Action Revisions Rev. 0 2016 10 21 Released A. Olejnik and D. Bellile, NA Controlled document at www.MATS Copyright 2016, Ford Global Technologies, LLC Page 1 of 5 Determination of Cyclical Volatile Siloxane Compounds in Seals Using Thermal Extraction and
2、Gas Chromatography-Mass Spectrometry (GC-MS) Application This procedure is used to quantitatively determine thermal extraction rates of selected cyclical volatile siloxane compounds (D3, D4, D5, D6 and D9) in elastomeric materials. Other cyclical volatile siloxane compounds (D7, D8, others) can be d
3、etermined semi-quantitatively. Apparatus and Materials Required Thermal Desorption Tubes for the Gerstel Twister Thermal Desorption System (supplier Gerstel): Tubes with glass frit (Gerstel part # 013742-005-0) Tubes with Tenax GR (Gerstel part # 015953-005-0) GC-MS Chromatographic Column: DB-5 MS,
4、column dimensions: length: 30 m, diameter: 0.25 mm, film thickness: 0.25 m (Supplier: Agilent or equivalent) Gerstel Twister Inlet Liner: CS4 liner with silanized glass wool; ID=1 mm; length= 76 mm; Gerstel Part# 013409-010-00 GC-MS Carrier and Gerstel ThermoExtractor Conditioning Unit Gas: Ultra Hi
5、gh Purity Helium (Supplier: AirGas or equivalent) Individual Standards: neat dimethylsilicone cyclic compounds (purity 99.8% or better): D3 (Cyclotrisiloxane, hexamethyl-; CAS # 541-05-9) D4 (Cyclotetrasiloxane, octamethyl-; CAS # 556-67-1) D5 (Cyclopentasiloxane, decamethyl-; CAS # 541-02-6) D6 (Cy
6、clohexasiloxane, dodecamethyl-; CAS # 540-97-6) D9 (Cyclononasiloxane, octadecamethyl-; CAS # 556-71-8) (Supplier: Ohio Valley Specialty Company or equivalent) Glass volumetric cylinders: 10 mL and 50 mL Pipets (9”, Fisher Brand) Stainless Steel Surgical Blade Syringe: 10 L Liquid Nitrogen: low pres
7、sure liquid nitrogen required for GC-MS “Cryo Cooling” (Supplier: AirGas or equivalent) Solvent: Methanol, certified (Supplier: Fisher Scientific or equivalent) Thermal Desorption System: Gerstel Twister Thermal Desorption Unit (TDU) with MPS2 auto-sampler Chromatography System: Agilent Gas Chromato
8、graph, 6890N Agilent Mass Spectrometer, 5975B VL MS FORD LABORATORY TEST METHOD AP 151-01 Copyright 2016, Ford Global Technologies, LLC Page 2of 5 Controlling Software: Agilent Chemstation GC-MS, Data Acquisition/Data Analysis Software Version: D.03.00, SP1 Gerstel MASter, Thermal Extraction Control
9、 Software, Version 2.8.0.0 Tube Conditioner: Gerstel TubeConditioner, model: TC (with a metal extension for short tubes) Analytical Balance: Mettler-Toledo, model: PM400 Note: All the above listed equipment and/or software may be substituted with alternatives of equivalent performance. Chemicals, ma
10、terials, parts, and equipment referenced in this document must be used and handled properly. Each party is responsible for determining proper use and handling in its facilities. Procedure Preparation of standard and quality control solutions: Prepare a stock standard solution in methanol (1,000 g/mL
11、 mixture of each target volatile siloxane compound). From the stock standard solution prepare calibration standards of 250 and 500 g/mL mixtures of each compound by serial dilutions by volume. Separately prepare a solution of the 500 g/mL mixture of each compound from the stock solution. This soluti
12、on will be used as a quality assurance /quality control check solution (QA/QC). Store the solutions in sealed vials in a refrigerator; the solutions are stable for 2 months. FORD LABORATORY TEST METHOD AP 151-01 Copyright 2016, Ford Global Technologies, LLC Page 3of 5 GC-MS Data Acquisition Methods
13、(Method 1 and Method 2): Method 1 Create a data acquisition Method 1. This method is a calibration method and will be used to thermally desorb the Twister GR thermal extraction tubes spiked with calibration standard solutions or QC/QA check solution. In Method 1 the thermal extraction temperature of
14、 the Twister unit is set at 300C to thermally desorb spiked/trapped target cyclical siloxane compounds onto the GC-MS system. It is recommended to exclude the data analysis segment from the Method 1. For the data analysis segment create Method 3 (see below). Instrument control parameters: Oven: Init
15、ial Temperature=50C; Initial Time=0.00 min; Temperature Ramps: 5C/min to 200C (hold 1.00 min); 20C/min to 300C (hold 5 min); Run Time=41.00 min; Cryo=On (enabled). Inlet: Mode=Split; Split Ratio: 20:1; Initial Temperature=OFF; Pressure=7.64 psi; Gas=Helium. Column: DB5-MS; 30.0 m; 0.25 mm ID; 0.25 u
16、m film thickness; Constant Flow 1.0 mL/min; Outlet Pressure=Vacuum. MS Acquisition Parameters: Tune File= stune.u; Acquistion Mode=Scan; MSD Transfer Line Heater=250C; Solvent Delay=2.00 min; Low Mass=33.0; High Mass=550.0; MS Quad=150C; MS Source=230C. Gerstel CIS: Initial Temperature=-90C; Equilbr
17、ation Time=0.10 min; Initial Time=0.20 min; Temperature Ramp=12.00C/sec; End Temperature=300C; Hold Time=3.00 min. Gerstel TDU: Initial Temperature=30C; Delay Time=0.40 min; Initial Time=0.10 min; Ramp: Rate=300C/min; End Temperature=300C; Hold Time=5.00 min; Splitless Mode. Method 2 Create a data a
18、cquisition Method 2. This method will be used to thermally desorb the volatile siloxane compounds from seal materials using glass frit thermal extraction tubes. In Method 2 the thermal extraction temperature of the Twister unit is set at 120C to thermally desorb the cyclical siloxane compounds (and
19、other VOC) from the tested seal specimen onto the GC-MS system. It is recommended to exclude the data analysis segment from the Method 2. For the data analysis segment create Method 3 (see below). Instrument Control Parameters: Same as in the Method 1 except one parameter in the Gerstel TDU: End Tem
20、perature=120C. GC-MS calibration of the cyclical siloxanes standards: Draw 1 L of 250 g/mL of the cyclical siloxane standard solution into a 10 L syringe and dispense it on the frit surface of the Tenax GR thermal extraction tube. Then apply a slight upstream helium flow into the tube to evaporate t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
10000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- FORDFLTMAP151012016DETERMINATIONOFCYCLICALVOLATILESILOXANECOMPOUNDSINSEALSUSINGTHERMALEXTRACTIONANDGASCHROMATOGRAPHYMASSSPECTROMETRYGCMSPDF

链接地址:http://www.mydoc123.com/p-746819.html