ASTM F2119-2007(2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants《评定无源植入物磁共振图像产物的标准试验方法》.pdf
《ASTM F2119-2007(2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants《评定无源植入物磁共振图像产物的标准试验方法》.pdf》由会员分享,可在线阅读,更多相关《ASTM F2119-2007(2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants《评定无源植入物磁共振图像产物的标准试验方法》.pdf(4页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F2119 07 (Reapproved 2013)Standard Test Method forEvaluation of MR Image Artifacts from Passive Implants1This standard is issued under the fixed designation F2119; the number immediately following the designation indicates the year oforiginal adoption or, in the case of revision, the ye
2、ar of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon () indicates an editorial change since the last revision or reapproval.1. Scope1.1 This test method characterizes the distortion and signalloss artifacts produced in a magnetic resonance (MR) ima
3、ge bya passive implant (implant that functions without the supply ofelectrical or external power). Anything not established to beMR-Safe or MR-Conditional is excluded.1.2 The values stated in SI units are to be regarded asstandard. No other units of measurement are included in thisstandard.2. Refere
4、nced Documents2.1 ASTM Standards:2F2052 Test Method for Measurement of Magnetically In-duced Displacement Force on Medical Devices in theMagnetic Resonance EnvironmentF2182 Test Method for Measurement of Radio FrequencyInduced Heating On or Near Passive Implants DuringMagnetic Resonance ImagingF2213
5、 Test Method for Measurement of Magnetically In-duced Torque on Medical Devices in the Magnetic Reso-nance EnvironmentF2503 Practice for Marking Medical Devices and OtherItems for Safety in the Magnetic Resonance Environment3. Terminology3.1 Definitions:3.1.1 artifact width, nthe maximum distance (m
6、m) fromthe edge of the implant to the fringe of the resulting imageartifact found in the entire set of images acquired using this testmethod.3.1.2 image artifact, na pixel in an image is considered tobe part of an image artifact if the intensity is changed by atleast 30 % when the device is present
7、compared to a referenceimage in which the device is absent.3.1.3 magnetic resonance (MR) environment, nvolumewithin the 0.50 mT (5 gauss (G) line of an MR system, whichincludes the entire three dimensional volume of space sur-rounding the MR scanner. For cases where the 0.50 mT line iscontained with
8、in the Faraday shielded volume, the entire roomshall be considered the MR environment.3.1.4 magnetic resonance imaging (MRI), nimaging tech-nique that uses static and time varying magnetic fields toprovide images of tissue by the magnetic resonance of nuclei.3.1.5 MR-Conditional, adjan item that has
9、 been demon-strated to pose no known hazards in a specified MR environ-ment with specified conditions of use. Field conditions thatdefine the specified MR environment include field strength,spatial gradient, dB/dt (time rate of change of the magneticfield), radio frequency (RF) fields, and specific
10、absorption rate(SAR).Additional conditions, including specific configurationsof the item, may be required.3.1.6 MR-Safe, adjan item that poses no known hazards inall MR environments.NOTE 1MR-Safe items include nonconducting, nonmagnetic itemssuch as a plastic petri dish.An item may be determined to
11、be MR-Safe byproviding a scientifically based rationale rather than test data.3.1.7 MR-Unsafe, adjan item that is known to posehazards in all MR environments.NOTE 2MR-Unsafe items include magnetic items such as a pair offerromagnetic scissors.3.1.8 tesla (T), nthe SI unit of magnetic induction equal
12、 to104G.4. Summary of Test Method4.1 Pairs of spin echo images are generated both with andwithout the implant in the field of view. Image artifacts areassessed by computing differences outside the region corre-sponding to the implant between reference and implant images.Once the worst case condition
13、s using the spin echo pulsesequence are ascertained, a pair of gradient echo images areacquired under the same conditions.5. Significance and Use5.1 This test method provides a quantified measure of theimage artifact produced under a standard set of scanningconditions.1This test method is under the
14、jurisdiction ofASTM Committee F04 on Medicaland Surgical Materials and Devices and is the direct responsibility of SubcommitteeF04.15 on Material Test Methods.Current edition approved March 1, 2013. Published March 2013. Originallyapproved in 2001. Last previous edition approved in 2007 as F2119 07.
15、 DOI:10.1520/F2119-07R13.2For referenced ASTM standards, visit the ASTM website, www.astm.org, orcontact ASTM Customer Service at serviceastm.org. For Annual Book of ASTMStandards volume information, refer to the standards Document Summary page onthe ASTM website.Copyright ASTM International, 100 Ba
16、rr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States15.2 This test method applies only to passive implants thathave been established to be MR-Safe or MR-Conditional.6. Apparatus6.1 An MR imaging system with a static field strength of 1.5T or 3.0 T is recommended. The MRI sys
17、tem must have theability to swap readout and phase-encode directions.6.2 A reference object made from a nondistorting medium,such as 0.5-in. diameter nylon rod.7. Test Specimen7.1 The implant for which image artifact is to be measuredshall serve as the test specimen.7.2 For the purposes of device qu
18、alification, the deviceevaluated according to this test method should be a finishedsterilized device.NOTE 3The device does not have to be sterile at the time of testing;however, it should have been subjected to all processing, packaging, andsterilization steps before testing because any of these ste
19、ps may affect themagnetic properties of the device.7.3 This test method may be used on prototype devices atany stage of production during product development. A justi-fication for using a prototype instead of the finished devicemust be provided.8. Procedure8.1 MR Imaging Parameters for Testing Artif
20、acts:8.1.1 The recommended MR imaging test environment forevaluation of artifacts are given as follows.An alternative maybe used if an adequate case can be made for relevance to thespecific device. Field of view, slice thickness, and matrix sizeshall be adjusted to achieve pixel dimensions to accura
21、telymeasure the artifact. Two example situations are described, onefor small implants, such as a coronary stent, and one for largerimplants such as an artificial hip joint.Static field strength: 1.5T (see 6.1)Bandwidth: 32 kHz (required)Field of view: sufficient to encompass the entire implantand th
22、e artifactSmall implant (for example, coronary stent):Matrix size: 256 256Slice thickness: 3 mmLarge implant (for example, hip implant):Matrix size: 256 128Slice thickness: 5 mmTwo different pulse sequences will be used:Pulse sequence: spin echoTR: 500 msTE: 20 msPulse sequence: gradient echoTR: 100
23、 500 msTE: 15 msFlip angle: 308.1.2 The device should be immersed in a solution. Forexample, a copper sulfate (CuSO4) solution (12 g/L) may beused to reduce T1and keep TR at a reasonable level. Thedevice may be suspended in nylon netting. If a copper sulfatesolution is inappropriate for a particular
24、 device, a substitutemay be used but a justification must be provided. Nickelchloride (NiCl2) and manganese chloride (MnCl2) are possiblesubstitutes. To achieve adequate field homogeneity, thereshould be at least 4 cm of clearance between the device andeach side of the container holding the solution
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF211920072013STANDARDTESTMETHODFOREVALUATIONOFMRIMAGEARTIFACTSFROMPASSIVEIMPLANTS 评定 无源 植入 磁共振 图像

链接地址:http://www.mydoc123.com/p-537796.html