ASTM F2052-2006e1 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment《测量磁共振环境中医疗设备磁感应位移力的标准试验方法.pdf
《ASTM F2052-2006e1 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment《测量磁共振环境中医疗设备磁感应位移力的标准试验方法.pdf》由会员分享,可在线阅读,更多相关《ASTM F2052-2006e1 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment《测量磁共振环境中医疗设备磁感应位移力的标准试验方法.pdf(6页珍藏版)》请在麦多课文档分享上搜索。
1、Designation: F 2052 06e1Standard Test Method forMeasurement of Magnetically Induced Displacement Forceon Medical Devices in the Magnetic ResonanceEnvironment1This standard is issued under the fixed designation F 2052; the number immediately following the designation indicates the year oforiginal ado
2、ption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. Asuperscript epsilon (e) indicates an editorial change since the last revision or reapproval.e1NOTEParagraph X1.3 was added editorially in May 2006.1. Scope1.1 This test metho
3、d covers the measurement of the mag-netically induced displacement force produced by static mag-netic field gradients on medical devices and the comparison ofthat force to the weight of the medical device.1.2 This test method does not address other possible safetyissues which include but are not lim
4、ited to issues of magneti-cally induced torque, RF heating, induced heating, acousticnoise, interaction among devices, and the functionality of thedevice and the MR system.1.3 This test method is intended for devices that can besuspended from a string. Devices which cannot be suspendedfrom a string
5、are not covered by this test method. The weightof the string from which the device is suspended during the testmust be less than 1 % of the weight of the tested device.1.4 This test method shall be carried out in a system inwhich the direction of the magnetically induced deflectionforce is horizonta
6、l.1.5 The values stated in SI units are to be regarded asstandard. Values in parentheses are for information only.1.6 This standard does not purport to address all of thesafety concerns, if any, associated with its use. It is theresponsibility of the user of this standard to establish appro-priate s
7、afety and health practices and determine the applica-bility of regulatory requirements prior to use.2. Referenced Documents2.1 ASTM Standards:2F2119 Test Method for Evaluation of MR Image Artifactsfrom Passive ImplantsF 2182 Test Method for Measurement of Radio FrequencyInduced Heating Near Passive
8、Implants During MagneticResonance ImagingF 2213 Test Method for Measurement of Magnetically In-duced Torque on Medical Devices in the Magnetic Reso-nance EnvironmentF 2503 Practice for Marking Medical Devices and OtherItems for Safety in the Magnetic Resonance Environment2.2 Other Standards:3IEC 606
9、01233 Ed. 2.0 Medical Electronic EquipmentPart 2: Particular Requirements for the Safety of MagneticResonance Equipment for Medical DiagnosisISO 13485:2003(E) Medical DevicesQuality Manage-ment SystemsRequirements for Regulatory Purposes,definition 3.73. Terminology3.1 Definitions:3.1.1 diamagnetic
10、materiala material whose relative per-meability is less than unity.3.1.2 ferromagnetic materiala material whose magneticmoments are ordered and parallel producing magnetization inone direction.3.1.3 magnetic field strength (H in A/m)strength of theapplied magnetic field.3.1.4 magnetic induction or m
11、agnetic flux density (B inT)that magnetic vector quantity which at any point in amagnetic field is measured either by the mechanical forceexperienced by an element of electric current at the point, or bythe electromotive force induced in an elementary loop duringany change in flux linkages with the
12、loop at the point. Themagnetic induction is frequently referred to as the magneticfield. Bois the static field in an MR system. Plain type indicatesa scalar (for example, B) and bold type indicates a vector (forexample, B).3.1.5 magnetic resonance diagnostic devicea device in-tended for general diag
13、nostic use to present images whichreflect the spatial distribution or magnetic resonance spectra, or1This test method is under the jurisdiction of ASTM Committee F04 on Medicalmetals (for surgical implants andmedical devices); MRI (magnetic resonance imaging); MRsafetyAPPENDIXES(Nonmandatory Informa
14、tion)X1. RATIONALE FOR DEVELOPMENT OF THE TEST METHODX1.1 The primary reason for this test method is to deter-mine the magnetically induced deflection force on medicaldevices that may be subjected to magnetic resonance imaging.Note that this test method only addresses the magneticallyinduced deflect
15、ion force and that the results of this test aloneare not sufficient to determine whether a particular device issafe in the MR environment. The deflection force is producedby the static magnetic field gradients. The static field alsoproduces a torque on a device that acts to align the object withthe
16、magnetic field (like a compass needle aligns itself with theEarths magnetic field). For a device to be safe in the MRenvironment, the magnetically induced deflection force andtorque should be less than forces and torques to which thedevice would be subjected if it were not in a large magneticfield,
17、for example, a force less than the weight of the deviceand a torque less than that produced by normal daily activities(which might include rapidly accelerating vehicles or amuse-ment park rides). Other possible safety issues include but arenot limited to RF heating, induced heating, acoustic noise,i
18、nteraction among devices, and the functionality of the deviceand the MR system. Although a commercial 1.5 T MR systemcurrently produces the conditions that would most commonlybe encountered by a medical device, 3 T MR systems havebeen cleared for market and are becoming more common inclinical situat
19、ions. It is important to note that a medical devicethat is safe in a 1.5 T scanner may not be so in a system withhigher or lower static field strength (for example, a 3 T systemor a 1 T system). Also, there can be major differences in thecharacteristics of open and cylindrical MR systems. Forinstanc
20、e, the static field spatial gradients may be significantlyhigher in open systems.X1.1.1 After the safety of a device has been determined, itshould be marked as MR Safe, MR Conditional, or MR Unsafeusing the definitions and icons given in Practice F 2503. Theterms are defined in Practice F 2503 as:X1
21、.1.2 MR Safean item that poses no known hazards inall MR environments.NOTE X1.1MR Safe items include nonconducting, nonmagneticitems such as a plastic Petri dish. An item may be determined to be MRSafe by providing a scientifically based rationale rather than test data.X1.1.3 MR Conditionalan item t
- 1.请仔细阅读文档,确保文档完整性,对于不预览、不比对内容而直接下载带来的问题本站不予受理。
- 2.下载的文档,不会出现我们的网址水印。
- 3、该文档所得收入(下载+内容+预览)归上传者、原创作者;如果您是本文档原作者,请点此认领!既往收益都归您。
下载文档到电脑,查找使用更方便
5000 积分 0人已下载
下载 | 加入VIP,交流精品资源 |
- 配套讲稿:
如PPT文件的首页显示word图标,表示该PPT已包含配套word讲稿。双击word图标可打开word文档。
- 特殊限制:
部分文档作品中含有的国旗、国徽等图片,仅作为作品整体效果示例展示,禁止商用。设计者仅对作品中独创性部分享有著作权。
- 关 键 词:
- ASTMF20522006E1STANDARDTESTMETHODFORMEASUREMENTOFMAGNETICALLYINDUCEDDISPLACEMENTFORCEONMEDICALDEVICESINTHEMAGNETICRESONANCEENVIRONMENT

链接地址:http://www.mydoc123.com/p-536413.html