1、Designation: G 59 97 (Reapproved 2009)Standard Test Method forConducting Potentiodynamic Polarization ResistanceMeasurements1This standard is issued under the fixed designation G 59; the number immediately following the designation indicates the year of originaladoption or, in the case of revision,
2、the year of last revision. A number in parentheses indicates the year of last reapproval. A superscriptepsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method covers an experimental procedure forpolarization resistance measurements which can be use
3、d for thecalibration of equipment and verification of experimentaltechnique. The test method can provide reproducible corrosionpotentials and potentiodynamic polarization resistance mea-surements.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are inclu
4、ded in thisstandard.1.3 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate safety and health practices and determine the applica-bility of regulatory limitations prior to
5、 use.2. Referenced Documents2.1 ASTM Standards:2G3 Practice for ConventionsApplicable to ElectrochemicalMeasurements in Corrosion TestingG5 Reference Test Method for Making Potentiostatic andPotentiodynamic Anodic Polarization MeasurementsG 102 Practice for Calculation of Corrosion Rates andRelated
6、Information from Electrochemical Measurements3. Significance and Use3.1 This test method can be utilized to verify the perfor-mance of polarization resistance measurement equipment in-cluding reference electrodes, electrochemical cells, poten-tiostats, scan generators, measuring and recording device
7、s. Thetest method is also useful for training operators in samplepreparation and experimental techniques for polarization resis-tance measurements.3.2 Polarization resistance can be related to the rate ofgeneral corrosion for metals at or near their corrosion potential,Ecorr. Polarization resistance
8、 measurements are an accurate andrapid way to measure the general corrosion rate. Real timecorrosion monitoring is a common application. The techniquecan also be used as a way to rank alloys, inhibitors, and so forthin order of resistance to general corrosion.3.3 In this test method, a small potenti
9、al scan, DE(t), definedwith respect to the corrosion potential (DE=EEcorr), isapplied to a metal sample. The resultant currents are recorded.The polarization resistance, RP, of a corroding electrode isdefined from Eq 1 as the slope of a potential versus currentdensity plot at i =0(1-4):3Rp5SDE iDi50
10、, dE/dt0(1)The current density is given by i. The corrosion currentdensity, icorr, is related to the polarization resistance by theStern-Geary coefficient, B. (3),icorr5 106BRp(2)The dimension of Rpis ohm-cm2, icorris muA/cm2, and B isin V. The Stern-Geary coefficient is related to the anodic, ba,an
11、d cathodic, bc, Tafel slopes as per Eq 3.B 5babc2.303ba1 bc!(3)The units of the Tafel slopes are V. The corrosion rate, CR,in mm per year can be determined from Eq 4 in which EW isthe equivalent weight of the corroding species in grams and ris the density of the corroding material in g/cm3.CR 5 3.27
12、 3 103icorrEWr(4)1This test method is under the jurisdiction of ASTM Committee G01 onCorrosion of Metals and is the direct responsibility of Subcommittee G01.11 onElectrochemical Measurements in Corrosion Testing.Current edition approved May 1, 2009. Published May 2009. Originallyapproved in 1978. L
13、ast previous edition approved in 2003 as G 5997(2003).2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.3The bo
14、ldface numbers in parentheses refer to the list of references at the end ofthis standard.1Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.Refer to Practice G 102 for derivations of the above equa-tions and methods for estimating Tafel
15、 slopes.3.4 The test method may not be appropriate to measurepolarization resistance on all materials or in all environments.See 8.2 for a discussion of method biases arising from solutionresistance and electrode capacitance.4. Apparatus4.1 The apparatus is described in Test Method G5.Itincludesa1Lr
16、ound bottom flask modified to permit theaddition of inert gas, thermometer, and electrodes. This stan-dard cell or an equivalent cell can be used. An equivalent cellmust be constructed of inert materials and be able to reproducethe standard curve in Test Method G5.4.2 A potentiostat capable of varyi
17、ng potential at a constantscan rate and measuring the current is needed.4.3 A method of recording the varying potential and result-ing current is needed.5. Test of Electrical Equipment5.1 Before the polarization resistance measurement is made,the instrument system (potentiostat, X-Y recorder or data
18、acquisition system) must be tested to ensure proper function-ing. For this purpose, connect the potentiostat to a testelectrical circuit (5). While more complex dummy cells aresometimes needed in electrochemical studies, the simple resis-tor shown in Fig. 1 is adequate for the present application.5.
19、2 Use R = 10.0 V. Set the applied potential on thepotentiostat to E = 30.0 mV and apply the potential. Thecurrent should be 3.0 mA by Ohms Law, I = E/R.NOTE 1When polarization resistance values are measured for systemswith different corrosion currents, the value of R should be chosen to coverthe cur
20、rent range of the actual polarization resistance measurement.Expected corrosion currents in the microampere range requireR=1to10kV.5.3 Record the potentiodynamic polarization curve at a scanrate of 0.6 V/h from DE = 30 mV to DE = +30 mV and backto DE = 30 mV. The plot should be linear, go through th
21、eorigin, and have a slope 10 V. The curves recorded for theforward and reverse scans should be identical.5.4 If the observed results are different than expected, theelectrochemical equipment may require calibration or servicingin accordance with the manufacturers guidelines.6. Experimental Procedure
22、6.1 The 1.0 N H2SO4test solution should be prepared fromAmerican Chemical Society reagent grade acid and distilledwater as described in Test Method G5. The standard test cellrequires 900 mL of test solution. The temperature must bemaintained at 30C within 1.6.2 The test cell is purged at 150 cm3/min
23、 with an oxygen-free gas such as hydrogen, nitrogen, or argon. The purge isstarted at least 30 min before specimen immersion. The purgecontinues throughout the test.6.3 The working electrode should be prepared as detailed inTest Method G5. The experiment must commence within 1 hof preparing the elec
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMG5919972009STANDARDTESTMETHODFORCONDUCTINGPOTENTIODYNAMICPOLARIZATIONRESISTANCEMEASUREMENTS 电势 极化

链接地址:http://www.mydoc123.com/p-540664.html